Page 1
Operating Instructions Version 5.0 HYDROMETTE BL E [Geben Sie Text ein]...
Page 2
Disclaimer GANN Mess- u. Regeltechnik GmbH does not make any assurances or warranties with respect to this document and, to the extent permitted by law, limits its liability for breach of any implied warranty to the replacement of this document with another. Furthermore, GANN Mess- u. Regeltechnik GmbH reserves the right to revise this publication at any time without having to notify anyone about such revisions.
Page 3
Table of Contents Foreword ............................6 User Description ........................6 Intended Use ........................... 6 Non-Intended Use ........................6 Explanation of the General Warnings ..................7 General Safety Instructions ..................... 8 1.5.1 Persons at Risk ........................8 1.5.2 Preparation and Start-up ......................9 1.5.3 Use / Operation ........................
Page 4
Automatic Switch-Off ......................26 Battery Monitoring ........................ 26 Querying the Device Firmware ....................26 Installation of the PC-Software GANN Dialog Pro ................. 27 USB-Communication with a PC ..................... 29 Application Notes .......................... 30 Comparative Measurement or Reference Measurement ............. 30 General Notes on Structural Moisture Measurement ............
Page 5
9.8.2 Push-in Electrode ET 10 BL ....................49 9.8.3 Surface Temperature Sensors OT 100 BL ................50 9.8.4 Immersion and Flue Gas Temperature Sensor TT 40 BL ............50 Accessories ..........................51 10.1 Accessories for Structural Moisture Measurement .............. 51 10.2 Accessories for Wood Moisture Measurement ..............
Page 6
B 55 BL active electrode. It can be used to measure the moisture content of various building, insulating and insulating materials as well as various hardwoods and softwoods. The Hydromette BL E is also used to detect moisture distribution in walls, ceilings and floors.
Page 7
1.4 Explanation of the General Warnings The following danger levels are used in this operating manual to indicate potentially dangerous situations and important safety instructions: Danger Level Description Danger / Indicates a hazardous situation which, if not avoided, will result in death or serious irreversible injuries. GEFAHR Warning / Indicates a hazardous situation which, if not avoided, could result in death or serious irreversible injuries.
Page 8
1.5 General Safety Instructions It must be ensured that the complete instructions and all safety instructions have been read and understood before using this device. All instructions must be followed. This prevents accidents that can result in property damage or minor or moderate injuries.
Page 9
1.5.2 Preparation and Start-up Never store or place the device in a location where it can fall or be drawn into water or other liquids. To avoid the risk of electric shock, never immerse the device in water or other liquids. Always remove all packaging before operating the device.
Page 10
Caution, there is a risk of explosion if the wrong type of battery is inserted. Handle used batteries according to the manufacturer's instructions. Gann Mess- u. Regeltechnik GmbH accepts no liability for damage caused by non-compliance with the operating instructions or by violation of a duty to care during transport, storage or operation of the instrument, even if this duty to care is not specifically discussed in the operating instructions.
Page 11
1.6 Specific Warnings There is a risk of injury from the measuring pins of the electrodes for resistance measurement. There is also a risk of injury due to careless handling when piercing / knocking into the material to be measured. Before the electrode pins are pressed / driven into walls or ceilings CAUTION (e.g.
Page 12
Ambient temperatures continuously too low (< 0 °C) 2.3 Transport- & Storage Conditions Die Hydromette BL E may only be stored in the packaging provided by the manufacturer or available from the manufacturer as an accessory. The manufacturer shall not accept any liability or warranty for damage that may occur to the device or to the sensor system as a result of non- compliance.
Page 13
2.4 Measuring Range Structural moisture (resistive measuring) Dimensionless parameter: 2 to 87 digits Cement screed: 2.3 to 4.8 % by weight 1.2 to 3.2 CM-% Anhydrite screed: 0.1 to 4.5 % by weight 0.1 to 4.5 CM-% Concrete C20/25: 0.9 to 2.8 % by weight Cement mortar: 1.2 to 8.9...
