Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Table of Contents Section 1: Introduction Brief Description ...................... 1 Intended Use ......................1 Contraindications ...................... 1 Section 2: Definitions Graphic Symbols and Meanings ................2 Glossary of Terms ....................3 Section 3: Identification Model Designation ....................
Page 4
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Setting the Sensor RF Channels................36 Section 13: Service and Repair Availability of Circuit Diagrams and Component Lists ..........37 Warranty Information ....................37 Submitting Service Requests ................... 37 Returning Equipment....................37 Section 14: List of Spare Parts and Consumables Consumable Items (electrodes and sensor elements) ..........
Noraxon U.S.A, Inc. TeleMyo DTS Belt Receiver Section 1: Introduction Brief Description The TeleMyo™ Direct Transmission System Belt Receiver for EMG and other biomechanical sensors directly transmits data from the electrode or sensor site to a Belt Receiver. This direct transmission concept greatly simplifies the arrangement of EMG measurements by eliminating cable connections between the EMG electrodes and EMG amplifier.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 2: Definitions Graphic Symbols and Meaning The following international icons and symbols are found on the TeleMyo DTS Belt Receiver enclosures and in this user manual. Their meaning is described below. Approval to market this product (#546,548) in the European Community was certified by Notified Body #2797 BSI.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Glossary of Terms DTS – (Abbreviation for Direct Transmission System) A network of short-range wireless sensors where measured data is transmitted directly from each sensor into a receiver for subsequent display and analysis on a computer or intelligent handheld device.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 3: Identification Model Designation The basic DTS Belt Receiver System consists of two primary components: Model 580 DTS Belt Receiver (1 per system) Model 546 DTS Belt Receiver EMG Sensor (up to 8 per system) Product Versions and Configurations The model 580 DTS Belt Receiver can accommodate up to 8 DTS Sensors.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 4: General Warnings and Cautions Risks and Benefits There is no identified risk of physical harm or injury with use of the TeleMyo DTS Belt Receiver product. The benefit provided by use of the device is the provision of objective...
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 5: Getting Started Quick Start Guides MR3 – Quick Start Tutorial MyoMuscle Step 1: Home/Start Screen Select a module Select/create a subject Select a configuration (to create a new configuration see next...
Page 11
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Step 3: Viewer Review/Replay the record Set a marker at each start and end of a desired analysis period Continue to next step with: MR3 – General Quick Start Tutorial Step 4: Report Selection and Analysis Period Definition...
Page 12
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Step 5: Read and Print a Report Scroll through the report results Use Print, View or Analysis Options Reports can be printed, copied to clipboard or exported to several files formats. The report can be Re-analyzed to adjust the period definition or to perform a comparison analysis.
Page 13
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver MyoResearch-XP – Quick Start Tutorial MyoMuscle P-5808 Rev M...
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 6: Preparing the Product for Use (Set-up Instructions) Unpacking and Component Identification The TeleMyo DTS Belt Receiver System is provided with a padded suitcase for protection during transport and for storage. The contents are secured in recessed areas of the padding as illustrated.
Page 15
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver DTS EMG Lead Set (part # 542AP) Qty: Matches number of EMG Sensors DTS Sensor Charging Station (part #543) DTS Sensor Charger Power Source (part #PSU1) Link Cable (part #580K) TeleMyo DTS Belt Receiver Charger 5V Supply...
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Component Inputs, Outputs and Indicators 1A Receiver (Front) Display – Indicates the status of the receiver, i.e. USB Ready, Measuring, etc. Keypad– Used to navigate device setup Charge Indicator – Green light indicates the device is charging Power Button –...
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver 2A EMG Sensor (front) Status – Sensor operational indicator flashes green. Flash rate is faster when measuring, slower when idle. Charge – Indicator illuminates steady amber while sensor is charging. When the battery is fully charged the indication is off.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Insert the opposite end of the USB cable into an available USB port on the computer. Step 2b If using WiFi with PC interface, Insert the CompactFlash WiFi cards. Insert one into the top of the DTS Belt Receiver and one into the PC Interface.
5. Please include the following: Your name, Company/Organization Name, Serial Number on TeleMyo DTS Belt Receiver and the Activation Key. 6. Noraxon Support will email or respond by phone with the Activation Code 7. Enter the provided Activation Code to remove any restrictions on use.
Page 20
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Step 2 Click on the Hardware button Step 3 Select (click on) the Telemyo DTS Belt Receiver Icon from the various devices at the top of the screen. Step 4 Click on the Settings Icon.
