Teledyne 200EU Training Manual
Teledyne 200EU Training Manual

Teledyne 200EU Training Manual

Noy nitrogen oxides analyzer

Advertisement

Quick Links

TRAINING MANUAL
MODEL 200EU / T200U NOy
NITROGEN OXIDES ANALYZER
© TELEDYNE ADVANCED POLLUTION INSTRUMENTATION
(TAPI)
9480 CARROLL PARK DRIVE
SAN DIEGO, CA 92121-5201
TOLL-FREE: 800-324-5190
TEL: 858-657-9800
FAX: 858-657-9816
E-MAIL: api-customerservice@teledyne.com
WEB SITE: www.teledyne-api.com
06041 Rev. B
DCN 6465
08 May 2012
Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 200EU and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Teledyne 200EU

  • Page 1 TRAINING MANUAL MODEL 200EU / T200U NOy NITROGEN OXIDES ANALYZER © TELEDYNE ADVANCED POLLUTION INSTRUMENTATION (TAPI) 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 TOLL-FREE: 800-324-5190 TEL: 858-657-9800 FAX: 858-657-9816 E-MAIL: api-customerservice@teledyne.com WEB SITE: www.teledyne-api.com 06041 Rev. B DCN 6465 08 May 2012 Information contained herein is classified as EAR99 under the U.S.
  • Page 2 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 3 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual TABLE OF CONTENTS PRINCIPLE OF OPERATION PNEUMATICS MENU STRUCTURE CALIBRATION PROCEDURE MAINTENANCE TROUBLESHOOTING & FAULTS SPECIFICATIONS MISC DIAGRAMS T SERIES ADDENDUM 06041B Page 3 of 96 05/08/12 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 4 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 5: Principle Of Operation

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 1. PRINCIPLE OF OPERATION 06041B Page 5 of 96 05/08/12 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 6 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 7: Measurement Principle

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Measurement Principle The M200EU NO Nitrogen Oxides Analyzer is a microprocessor controlled instrument that determines the concentration of nitric oxide (NO), total nitrogen oxides (NO , the sum of NO and NO...
  • Page 8 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual The system is composed of 3 modules:  The M200EU, without a moly converter  A Bypass pump chassis containing:  Bypass pump  Flow control  Sample filtration  Moly temperature control ...
  • Page 9 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual The Sample Particulate Filters The particulate filters should be inspected often for signs of plugging or excess dirt. It should be replaced regularly even without obvious signs of dirt. Because the NOy analyzer uses a 5 micron pore size, it can clog up while retaining a clean look.
  • Page 10 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual The Reaction Chamber The reaction cell of the M200EU NOy does not have critical flow orifices like all of the other chemiluminescent analyzers do. The orifices are located in different locations throughout the analyzer as specified in the critical flow orifice section of this manual.
  • Page 11: The Photo Multiplier Tube

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual The Photo Multiplier Tube The M200EU NOy uses a photo-multiplier tube (PMT) to detect the amount of light created by the NO and O reaction in the reaction cell. A PMT is typically a vacuum tube containing a variety of specially designed electrodes. Photons enter the PMT and strike a negatively charged photo cathode causing it to emit electrons.
  • Page 12 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Schematic of Basic PMT Design and Functionality A significant performance characteristic of the PMT is the voltage potential across the electron multiplier. The higher the voltage, the greater is the number of electrons emitted from each dynode of the electron multiplier, making the PMT more sensitive and responsive to small variations in light intensity but also increases random noise (dark noise).
  • Page 13 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual M200EU NOy Sensor Assembly 06041B Page 13 of 96 05/08/12 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 14 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 06041B Page 14 of 96 05/08/12 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 15: Optical Filter

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Optical Filter Another critical component in the method by which your M200EU NOy detects chemiluminescence is the optical filter that lies between the reaction cell and the PMT. This filter is a high pass filter that is only transparent to wavelengths of light above 665nm.
  • Page 16: Ozone Gas Air Flow

