Regularly inspect the tubing and replace it when necessary to safeguard both operators and equipment. 1.2 Repair Notes Please contact Longer Pump or its distributor, and provide the product serial number before returning the product. Products which has been contaminated with, or exposed to, toxic chemicals or any other substance hazardous to health must be decontaminated before returning to Longer Pump or its distributor.
2 Product Introduction 2.1 Main Features The dPOFLEX BPH01 is a peristaltic pump head designed and manufactured by Longer Pump. The pump head is compatible with silicone tubing and Pharmed tubing in sizes 13#, 14#, 19#, 16# and 25#, and the max flow rate is 530mL/min.
To meet the requirements of different pressure conditions, the pump head compression block can be set for standard pressure and high pressure operation. The factory default setting is standard pressure. If a high-pressure setting is required, please inform Longer Pump or your local distributor when placing your order. Max System Pressure(MPa) Compression...
BPH01 Pump Head Operating Manual 3 Installation 3.1 Pump Head Mounting Pump Head Mounting Dimensions 3.1.1 Unit: mm 3.1.2 Panel Mounting OD of the adapter: minФ6mm, max Ф10mm Adapter 7 / 11...
BPH01 Pump Head Operating Manual 3.1.3 Mounting on Standard Pump Drive of Longer Pump Mounting Bracket Mounting Bracket Adapter Pump Drive 4pcs M4x10 Cross countersunk head screws 2pcs M4x10 Hexagon socket head screws a) Mount the mounting plate on the pump drive by tightening four mounting screws (M4x10 cross countersunk head screw).
BPH01 Pump Head Operating Manual 3.2 Tubing Loading The spring-loaded tubing clamps can be set at two positions to accommodate various tubing inner diameters. Tubing Size ID (mm) Inner Position √ √ √ √ × Outer Position × ×...
BPH01 Pump Head Operating Manual (From the inner position to the outer position) (From the outer position to the inner position) 1. From the inner position to the outer position (from small tubing to large tubing) Fully open the flip cover. Place a pointed tool into the groove of the tubing clamp. Press the clamp down, then push it forward towards the front casing of the pump head.
Page 11
BPH01 Pump Head Operating Manual Recommended tubing length (65-70mm) Fully open the flip cover. (Figure (1)) Check if the tubing clamp position on both sides of the pump head matches the tubing size to be loaded. If not, set the tubing clamp position accordingly. (Figure (2)) ...
Need help?
Do you have a question about the dPOFLEX BPH01 and is the answer not in the manual?
Questions and answers