Eizo FLEXSCAN L560T-CB Manual page 42

Touch panel color lcd monitor
Hide thumbs Also See for FLEXSCAN L560T-CB:
Table of Contents

Advertisement

Recommended separation distances between portable and mobile RF communications
equipment and the FlexScan L560T-CB
The FlexScan L560T-CB is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the FlexScan L560T-CB can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the FlexScan L560T-CB as recommended below, according
to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
"d" in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where
"P" is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Cable length
Power Cord :
Signal Cable (MD-C87) :
USB Cable (MD-C93) :
Signal Cable (FD-C39) :
Separation distance according to frequency of transmitter
150kHz to 80MHz
d = 1.2 √ P
0.12
0.38
1.2
3.8
12
Accessary
2.0m
Accessary
1.8m
Accessary
1.8m
Option
2.0m
m
80MHz to 800MHz
d = 1.2 √ P
0.12
0.38
1.2
3.8
12
800MHz to 2.5GHz
d = 2.3 √ P
0.23
0.73
2.3
7.3
23

Advertisement

Table of Contents
loading

Table of Contents