Page 14
Operational reliability and functionality are no longer ensured if the device is modified or adapted. Gann Mess- u. Regeltechnik GmbH shall not be liable for any damage arising from such modifications or adaptations. The risk is borne solely by the user.
Page 15
In addition to surface temperatures, material temperatures (based on Pt100) can also be measured. An internal memory is available for storing data. The Hydromette BL E has a 3-line LCD display. The silicone buttons give haptic feedback for important functions.
Page 16
3.5 mm – jack receptacle LCD-segment display Up button On / Off button Down button Measure button ("M" button) Mini-USB port valid from January 2025 Figure 5-1: Front view of the Hydromette BL E ® Hydromette BL E Version 5.0...
Page 17
5.1 Display Symbols 5.1.1 Main Menu Symbols Material moisture in % by weight or CM-% Oder CM-% "Hold" symbol Material index Building material symbol Figure 5-2: Main menu symbols 5.1.2 Other Symbols (Resistive Measuring) Symbols for minimum / maximum value display Display of the upper and lower material limits ("HI"...
Page 18
5.1.3 Symbols for non-destructive Measurement Symbols for minimum / maximum value display Actual measured value in digits "Hold" symbol Material code "0" and measuring mode symbol "c" (capacitive) for non-destructive measurement Figure 5-2: Symbols for non-destructive measurement 5.1.4 Symbols when Measuring Temperature Symbols for minimum / maximum value display Temperature display in °C...
Page 19
[see Chapter 5.3.1 "Measurement menu (main menu)"]. By default, the Hydromette BL E is delivered with the material code "r 0" (resistance measurement or resistive measurement in digits) as the factory setting. Last measured value in digits "Hold" symbol; Measured value is held;...
Page 20
If the measuring range of a selected material is exceeded or not reached, a flashing measured value warns you, which is also marked alternately with "LO" or "HI". By default, the Hydromette BL E is supplied with the material code "0" (scan mode) as the factory setting.
Page 21
5.3.2 Material Setting Measuring menu Press Down button once The device is delivered with the measuring mode "r0" (resistive measurement in digits) as factory setting. To be able to make the material settings, the device must be switched on and in main measuring mode. Then press the "Down" button once to access the material settings.
Page 22
The Hydromette BL E has a 2-stage wood type correction (type 2 and 3). The setting for the wood to be measured is taken from column "1..4". This type selection covers the types of wood commonly found on the building site.
Page 23
5.3.3 Maximum Value Display Press Down button twice Measuring menu The highest measured value of a measurement series is displayed together with the "Max" display symbol. A dash at the position of the measured value indicates that there is no maximum value (yet). If an existing maximum value should be deleted, the displayed value must be selected by briefly (<...
Page 24
5.3.4 Minimum Value Display Press Down button 3 times Measuring menu The lowest measured value of a measurement series is displayed together with the “Min" display symbol. A dash at the position of the measured value indicates that there is no minimum value (yet). If an existing minimum value should be deleted, the displayed value must be selected by briefly (<...
Page 25
5.3.5 Memory Menu Measuring menu Press Down button 4 times The ring memory symbol "o" and the corresponding memory location number are displayed. As soon as you select the memory menu, the memory location number "o1" is displayed for approx. 1 second, and then the last measured saved value contained there is displayed.
Page 26
6 Other Functions 6.1 Automatic Switch-Off If no button is pressed within approx. 90 seconds, the device switches off automatically. The current values are retained and are displayed again after the device is switched on again. 6.2 Battery Monitoring If the battery symbol is shown in the display, the battery is dead and must be replaced.
Page 27
Minimum screen resolution 1280 x 800 (1920 x 1080 is recommended) Internet connection for software downloads. updates and upgrades The PC software GANN Dialog Pro is available for download free of charge at the following link: http://download-ota.gann.de/dlg Detail information about the PC software GANN Dialog Pro can be found in the associated user manual.
