Platinum ® Oxygen Concentrator IRC10LXO2 HomeFill System ® Compatible User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
Contents 1. Introduction ..................3 7.3.1 Setting Up the Humidifier (If Prescribed).....17 2. General ....................4 7.3.2 Connecting/Positioning the Nasal Cannula ....20 7.3.3 Turning the Concentrator On ........22 2.1 Symbols ..................4 7.3.4 Flowrate...............22 2.2 Intended Use ................5 7.3.5 SensO2 Oxygen Purity Indicator .........23 2.3 Indications For Use ..............6 7.3.6 Initial Startup of the Concentrator.......23 2.4 Description ................6...
React Health Platinum 10L Oxygen Concentrator User Manual ® 2. General Symbols in Documentation 2.1 Symbols General Warning Sign Signal words are used in this manual and apply to hazards or unsafe prac- The background color inside the triangle is tices which could result in personal injury or property damage.
React Health Platinum 10L Oxygen Concentrator User Manual ® Indicators Transport and Storage Humidity Symbol O 2 Purity Indicator Lights (LED) Alternating Current SYSTEM OKAY GREEN Indicator Light O2 over 85% YELLOW Indicator Light Type BF equipment O2 between 73% A.
React Health Platinum 10L Oxygen Concentrator User Manual ® The React Health Platinum concentrator uses a molecular sieve and pres- WARNING! sure swing adsorption methodology to produce the oxygen gas output. Risk of Injury or Damage Ambient air enters the device, is filtered and then compressed.
React Health Platinum 10L Oxygen Concentrator User Manual ® 3. Safety Serial Number Label is located on the resonator intake 3.1 Label Locations assembly. Specification and Double Insulation Label is located on the back of the concentrator at the base...
In case of damage, or if the product is not working Textiles, oil or petroleum substances, grease, greasy sub- correctly, contact a technician or React Health for repair. stances and other combustibles are easily ignited and burn with great intensity in oxygen enriched air and when in con- •...
Page 9
Accessories designed by other manufactur- in contact with oxygen under pressure. Smoking during ers have not been tested by React Health and are not oxygen therapy is dangerous and is likely to result in burns recommended for use with React Health products.
Page 10
React Health Platinum 10L Oxygen Concentrator User Manual ® DANGER! WARNING! Risk of Injury or Death Risk of Injury or Damage To avoid choking or ingestion of chemicals from airway To prevent injury or damage from cord misuse: contamination: •...
WARNING! • Users should contact their dealer or React Health for service. Use of accessories and cables other than those specified or provided by React Health could result in increased electro- magnetic emissions or decreased electromagnetic immuni- ty of this equipment and result in improper operation.
React Health Platinum 10L Oxygen Concentrator User Manual ® 4. Components Power Switch 4.1 Component Identification Circuit Breaker Elapsed Time Meter Cabinet Filter/Air Inlet Rear View Power Cord HomeFill Outlet Fitting Exhaust Accessories (Not Shown): HomeFill home oxygen compressor — IOH200 The HomeFill outlet fitting I is to be used only for filling oxygen cylinders with the HomeFill home oxygen compressor.
React Health Platinum 10L Oxygen Concentrator User Manual ® 5. Accessories 5.1 Optional Accessories There are many different types of humidifiers, oxygen tubing, cannulas and masks that can be used with this device. You should contact your local home care supplier for recommendations on which of these devices will be best for you.
React Health Platinum 10L Oxygen Concentrator User Manual ® Unless the oxygen concentrator is to be used IMMEDIATELY, leave concentrator in its packaging for storage until use is 6. Setup required. 1. Check for any obvious damage to the carton or its contents. If dam- 6.1 Unpacking...
React Health Platinum 10L Oxygen Concentrator User Manual ® 7. Usage 7.2 Select a Location 7.1 Introduction WARNING! Risk of Injury Your oxygen concentrator is indicated for individual use indoors. It is an electronically operated device that separates oxygen from room air. It...
React Health Platinum 10L Oxygen Concentrator User Manual ® Humidifier Bottle with Cap Humidifier Bottle without Cap You may select a room in your house where using your oxygen concentra- tor would be most convenient. Your concentrator can be easily rolled from room to room on its casters.
Page 18
React Health Platinum 10L Oxygen Concentrator User Manual ® 3. Insert a flathead screwdriver in the plate groove on the top edge of the filter access door and gently pry the filter access door off. WARNING! Risk of Injury To avoid impacting the flow of oxygen: •...
Page 19
React Health Platinum 10L Oxygen Concentrator User Manual ® When using tubing connections longer than 7 ft (2.1 m) in length, position the humidifier as close to the patient as pos- sible to allow for maximum humidification output. Ensure the humidifier is stable and will not tip-over.
React Health Platinum 10L Oxygen Concentrator User Manual ® 7.3.2 Connecting/Positioning the Nasal Cannula WARNING! Risk of Injury or Death DANGER! Risk of Injury or Death To reduce the risk of injury or death from illness: The cannula can cause tripping, falling, or other injury if im- •...
Page 21
React Health Platinum 10L Oxygen Concentrator User Manual ® Fig. 7-1 Concentrator without Humidifier Fig. 7-2 Concentrator with Humidifier 3. Perform one of the following: 1. Remove the cannula from its packaging. • For use with a humidifier bottle—Attach the patient’s nasal can- 2.