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Ozone Gas Air Flow The excess ozone needed for reaction with NO in the reaction cell is generated inside the analyzer because of the instability and toxicity of ozone. Besides the ozone generator itself, this requires a dry air supply and filtering of the gas before it is introduced into the reaction cell.
  • Page 17 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Generator The M200EU NOy uses a corona discharge (CD) tube for creating its O . Corona discharge generation is capable of producing high concentrations of ozone efficiently and with low excess heat. Although there are many cell designs, the fundamental principle remains the same.
  • Page 18 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual ® Perma Pure Dryer The air supplied to the O generation system needs to be as dry as possible. Normal room air contains a certain amount of water vapor, which greatly diminishes the yield of ozone produced by the ozone generator.
  • Page 19 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual ® M200EU NOy Perma Pure Dryer Just like on startup, if the instrument is turned on after having been off for more than 30 minutes, it takes ® a certain amount of time for the humidity gradient to become large enough for the Perma Pure Dryer to adequately dry the air.
  • Page 20 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Critical Flow Orifices In order to maintain constant flow rates for both the O supply air and the sample gas, the M200EU NOy uses a variety of critical-orifice flow control assemblies located at these places in the pneumatic system of the instrument: ...
  • Page 21 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual M200EU NOy Gas Flow Rates Flowrate Location Purpose Orifice Diameter (nominal) On port 2 of Pre- Controls rate of flow of sample gas 0.014” (14mil) 1000 cm³/min reactor valve into the reaction cell.
  • Page 22 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 23 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 2. PNEUMATICS 06041B Page 23 of 96 05/08/12 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 24 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 25 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 06041B Page 25 of 96 05/08/12 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 26: User Notes

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual USER NOTES: 06041B Page 26 of 96 05/08/12 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 27: Menu Structure

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 3. MENU STRUCTURE 06041B Page 27 of 96 05/08/12 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 28 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 29 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual SAMPLE TEST SETUP Primary Setup Menu) Only appear if reporting range HIGH <TST TST> is set for AUTO range mode. RANG PASS MORE (Secondary Setup Menu) ZERO SPAN CONC RANGE NOx STB...
  • Page 30 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual SAMPLE CALZ SETUP TEST CALS Only appear if reporting range HIGH HIGH HIGH <TST TST> is set for AUTO range mode. RANGE NOX STB ZERO SPAN CONC ZERO SPAN CONC SAMP FLOW...
  • Page 31 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual SAMPLE RNGE PASS MORE ACAL PREV NEXT Go To iDAS MENU TREE (Fig. A-4) PREV NEXT MODE MODEL TIME DATE DESIGNATION SEQ 1) SERIAL NUMBER SEQ 2) SOFTWARE REVISION SEQ 3 MODE...
  • Page 32 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual SAMPLE RNGE PASS MORE ACAL VIEW EDIT Password required PREV NEXT CONC01 PREV NEXT EDIT PRNT ENGDATA CALDATA CONC60 CONC01 VIEW ENGDATA CALDATA <SET SET> EDIT PRNT CONC60 <PRM PRM> PV10 PREV...
  • Page 33 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual SAMPLE RNGE PASS MORE ACAL COMM VARS DIAG Password required Password required INET COM1 COM2 GTWY SNET START STOP PREV NEXT JUMP EDIT PRINT DAS_HOLD_OFF <SET SET> EDIT MEMS-MODE TPC_ENABLE DYN_ZERO OFF...
  • Page 34 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual SAMPLE RNGE PASS MORE ACAL DIAG COMM VARS Password required PREV NEXT SIGNAL ANALOG ANALOG I/O TEST OPTIC ELECTRIC OZONE GENERATOR FLOW OUTPUT CONFIGURATION CHANNEL TEST TEST OVERRIDE CALIBRATION OUTPUT ENTR ENTR...
  • Page 35: Calibration Procedure

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 4. CALIBRATION PROCEDURE 06041B Page 35 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 36 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 37: Calibration Information