Page 28
Hydromette-Settings Check for a BL E firmware update! Die Hydromette BL Compact B2-Treiber installieren BL Compact B2 Gerätetreiber Install the Hydromette BL E Drivers BL E Drivers BL E Workspace Figure_7_2: Download device drivers of the Hydromette BL E To download the device drivers, the working range of the desired Hydromette must first be selected in the menu item "Select working range".
Page 29
Figure 7-2) If the Hydromette BL E is connected to a PC with Windows operating system when it is switched off, the Hydromette starts in USB mode. No measurements can be performed during the communication with the PC. The GANN Dialog Pro software now provides the possibility to update the firmware of the Hydromette BL E via the Internet.
Page 30
9 Application Notes 9.1 Comparative Measurement or Reference Measurement With this type of measurement, almost all (set) building materials or mixed materials or mixed structures can be measured comparatively. It is important that these measurements are only carried out on the same materials or structures. A deliberately dry spot must be determined on the structure to be measured.
Page 31
(wt. %) in relation to the dry weight or also in CM-% (moisture determination according to the calcium carbide method). Depending on the type of GANN Hydromette used, this is done using programmed characteristic curves or independent conversion using tables.
Page 32
9.3 Notes on resistance-based Structural Moisture Measurement The GANN Hydromettes work according to the electrical resistance and conductivity measurement method that has been known for years. This process is based on the fact that electrical resistance is highly dependent on the amount of moisture in the material.
Page 33
The following applies in principle: - Generally, both electrode pins are to be applied to the same contiguous material to be measured. INFORMATION - Before drilling holes for probes or hammering electrode pins into walls, ceilings, floors, etc. make absolutely sure by suitable means that there are no electrical cables, water pipes or other supply lines in this location.
Page 34
9.4 Using the Hydromette BL E The Hydromette BL E is a multi-purpose measuring instrument for measuring structural moisture, wood moisture and temperature. The device utilises both the resistance measuring principle and the non-destructive capacitive measuring principle. This means that the device can be used in conjunction with different (active) electrodes depending on the measuring task.
Page 35
Measure: Press and hold the "M" button for longer than 2 seconds. A measuring process is carried out as long as the measurement button is pressed. After releasing the "M" button, the measuring process is interrupted and the "Hold" symbol is INFORMATION displayed.
Page 36
Please note: Any surface contaminants (e.g. formwork release oil) may cause errors. INFORMATION 9.4.3 M 6 Stick-in Electrode The two electrodes that are intended to measure set construction materials only are to be pushed into the material to be measured with approx.8 - 10 cm clearance. Where this is not possible because of the hardness of the material to be measured (floor screed, concrete etc.), holes of approx.
Page 37
For the M 6-150 probes with 3 mm Ø, which were specially developed for measuring through a spacer cross, a special carbide drill bit with a length of 160 mm and a diameter of 3 mm is required. This can be used to drill through the screed layer to the insulation. The distance between the probes should be approx.
Page 38
9.5 General Information on Wood Moisture Measurement The wood moisture in the GANN Hydromettes® is displayed in per cent by weight (wt.-%) in relation to absolutely dry wood (bone-dry). If wood is stored for extended periods in a given climate it assumes a moisture content corresponding to this climate, which is also termed equilibrium moisture or wood moisture equilibrium.
Page 39
9.6 Notes on resistance-based Measuring Wood Moisture The GANN Hydromettes works by measuring electrical resistance or conductivity, a principle which has been well-known for many years. This process is based on the fact that electrical resistance is highly dependent on the amount of moisture in the wood. The conductivity of oven-dried wood is very low, or the resistance is so high that no current worth mentioning can flow.
