This unit must be used with the guidance of a Physiotherapist or Doctor.
*
Type BF equipment, Continuous Operation.
*
Do not insert lead wires into a mains power supply.
*
Do not immerse unit into water or any other substance.
*
The unit is not protected from the ingress of water droplets from a shower of rain if used
outside the carrying case.
*
Do not use the NeuroTrac® MyoPlus Pro unit in the presence of a flammable anaesthetic
gas mixture and air or with Oxygen or Nitrous Oxide.
*
If using rechargeable 9 Volt PP3 Nickel Metal Hydride batteries, be sure to use a CE
approved battery charger. Never connect the NeuroTrac® MyoPlus Pro directly to a
battery charger or to any other mains powered equipment. We advise not to use Ni-Cad
rechargeable batteries.
*
Patient Electrodes are for single patient use only.
*
Keep out of reach of children.
*
Do not use this stimulator on your facial area unless you are under strict guidance from a
qualified Clinician.
*
Application of electrodes near the thorax may increase the risk of cardiac fibrillation.
*
Operation in close proximity (e.g. 1m) to shortwave or microwave therapy equipment
may produce instability in the stimulator output.
*
Simultaneous connection of a patient to high frequency surgical equipment may result in
burns at the site of the stimulator electrodes and possible damage to the stimulator.
*
This device can deliver current densities in excess of 2 mA/cm
intensity with small electrodes.
*
No modification of this equipment is allowed!

Warnings