React Health Platinum 10L Oxygen Concentrator User Manual ® 7.3.4 Flowrate WARNING! Risk of Injury or Damage WARNING! Risk of Minor Injury or Discomfort Failure to properly install the humidifier bottle or other accessories to the concentrator will impact the flow of It is very important to select the prescribed oxygen flow oxygen.
React Health Platinum 10L Oxygen Concentrator User Manual ® 2. If the flowrate on the flowmeter ever falls below 2 L/min for more than about one minute, the Potential Obstruction Alert will be triggered. This is a rapid beeping of the audible alarm. Check your tubing or accessories for blocked or kinked tubing or a defective hu- midifier bottle.
React Health Platinum 10L Oxygen Concentrator User Manual ® 7.3.7 Explanation of Oxygen Purity Indicator Lights O2 Purity Indicator Light (LED) Explanation SYSTEM OKAY O2 GREEN light No action is needed. Normal for operation. over 85% O2 between 73% YELLOW light (Solid) Immediately call supplier.
React Health Platinum 10L Oxygen Concentrator User Manual ® 7.3.8 Elapsed Time Meter (Hour Meter) The hour meter displays the cumulative number of hours the unit has operated. Refer to “4.1 Component Identification” on page 12 for specific location.
At a minimum, preventive maintenance MUST be performed Risk of Injury or Damage according to the maintenance record guidelines. In places React Health oxygen concentrators are specifically de- with high dust or soot levels, maintenance may need to be signed to minimize routine preventive maintenance. To performed more often.
React Health Platinum 10L Oxygen Concentrator User Manual ® 8.3 Cleaning the Cabinet Filter CAUTION! Risk of Damage To avoid damage to the internal components of the unit: • DO NOT operate the concentrator without the filter in- stalled or with a dirty filter. There are two cabinet filters located on each side of the cabinet.
React Health Platinum 10L Oxygen Concentrator User Manual ® 8.4 Humidifier Cleaning and Thermic Disinfection 1. Dispose of and replace all patient side accessories not suitable for multiple patient use, including but not limited to: Clean and disinfect the oxygen humidifier daily to reduce limestone •...
React Health Platinum 10L Oxygen Concentrator User Manual ® 8.6 Preventive Maintenance Checklist Model No: Serial No: ON EACH INSPECTION Record Date of Service Record Elapsed Hours on Hour Meter Clean Cabinet Filter(s) (Refer to Cleaning the Cabinet Filter.) Check Prescribed L/min. Flowrate...
® 9. After Use 9.2 Wear and Tear React Health reserves the right to ask for any item back that has an al- 9.1 Recycling Information leged defect in workmanship. See Warranty that shipped with the product for specific warranty information.
React Health Platinum 10L Oxygen Concentrator User Manual ® 10. Troubleshooting 10.1 Troubleshooting SYMPTOM PROBABLE CAUSE SOLUTION Alarm: Main Power Loss: 1. Insert plug into outlet. No lights illuminated. 1. Power cord not plugged in. 2. Inspect house circuit breakers or fuses. If problem recurs, use a different 2.
React Health Platinum 10L Oxygen Concentrator User Manual ® 11. Technical Data 11.1 Specifications Electrical Requirements: 120 VAC + 10, –15% (132 VAC/102 VAC), 60 Hz Rated Current Input: 4.9 A < 58 dBA Average weighted for 10 L/min flow (Tested per ISO 3744:2010 with microphone location as specified in ISO 8359:1996 subclause 4.6)
Page 33
React Health Platinum 10L Oxygen Concentrator User Manual ® Current overload or line surge shutdown. High temperature compressor shutdown. High Pressure Alarm with compressor shutdown. Safety System: Low Pressure Alarm with compressor shutdown. Battery Free Power Loss Alarm. SensO2 Oxygen System.
React Health Platinum 10L Oxygen Concentrator User Manual ® 11.2 Electromagnetic Compliance (EMC) Guidance and manufacturer´s declaration — electromagnetic emission The Device is intended for use in the electromagnetic environment specified below. The customer or the user of the Device should assure that it is used in such an environment.
React Health Platinum 10L Oxygen Concentrator User Manual ® Guidance and manufacturer´s declaration — electromagnetic immunity The Device is intended for use in the electromagnetic environment specified below. The customer or the user of the Device should assure that it is used in such an environment.
Page 36
React Health Platinum 10L Oxygen Concentrator User Manual ® Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance Portable and mobile RF communications equipment should be used no closer to any part of the Device including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Page 37
React Health Platinum 10L Oxygen Concentrator User Manual ® Recommended separation distances between portable and mobile RF communications equipment and the Device. The Device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Device as recommend- ed below, according to the maximum output power of the communications equipment.
This warranty is extended only to the original purchaser who purchases this product when new and unused from React Health or a dealer. This LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL warranty is not extended to any other person or entity and is not trans- NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, ferable or assignable to any subsequent purchaser or owner.
Page 39
React Health Platinum 10L Oxygen Concentrator User Manual ® NOTES:...
Page 40
React Health 5101 Fruitville Road, Suite 200 Sarasota, FL 34232 United States (863) 226-6285 www.reacthealth.com Manufacturer React Health 5101 Fruitville Road, Suite 200 Sarasota, FL 34232 USA LBL-00373-000, Rev A...
Need help?
Do you have a question about the Platinum 10L and is the answer not in the manual?
Questions and answers