    Unlike most NOx analyzers, the M200EU with the NOy option does not have a sample inlet port on the rear panel of the 200EU. The sample port is located on the External Converter; therefore zero/span calibration is different than a normal NOx instrument.
  • Page 38 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Factory Calibration for a M200EU NOy The sensor module hardware calibration adjusts the slope of the PMT output when the Instruments slope and offset values are outside of the acceptable range and all other more obvious causes for this problem have been eliminated.
  • Page 39 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual NORM PMT value. If you are on a range 2000ppb and below you will want to use the first equation, with a range 2000.1ppb and above you will use the second equation.
  • Page 40 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Service Note 9480 Carroll Park Dr., San Diego, CA 92121-2251 Phone (858) 657-9800 Fax: (858) 657-9818 Toll Free 1800 324-5190 E-mail: api-customerservice@teledyne.com http://www.teledyne-api.com 03-020A 9 SEPTEMBER, 2003 HOW TO PERFORM A MANUAL DAC CALIBRATION ON “E” SERIES MACHINES...
  • Page 41 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 11. Press CAL on the front panel. 12. You should have some DN and UP buttons. And the top line should be say ZERO ADJUST or something similar. 13. The output on the meter should be as close as possible to 0mV ±0.3mV.
  • Page 42 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 17. That channel is now calibrated. 18. Do this for all remaining channels that contain the Current option and ensure that you move the meter on the output connector to the proper pins for that channel.
  • Page 43: Pressure Calibration

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Pressure Calibration To calibrate the pressure in the analyzer the first thing you will want to do is disconnect the pump. You can either do this pneumatically or electrically. Also disconnect any tubing connected to the sample inlet port on the back of the analyzer.
  • Page 44 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 45: Maintenance

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 5. MAINTENANCE 06041B Page 45 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 46 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 47: Instrument Maintenance

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual INSTRUMENT MAINTENANCE Predictive diagnostic functions including data acquisition, failure warnings and alarms built into the analyzer allow the user to determine when repairs are necessary without performing unnecessary, preventative maintenance procedures. There is, however, a minimal number of simple procedures that, when performed regularly, will ensure that the analyzer continues to operate accurately and reliably over its lifetime.
  • Page 48: Maintenance Schedule

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual MAINTENANCE SCHEDULE The table below is the recommended maintenance schedule for the M200EU NOy. Please note that in certain environments with high levels of dust, humidity or pollutant levels some maintenance procedures may need to be performed more often than shown.
  • Page 49: Troubleshooting & Faults

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 6. TROUBLESHOOTING & FAULTS 06041B Page 49 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 50 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 51: Predictive Diagnostics

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Predictive Diagnostics The analyzer’s test functions can be used to predict failures by looking at trends in their values. Table below can be used as a basis for taking action as these values change with time. The internal data acquisition system (iDAS) is a convenient way to record and track these changes.
  • Page 52: Non-Linear Response