Page 21
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Step 5 When the DTS Belt Receiver is attached to the computer’s USB port, the DTS Belt Receiver Settings Dialog will appear. Be sure the USB Gen2 device box is checked if using the DTS Belt Receiver with the computer via USB.
Page 22
USB port of the computer. Click on the Insert Device button Step 3 Double-Click on the Noraxon DTS Belt Receiver Icon to bring up the dialog of step 4. Note: The DTS Belt Receiver Icon will not be displayed if the device is not attached to the USB port of the computer.
Page 23
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Step 6 For each DTS Belt Receiver Sensor identify its 4-character serial number and enter the value into the corresponding Serial column field. This also needs to be done in the DTS Belt Receiver. See TeleMyo DTS Belt Receiver Configuration for instructions.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver TeleMyo DTS Belt Receiver Configuration Setting Time and Date The TeleMyo DTS Belt Receiver contains a real-time clock to keep track of the current Time and Date. The clock is used primarily to create timestamps for datalogging sessions.
Page 25
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Assigning Sensors to Channels Before a measurement can be started the TeleMyo DTS must be informed which sensors it will be using. This is accomplished by assigning sensor serial numbers to each available channel.
Page 26
Press Enter. NOTE: If the RF channel you select overlaps with the WiFi frequency you will receive a warning message. Noraxon recommends that you press any key to cancel and select a different RF Channel that does not overlap. For more information, see Appendix A for a chart listing recommended combinations of Sensor RF and WiFi channels.
Enter Button NOTE: If the WiFi Channel you select overlaps with the Sensor RF channel you will receive a warning message. Noraxon recommends that you press any key to cancel and select a different WiFi Channel that does not overlap. For more information, see Appendix A for a chart listing recommended combinations of Sensor RF and WiFi channels.
EMG electrodes. The 2-pin lead wire connector can be inserted either way into the EMG Sensor to facilitate attachment to the surface electrodes. Both snap/button style and pinch (or clip) style wire terminations are available. Noraxon also offers longer lead wires for special needs.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 8: Operating Instructions Safety Information Summary Strictly follow all safety practices given in section 4 of this manual. The most critical ones are repeated here. CAUTIONS • Never use the TeleMyo DTS Belt Receiver System on a person with an implanted pacemaker •...
CompactFlash WiFi card The second mode provides a longer-range connection via a CompactFlash WiFi card. Using the WiFi card the user may connect the DTS system to any of Noraxon’s G2 receivers: TeleMyo 2400R G2 Analog Input/Output Receiver (Part #291)
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver This file can later be imported into MR-XP or MR3 for analysis. Exceptional Functions/Situations (error messages) Error Message Meaning Channel n (1<n<8) was selected for a measurement but a sensor serial number for...
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 9: Accessories and Optional Modules Accessories Part No. Image Description More… DTS Sensor Charging Station (8 sensors) PSU1 Sensor Charging Station Power Supply PSU2 TeleMyo DTS Belt Receiver Charger 5V Supply 290B...
DTS Respiration Sensor Mobie Force Sensor 542FW Fine Wire Leads 500-IS DTS Insoles-Pair As new accessories may be available after the time of printing, please check Noraxon’s website at this link for the latest offerings. http://noraxon.com/products/ Options Part No. Image Description...
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver 2400R Analog Output Receiver (16 or 32 channel options are available) TeleMyo PC Interface Receiver (default option) Interfaces to Other Devices Package Image Description Medilogic Measures foot pressure profiles using insoles Insoles Pressure Plate...
Page 35
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Digital camera For recording the test subject while they perform trials P-5808 Rev M...
The EMG Sensors are not warranted against exposure to any of the conventional forms of sterilization(autoclave, heating, etc.). Users wishing to utilize this equipment in a sterile environment, such as an operating theater, should consult Noraxon for other options. P-5808 Rev M...
Companion Software Updates • Perform a backup of the data folders to a separate drive as a precaution. • Click on the Patch/Update link provided in the email or as given on the Noraxon website: http://noraxon.com/software-downloads • Download the Patch/Update file.
Page 38
Device Software (firmware) Updates The internal program (firmware) inside the various DTS devices can be updated through the use of a special utility program available through a supplied link through the Noraxon website: http://noraxon.com/drivers-and-firmware The installed program will permit updates to both the DTS Belt Receiver and the DTS Sensors Attention All DTS sensors should be fully charged before firmware update is performed.
