STIRLING VAULT100 ™ OPERATING MANUAL -20°C to -100°C Upright Ultra-Low REDEFINING COLD Temperature Freezer...
Page 2
THIS PAGE INTENTIONALLY LEFT BLANK Used, Risk of Fire or To disable, enter Current PIN and complete the Explosion process leaving New PIN and Confirm New PIN blank. VAULT100 uses 90 grams of Tap Configurations Icon > Enter PIN* >...
Page 3
The Stirling Ultracold Model M D free-piston Stirling engine produces less heat than a traditional compressor-style engine. It provides quiet, high-e ciency, ULT cooling without the use of HCFC or HFC refrigerants. The M D Stirling engine uses approximately grams ( .
Page 4
Always take proper precautions when operating ultra-low temperature freezers. is sealed with a single 7-chamber, 2-rib gasket, that is replaceable. An integrated The Stirling VAULT100 is intended for storage of frozen sample products or vials at gasket heater minimizes ice deposits on the gasket to cabinet interface.
Page 5
2.4.1 Ultra-Low Temperature (ULT) Precautions to indicate the potential hazards associated with operating the Stirling VAULT The Stirling VAULT100 can operate as low as -100°C (-148°F). The Freezer. following precautions are related to hazards working at such low Information Symbol: Consult the Operating Manual in all cases where the temperatures and the impact it may have on sample storage materials.
Page 6
Adequate ventilation must be maintained when Potential hazards associated with installation and maintenance of the using these systems. Stirling VAULT100 may impact the safety of persons who install, maintain, and transport the freezer. All personnel who intend to perform these 2.4.2...
Page 7
If none are available, it is recommended to contact Stirling • CAUTION: Do not cut, change, or modify the power cord.
Page 8
• CAUTION: Do not drill or puncture the cabinet liners when mounting • CAUTION: Do not modify the cabinet, controls, or free-piston Stirling earthquake straps or other accessories. This will likely result in irreparable engine.** damage to the insulation.
Page 9
Ethernet Port ACCESSORY PANEL Recessed Accessory Panel Access port, with retained plug Top covers Mechanical compartment containing Stirling engine and control boards (not shown, covered by top covers). LCD, GUI, and flash drive connector Outer door and gasket Vacuum relief port...
Page 10
INSTALLATION AND SETUP PROCEDURE INSTALLATION AND SETUP PROCEDURE (Continued) Prior to installing your Stirling VAULT100, inspect the unpacked unit and any EXTERNAL DIMENSIONS AND OPERATING FLOOR PLAN included items for shipping damage. Compare all contents to the Packing List Be aware that the shipping packaging dimensions 84 H x 43 D x 46 W in. (2134 x (Section 4.3) for completeness.
Page 11
Disinfect with suitable sterilizing agent if the freezer has been used for biohazards. Do not use bleach or abrasive cleaning pads. stirlingultracold.com 20 OF 65 12.4.2024 VAULT100 LIT OM A 66 OF 54 A OF 54 OPERATING MANUAL VAULT100 21 OF 65...
Page 12
Use this product only in the way described in the product literature and in this manual. Before using your Stirling VAULT100 verify that this product is suitable for its Arrange and install the shelves as desired. See Section 9 for details. Each shelf intended use.
Page 13
Connections: View connectibility mode, IP address, and BACnet ID. Data Export: Allows users to transfer data files to a flash drive. > stirlingultracold.com A OF 54 24 OF 65 OPERATING MANUAL VAULT100 12.4.2024 VAULT100 LIT OM A 25 OF 65 66 OF 54...
Page 14
OPERATION (Continued) OPERATION (Continued) Guide to Basic GUI Functions Contact: Displays Stirling Ultracold contact information. 5.4.3 > > About: Displays various information including freezer model & serial number, ON / OFF Switch the main power on, located at the back of the unit engine serial number, power up &...
Page 15
Alarms. Audible alarms will sound, and a Mute icon will appear in the bottom-right corner. Select this icon to temporarily silence the alarms for a set duration. stirlingultracold.com A OF 54 28 OF 65 OPERATING MANUAL VAULT100 12.4.2024 VAULT100 LIT OM A 29 OF 65 66 OF 54...
Page 16
52 weeks. Alarm Descriptions The Stirling VAULT100 is equipped with several alarm codes that display when ap- plicable on the freezer's GUI screen. Freezer alarms are accompanied by an audible sound in the unit enclosure and can be connected to an external alarm. The GUI provides settings that allow the user to temporarily suppress or delay alarms.
Page 17
OPERATION (Continued) OPERATION (Continued) Connections screen displays the connectivity mode, IP address, and BACnet ID. Contact screen displays contact information for Stirling Ultracold. 5.4.9 Information The Information menu includes five tabs: Event Log, Connections, Data Export, Contact, and About. About screen displays general information including freezer mode, freezer serial Event Log screen displays all events.
Page 18
Navigating to the Predictive Analytics menu will prompt this summary message - tap the screen to dismiss. stirlingultracold.com 35 OF 65 A OF 54 34 OF 65 OPERATING MANUAL VAULT100 12.4.2024 VAULT100 LIT OM A 66 OF 54...