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Non-Linear Response The M200EU NOy was factory calibrated to a high level of NO and should be linear to within 1% of full scale. Common causes for non-linearity are:  Leaks in the pneumatic system. Leaks can add a constant of ambient air, zero air or span gas to the current sample gas stream, which may be changing in concentrations as the linearity test is performed.
  • Page 53 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual M200EU NOy Troubleshooting Tree Troubleshooting Trees No Power No front panel or locking up Display Optical Test bad, Electrical Test OK Bad Electrical Test Unstable Reading at Zero, Zero noise Unstable Reading at Span, Span noise...
  • Page 54 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual NO POWER Incorrect voltage or Frequency Verify that the correct voltage and powering the analyzer frequency are going to the analyzer Unit plugged into a good socket Unit must be plugged into a 3 wire...
  • Page 55 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual NO FRONT PANEL OR A LOCKING UP DISPLAY Check if the DC voltage is bad Verify proper DC voltages on the relay board test points. Something bringing down I Follow service note 03-028, “Troubleshooting the I...
  • Page 56 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual OPTICAL TEST BAD, ELECTRICAL TEST OK Bad HVPS Perform an HVPS test, measuring across each pin of the HVPS Check optical test LED. Pull the LED Bad Optical LED out of the cold block and verify...
  • Page 57 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual BAD ELECTRICAL TEST Bad Preamp Replace Preamp Bad Cabling Replace cables going from preamp to motherboard Bad Motherboard Replace Motherboard (5 Year warranty part) 06041B Page 57 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 58 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual UNSTABLE READING AT ZERO, ZERO NOISE Perform a Leak check on the analyzer. Cap the Leak in the analyzer sample inlet and the drier inlet and make sure the sample pressure drops <10” and is equal to the Rcell pressure to within 0.2”...
  • Page 59 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual UNSTABLE READING AT SPAN, SPAN NOISE Perform a Leak check on the analyzer. Cap the Leak in the analyzer sample inlet and the drier inlet and make sure the sample pressure drops <10” and is equal to the Rcell pressure to within 0.2”...
  • Page 60 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual UNABLE TO ZERO (NO ZERO BUTTON) Perform a Leak check on the analyzer. Cap the Leak in the analyzer sample inlet and the drier inlet and make sure the sample pressure drops <10” and is equal to the Rcell pressure to within 0.2”...
  • Page 61 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual UNABLE TO SPAN (NO SPAN BUTTON OR NO RESPONSE TO SPAN) Perform a Leak check on the analyzer. Cap the Leak in the analyzer sample inlet and the drier inlet and make sure the sample pressure drops <10”...
  • Page 62 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual NON-LINEAR RESPONSE Perform a Leak check on the analyzer. Cap the Leak in the analyzer sample inlet and the drier inlet and make sure the sample pressure drops <10” and is equal to the Rcell pressure to within 0.2”...
  • Page 63 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual SLOW RESPONSE TO ZERO OR SPAN Perform a Leak check on the analyzer. Cap the Leak in the anal yzer sample inlet and the drier inlet and make sure the sample pressure drops <10” and is equal to the Rcell pressure to within 0.2”...
  • Page 64 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual HVPS WARNING AFTER FACTORY CALIBRATION Perform a Leak check on the analyzer. Cap the Leak in the analyzer sample inlet and the drier inlet and make sure the sample pressure drops <10” and is equal to the Rcell pressure to within 0.2”...
  • Page 65 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual AUTO-ZERO WARNING Record the A-Zero reading after 15minutes on span Auto-Zero Valve bad and then 15minutes on zero. It should be the same. If it is more then 2mV difference, replace the valve.
  • Page 66 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual NO FLOW Front panel particulate filter dirty Replace front panel particulate filter Stainl ess steel sintered filter Replace the sintered filter in front of the critical flow dirty orifices, located underneath the fittings on top of the Reaction Chamber.
  • Page 67 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual NO ANALOG OR INCORRECT ANALOG OUTPUT Outputs not properly Perform an analog output calibration. Refer to calibrated service note 03-020 Look at the motherboard on the rear panel and J19, Output set for Voltage or J21, and J23.
  • Page 68 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual ANY TEMPERATURE WARNING Toggle the heater on and off in the Signal I/O menu while measuring the AC voltage going to the heater. Bad Relay This should switch between 0V and line voltage. If the LED is turning on and off but the voltage isn’t...
  • Page 69 Model 200EU NOy CUSTOMER: __________________________________ PHONE: ___________________________________________ CONTACTNAME: ______________________________ FAX NO. ___________________________________________ SITE ADDRESS: __________________________________________________________________________________ MODEL 200EU NOy SERIAL NO.:________________________FIRMWARE REVISION: ________________________ 1. ARE THERE ANY FAILURE MESSAGES? __________________________________________________________ _________________________________________________________________________________________________ _________________________________________________________________________________________________ PLEASE COMPLETE THE FOLLOWING TABLE: (NOTE: DEPENDING ON OPTIONS INSTALLED, NOT ALL TEST...
  • Page 70 5. IF POSSIBLE, PLEASE INCLUDE A PORTION OF A STRIP CHART PERTAINING TO THE PROBLEM. CIRCLE PERTINENT DATA. 6. THANK YOU FOR PROVIDING THIS INFORMATION. YOUR ASSISTANCE ENABLES TELEDYNE API TO RESPOND FASTER TO THE PROBLEM THAT YOU ARE ENCOUNTERING.
  • Page 71: User Notes

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual USER NOTES: 06041B Page 71 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 72 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 73: Specifications

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 7. SPECIFICATIONS 06041B Page 73 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 74 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 75 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Specifications Model 200E Basic Unit Specifications Min/Max Range Min: 0-50 ppb (Physical Analog Output) Max: 0-20 ppm Measurement Units ppb, ppm, µg/m , mg/m (user selectable) Zero Noise 0.2 ppb (RMS) Span Noise <0.5% of reading above 50 ppb or 0.2 ppm, whichever is greater...
  • Page 76 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual M200EU NOy converter option specifications Converter Remotely mounted molybdenum converter with temperature controller in a NEMA – 4 enclosure. Converter Temperature C  7 C with read-out on front panel of bypass pump chassis.
  • Page 77 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 8. MISC DIAGRAMS 06041B Page 77 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 78 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual M200EU NOy Electronic Block Diagram 06041B Page 78 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 79 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual USER NOTES 06041B Page 79 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 80 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual USER NOTES 06041B Page 80 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 81 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual USER NOTES 06041B Page 81 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 82 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual 9. T SERIES ADDENDUM 06041B Page 82 of 96 6/27/07 Information contained herein is classified as EAR99 under the U.S. Export Administration Regulations. Export, reexport or diversion contrary to U.S. law is prohibited.
  • Page 83: Getting Started