Page 40
> 20 °C they are too high. It is therefore necessary to correct the values obtained using an appropriate correction table. Various GANN Hydromettes® are already equipped with such temperature compensation, i.e. the wood temperature can be set directly on the measuring instrument and is automatically taken into account in the wood moisture display.
Page 41
using a spanner (WAF 12) or pliers. Loose electrode pins tend to break more easily and cause unstable measured values. If thicker wood is to be measured, the electrode pins can be replaced with a suitably longer version. With increasing tip length, however, increased danger of breakage or bending (particularly when pulling out) must be expected.
Page 42
Any wood particles adhering to the measuring surface must be regularly removed. If the elastic, plastic measurement sensors are damaged, they can be reordered (no. 31004316) and glued on using standard cyanate-based instant adhesive. 9.6.5 M 20-DS 16 and M 20-DS 16-i Conversion Kit The M 20-DS 16 conversion kit is used to measure the moisture content of wood up to 30 mm thick.
Page 43
9.6.7 M 18 Ram-in Electrode The ram-in electrode M 18 is made of corrosion-resistant V2A steel and is suitable for (depth) measurements in up to 180 mm thick or hard wood, wood fibre insulation materials and wood composites. When first delivered, the ram-in electrode M 18 is supplied with 10 replacement pins of 40 and 60 mm length (not insulated).
Page 44
Therefore, no binding statements can be made in this regard. Of course, this problem does not only apply to capacitive measuring instruments from GANN, but is the physical basis for all moisture probes and sensors that work dielectric constant (DK), high- frequency or microwave basis.
Page 45
"M" button for longer than 2 seconds. In conjunction with the active electrode B 55 BL, the Hydromette BL E only works in digit scan mode (setting "c 0"). A material setting or direct display in weight or CM % is not possible.
Page 46
Measured value in digits for control measurement in the air Non-destructive measurement mode "c 0" Figure 9-4: Capacitive measuring mode "c 0" display Measuring: Press the measurement button “M” for longer than 2 seconds, and scan the area to be inspected. The electrode must rest firmly on the building material and be held as vertically as possible (approx.
Page 47
In the corner/angle area, it is essential to maintain a distance of approx. 8 - 10 cm from the edge/angle. Figure 9-6: Correct use of distances when measuring Measurements directly in the corner/angle area lead to an overlapping of the measuring field and thus change the measured value! Figure 9-7: Incorrect use in corner/angle area To (re-)activate the BNC socket or the resistance-based measurement, the...
Page 48
The measuring instrument now automatically recognises the connected accessories. To activate the temperature measuring mode, press now the "M" button for longer than 2 seconds. In conjunction with the Pt100 temperature sensors, the Hydromette BL E [BL UNI 11] operates exclusively in temperature display mode. A material setting or direct display in weight or CM % is not possible.
Page 49
Measured value in °C Figure 9-8: Temperature measurement display Measure: Press and hold the "M" button for longer than 2 seconds. A measuring process is carried out as long as the measurement button is pressed. After releasing the "M" button, the measuring process is interrupted and the "Hold" symbol is displayed. INFORMATION 9.8.2 Push-in Electrode ET 10 BL...
Page 50
9.8.3 Surface Temperature Sensors OT 100 BL The OT 100 BL is a special sensor with a particularly low mass for measuring temperatures on surfaces. It has a spring-mounted sensor tip with thermal separation and the resulting optimised measured value acquisition, e.g. on solid wall surfaces. The sensor plate must be in full contact.
Page 51
10 Accessories 10.1 Accessories for Structural Moisture Measurement Structural moisture test adapter (order no. 31006071) Test adapter for structural moisture, for checking structural moisture measuring instruments with accessories. M 20 drive-in electrode (order no. 31003300) The drive-in electrode M 20 is made of impact-resistant plastic and can be used for measurements in soft, set building materials (e.g.
Page 52
M 6 stick-in electrode pair (order no. 31003700) The M 6 stick-in electrode pair can be used for resistance-based building moisture measurement in hard, set building materials (e.g. sand-lime brick, cement screed or concrete) in conjunction with contact compound or in soft building materials (such as mortar joints, plaster or drywall panels).