(esp. at long distances) of-sight relationship between sensor and receiver Website Link to FAQ Answers to common questions can be found at Noraxon’s Frequently Asked Questions (FAQ) website page at this link: http://noraxon.com/faq Other educational material is available at this link: http://noraxon.com/educational-materials...
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Radio Considerations The TeleMyo DTS Belt Receiver radio system operates in the 2400 MHz ISM (Industrial, Scientific and Medical) radio band reserved for use in most countries of the world. The radio transfers data digitally using a proprietary wireless sensor protocol. Other devices operating in this frequency band include computer networks, microwave ovens, cordless phone sets and other WiFi enabled devices.
If you are shipping from outside the USA please use UPS, FedEx, DHL, or EMS (US Postal Service) and not a freight-forwarder. Using a freight-forwarder incurs additional brokerage fees. If a package is shipped to Noraxon via a carrier other than the ones listed above, it may be refused.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 14: Spare Parts and Consumables Consumable Items Part Image Description Dual electrodes 8 per pouch or 200 per box 542C Double sided tape for attaching DTS sensors, 504 per package Replaceable Items...
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 15: Taking Product out of Operation Disposal of Equipment and Batteries The DTS Belt Receiver EMG Sensors contain Li-Polymer batteries, which may be hazardous if disposed of incorrectly. Please check with the governing authorities in your location before disposing of the TeleMyo DTS Belt Receiver and its contents.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 16: Specifications of the Product Expected Useful Lifetime Both the TeleMyo DTS Belt Receiver (580) and EMG Sensors (542) have a usable life of seven years. The DTS EMG sensors (#542) operate with a rechargeable Lithium Ion battery, as do all DTS Sensors.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver • Relative Humidity: 10% to 100% • Atmospheric Pressure: 500hPa to 1060hPa IP (Ingress Protection) Rating The DTS Belt Receiver device enclosures have a low ingress protection rating (IP20). The TeleMyo DTS Desktop Receiver, and in particular the various DTS Sensors are not waterproof.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Section 17: Technical Information Block Diagram Model 542 DTS EMG Sensor Model 580 DTS Belt Receiver P-5808 Rev M...
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Theory of Operation The DTS Belt Receiver wireless system is based on a pre-certified transceiver module: UGWG4USBB33 by Unigen. This radio module operates in the 2.4 GHz bands with an output power level of 1 mW and is based on a Wireless USB product by Cypress Semiconductor.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Electro-Magnetic Compatibility Tables Guidance and manufacturer’s declaration – electromagnetic emissions The TeleMyo DTS is intended for use in electromagnetic environment specified below. The customer or the user of the TeleMyo DTS should assure that it is used in such an environment.
Page 49
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Guidance and manufacturer’s declaration – electromagnetic immunity The TeleMyo DTS is intended for use in electromagnetic environment specified below. The customer or the user of the TeleMyo DTS should assure that it is used in such an environment.
Page 50
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Recommended separation distances between portable and mobile RF communications equipment and the TeleMyo DTS The TeleMyo DTS is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the TeleMyo DTS can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the TeleMyo DTS as recommended below, according to the maximum output power of the communications equipment.
Use a network sniffer program to determine which WiFi RF channels are being used in your area. “InSSIDer” is a network sniffer with a graphical display that is available as a free download from the Noraxon website at: http://www.noraxon.com/docs/default-source/Utility-Downloads/noraxon-inssider2- installer.zip?sfvrsn=0 This network sniffer is compatible with Windows XP, Vista and 7 (32 and 64-bit).
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Appendix B – Sensor RF Channel Frequencies The EMG Sensors and Biomechanical Sensors operate on a RF channel. The RF Channels (A- H) are assigned to the RF frequencies according to the table below.
It will usually disappear within 24 hours. Noraxon discourages any attempt to reuse a disposable electrode, even if it is simply pulled off to slightly reposition the electrode’s muscle placement. Some of the electrode gel may remain on the original site and the EMG signal may be affected.
Page 54
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver 7. It may take up to 5 minutes for disposable electrodes to fully stabilize electrically once applied to the skin. If extremely critical or precise measurements are intended, the electrodes should be applied several minutes in advance of the recording.
(next to the person’s head). At present, Noraxon identifies no restrictions on use and placement of the DTS sensors on any portion of the human body. The DTS sensors operate at radio frequencies known to effect older style pacemakers.
Noraxon U.S.A., Inc. TeleMyo DTS Belt Receiver Appendix E – Radio Regulatory Statements FCC Statement This device complies with part 15 of the FCC rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Need help?
Do you have a question about the TeleMyo DTS 580 and is the answer not in the manual?
Questions and answers