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Front panel, rear panel, and display Getting Started This section introduces you to the instrument components of the front and rear panel, which are unique to the T series analyzers. Front Panel Figure10-1 shows the analyzer’s front panel layout, followed by a close-up of the display screen in...
  • Page 84 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Figure 10-2: Display Screen and Touch Control The front panel liquid crystal display screen includes touch control. Upon analyzer start-up, the screen shows a splash screen and other initialization indicators before the main display appears, similar to Figure 10-2 above (may or may not display a Fault alarm).
  • Page 85 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Table 10-1: Display Screen and Touch Control Description Field Description/Function Status Lights indicating the states of Sample, Calibration and Fault, as follows: Name Color State Definition Unit is not operating in sample mode, DAS is disabled.
  • Page 86: Front Panel/Display Interface

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Figure 10-3: Display/Touch Control Screen Mapped to Menu Charts Front Panel/Display Interface Users can input data and receive information directly through the front panel touch-screen display. The LCD display is controlled directly by the CPU board. The touchscreen is interfaced to the CPU by means of a touchscreen controller that connects to the CPU via the internal USB bus and emulates a computer mouse.
  • Page 87: Lvds Transmitter Board

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual LCD Display Touchscreen Back-Light Supply TFT BIAS Supply 10.4, -7.0, 16, 4V 3.3V LVDS Touch Screen Controller LVDS Transmitter 18 Bit TTL Data Receiver Board Remote Local USB & 5V Utility COM4 USB4 Lang.
  • Page 88: Rear Panel

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Rear panel Figure 9-5: Rear Panel Layout Table 10-2 provides a description of new components on the rear panel. Table 10-2: Rear Panel Description Component Function Option for external voltage signals from other instrumentation and for logging these...
  • Page 89 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Figure 10-6: Analog In Connector Pin assignments for the Analog In connector are presented in Table 10-3. Table 10-3: Analog Input Pin Assignments DESCRIPTION PARAMETER Analog input # 1 AIN 1 Analog input # 2...
  • Page 90: Analog Input Calibration

    A Teledyne logo will appear and flash, wait until a logo appears again with the words System Booting and a loading bar appear below the logo, and hold down the left shift and left control key on the keyboard throughout the rest of the boot up.
  • Page 91: Ain Calibration

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual AIN Calibration This is the sub-menu to conduct the analog input calibration. This calibration should only be necessary after major repair such as a replacement of CPU, motherboard or power supplies. Navigate to the ANALOG I/O CONFIGURATION MENU from the DIAG Menu, then press:...
  • Page 92: Usb Configuration

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual To adjust settings for the Analog Input option parameters press: DIAG ANALOG I / O CONFIGURATION PREV NEXT ENTR EXIT DIAG AIO AOUTS CALIBRATED: NO Press SET> to scroll to the first channel.
  • Page 93 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Firmware Updates via USB The T series analyzers can receive firmware updates using a flash drive and the USB ports on the front panel. To update the firmware, locate the file you want to use for the update, and rename it to “update.exe”...
  • Page 94: Troubleshooting Faults

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Troubleshooting faults Touch-screen Interface Verify the functioning of the touch screen by observing the display when pressing a touch-screen control button. Assuming that there are no wiring problems and that the DC power supplies are...
  • Page 95: Diagrams And Schematics

    MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual Diagrams and schematics RS232 COM2 USB COM Ethernet Male Female port Analog Outputs COM1 (RS–232 ONLY) Optional Analog In 4-20 mA Control Inputs: Touchscreen 1 – 6 Display Status Outputs: 1 – 8...
  • Page 96 MODEL 200EU NOy NITROGEN OXIDES ANALYZER Training Manual “E” series compatibility Incompatible components The following components are not compatible between E series and T series analyzers: Multidrop Display and Keyboard components Ethernet Analog Inputs 06041B Page 96 of 96 6/27/07...