Page 53
M 21-100/250 electrode pair for deep measurement The depth measuring electrode pairs M 21-100 and M 21-250 have been developed for measuring building moisture, especially for depth measurements in building materials. By using the depth measuring electrodes, a moisture profile of the material to be measured can be created by taking measurements layer by layer.
Page 54
MK 8 Measuring cable - length 1 m (order no. 31006210) For connection of electrodes for resistance measurement Connection cable MK 26 – Length: 1.80 m (order no. 31016920) For device connection with a USB port. valid from January 2025 Active Electrode B 55 BL (order no.
Page 55
10.2 Accessories for Wood Moisture Measurement Wood moisture test adapter (order no. 31006070) Test adapter for wood moisture, for check of wood moisture measuring instruments with accessories. M 20 drive-in electrode (order no. 31003300) The drive-in electrode M 20 is made of impact-resistant plastic and is suitable for measuring wood thicknesses of up to 50 mm (e.g.
Page 56
M 20-OF 15 surface measuring caps (order no. 31004315) The surface measuring cap pair is suitable for moisture measurements on surfaces without damaging the material to be measured in conjunction with the M 20 electrode (e.g. veneers, solid or multi-layer parquet). Depth effect 2–5 mm.
Page 57
M 18 ram-in electrode (order no. 31003500) The ram-in electrode M 18 is made of corrosion-resistant V2A steel and is suitable for (depth) measurements in up to 180 mm thick or hard wood, wood fibre insulation materials and wood composites. When first delivered, the ram-in electrode M 18 is supplied with 10 replacement pins of 40 and 60 mm length (not insulated).
Page 58
10.3 Temperature Measurement Accessories ET 10 BL stick-in temperature sensor (order no. 31013165) The ET 10 BL is a simple penetration temperature sensor for measuring temperatures in semi-solid materials (e.g. frozen goods), bulk materials, liquids and for measuring core temperatures in a drill hole. OT 100 BL Surface temperature sensor (order no.
Page 59
11 Appendix 11.1 Material Table Material index Material Display in digits / Scan mode resistive Wood type 2 Wood type 3 Cement screed in weight-% Anhydrite screed in weight-% Cement mortar in weight-% Lime mortar in weight-% Gypsum plaster in weight % Cement screed in CM % Lime sand brick in weight-% Styrofoam in weight %...
Page 60
11.2 Display Values (Digits) in Relation to the Bulk Density Corresponding Relative Air Humidity in % 30 ------------ 50 ---------------- 70 --------------- 80 ---------------- 90 --------------- 95-----------100 Display in Digits* Bulk density kg/m³ very dry normal dry semi dry moist very moist up to 600 10 –...
Page 62
11.4 Equilibrium Wood Moisture Table Wood Moisture Equilibrium Air temperature in °C 10 °C 15 °C 20 °C 25 °C 30 °C Relative air humidity Wood moisture content 4.70% 4.70% 4.60% 4.40% 4.30% 6.30% 6.20% 6.10% 6.00% 5.90% 7.90% 7.80% 7.70% 7.50% 7.50%...
Page 64
11.6 General Concluding Remarks The notes and tables in these operating instructions on permitted or normal humidity conditions in practice and the general definitions of terms have been taken from the specialist literature. No responsibility can therefore be taken by the manufacturer of the measuring device for the correctness of this information.
Page 65
Applied harmonized standards: ☒ EN 61326-1 : 2013 General EMC requirements ☒ EN IEC 63000 : 2018 Restriction of hazardous substances This declaration is given in responsibility for: Gann Mess- und Regeltechnik GmbH Schillerstr. 63 70839 Gerlingen Germany issued by:...
Need help?
Do you have a question about the HYDROMETTE BL E and is the answer not in the manual?
Questions and answers