Page 19
Gasket Heater screen allows users to change the gasket heater duty cycle from 0% to 100%. Increase the percentage for more defrost heating when there is ice buildup around the door. stirlingultracold.com 36 OF 65 37 OF 65 A OF 54 OPERATING MANUAL VAULT100 12.4.2024 VAULT100 LIT OM A 66 OF 54...
Page 20
PIN to confirm. If users go through the process without entering any numbers, the PIN will be disabled and will not be required to enter the Settings menu. stirlingultracold.com .39 OF 65 A OF 54 38 OF 65 OPERATING MANUAL VAULT100 12.4.2024 VAULT100 LIT OM A 66 OF 54...
Page 21
RTD Calibration Screen is used to compensate for discrepancies between the Display RTD and a user provided standard. This screen requires Authorized Service & PIN, if enabled. stirlingultracold.com 41 OF 65 A OF 54 40 OF 65 OPERATING MANUAL VAULT100 12.4.2024 VAULT100 LIT OM A 66 OF 54...
Page 22
PIN will be disabled and will not be required to enter the Service menu. MANAGING FREEZER CONTENTS The Stirling VAULT100 freezer is designed for long-term storage of sample materials with ultra-low, well-regulated temperature storage requirements. The temperature setpoint for the freezer should be changed appropriately for the materials being stored using the Graphical User Interface (GUI) –...
Page 23
Air Filter Replacement PREVENTATIVE MAINTENANCE SCHEDULE Recommend air filter replacement every 6 months. The Stirling VAULT100 freezer is designed for years of trouble-free operation. To Turn the power switch OFF. > prevent costly and inconvenient repairs, and to maintain your freezer at an optimum level of performance, follow the recommended preventative maintenance Remove air filter cover by using a screwdriver or turning thumb knobs...
Page 24
OPERATOR MAINTENANCE 6.2.1 Cleaning and Disinfection Clean the exterior surfaces of the Stirling VAULT100 cabinet as needed by using a soft cloth and mild detergent. Do not use solvents (such as bleach) or harsh abrasive cleansers or pads. 46 OF 65 stirlingultracold.com...
Page 25
Platform. For details on how to configure the freezer please refer to the Building LN2 BACKUP SYSTEM OPTIONAL : Management Software Platform and its software administrator. Contact Stirling Ultracold Service or an authorized service provider to install an DRY CONTACTS optional LN2 Backup System.
Page 26
Shelf #2 60.7” Height (top and bottom clearance), without shelves. One inner door to access 2 compartments. Shelf #3 Refer to Stirling Ultracold Rack Solutions catalog – contact Stirling Ultracold sales for details. Shelf #4 Shelf #5 2 Shelves, 3 Compartments Storage: Custom size >...
Page 27
The angled section of the clip is inserted in slot #3 The top inner door is used to access shelf #1 4th rivet Shelf #2 from top stirlingultracold.com 53 OF 65 A OF 54 52 OF 65 OPERATING MANUAL VAULT100 12.4.2024 VAULT100 LIT OM A 66 OF 54...
Page 28
The angled section of the clip is inserted in slot #2 The bottom inner door is used to access shelf #5 7th rivet from top Shelf #4 stirlingultracold.com 55 OF 65 A OF 54 54 OF 65 OPERATING MANUAL VAULT100 12.4.2024 VAULT100 LIT OM A 66 OF 54...
Page 29
Maximum current is 10 Amps at 120 V. No surge Current Fuse current occurs at start and stop (3AB fuse 12A/250 Cooling Engine R704 (Helium) charged free-piston Stirling engine VAC slow) 10 g, (0.35 oz) Control Sensor One RTD (PT...
Page 30
Powder Coated Hinges Steel White Powder Coated LCD Bezel Black Injection Molded Conduit (Cable Stainless Steel Silver Natural Routing*) stirlingultracold.com 59 OF 65 A OF 54 58 OF 65 OPERATING MANUAL VAULT100 12.4.2024 VAULT100 LIT OM A 66 OF 54...
Page 31
(handle side). Change air filter. recharge. Ambient air temperature Ensure ambient air conditions * Stirling VAULT100 is equipped with a battery backup to maintain display in the above 35°C are cooler than 35°C. event of power loss. Ensure gasket is clean and free of debris and ice.
Page 32
Global Cooling, Inc. service, Global Cooling, Inc. reserves the right to exchange the Unit with a The Stirling Ultracold freezer is warranted for a full SIX YEARS for all labor and serviceable new or previously used replacement at its discretion.
Page 33
CLAIMED TO ARISE BY OPERATION OF LAW, CUSTOM, TRADE USAGE, OR COURSE OF DEALING. THIS PAGE INTENTIONALLY LEFT BLANK stirlingultracold.com 65 OF 65 A OF 54 64 OF 65 OPERATING MANUAL VAULT100 12.4.2024 VAULT100 LIT OM A 66 OF 54...
Page 34
Stirling Ultracold, Global Cooling, Inc. All Rights Reserved. Global Cooling technology is manufactured under U.S. and International patents. Stirling Ultracold is a trademark of Global Cooling, Inc. Specifications subject to change without notice. Refer to stirlingultracold.com for the latest specifications.