Tornado 1 / Tornado 2
EN
Installation and operating instructions
0297

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Tornado 1 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Durr Dental Tornado 1

  • Page 1 Tornado 1 / Tornado 2 Installation and operating instructions 0297...
  • Page 2 The current version of the installation and operating instructions is available in the Download Center: http://qr.duerrdental.com/9000-610-60 © DÜRR DENTAL SE...
  • Page 3: Table Of Contents

    Check the motor protection Tornado 1 ....switch .....
  • Page 4 Contents Usage 12 Operation ..... . . 12.1 Switching the unit on/off ..13 Maintenance ..... 13.1 Maintenance schedule .
  • Page 5: Important Information

    – NOTICE These installation and operating instructions Risk of extensive material/property damage apply to: Tornado 1 Other symbols REF: 5180-01; 5180-02; 5180-03; 5185-01; These symbols are used in the document and on 5182-01; 5182-02; 5182-03; 5186-01...
  • Page 6: Copyright Information

    Important information Safety Conformity mark for the United Kingdom of Great Britain and Northern Ireland, with The unit has been developed and designed in the number of the designated authority such a way that dangers are effectively ruled out if used in accordance with the Intended Use. Authorised representative for Switzerland Despite this, the following residual risks can remain:...
  • Page 7: General Safety Information

    Important information Electrical safety WARNING Observe and comply with all the relevant elec- ❯ Risk of explosion due to ignition of trical safety regulations when working on the combustible materials unit. Do not operate the unit in any rooms in ❯...
  • Page 8: Disposal

    Important information 2.10 Disposal The unit must be disposed of properly. Within the European Union, the unit must be disposed of in accordance with EU Directive 2012/19/EU (WEEE). If you have any questions about the correct ❯ disposal of parts, please contact your dental trade supplier.
  • Page 9: Product Description

    – Appliance log book cific requirements and/or import regulations): Tornado 1, 230 V ....5180-01 – Collector tray Tornado 1, 100-110 V, 50 Hz / Optional items 100-127 V, 60 Hz .
  • Page 10: Technical Data

    Product description Technical data Tornado 1 Electrical data 5180-01 5185-01 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1...
  • Page 11 Product description Tornado 1 Electrical data 5180-02 Rated voltage 100 - 110 100 - 127 Mains frequency Nominal current at 8 bar (0.8 MPa) 8.5 - 8.9 9.0 - 7.9 Motor protection switch, recommended settings 8.5 - 10 9.0 - 9.0...
  • Page 12 Product description Tornado 1 Electrical data 5180-03 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1...
  • Page 13 Product description Tornado 1 with membrane drying unit Electrical data 5182-01 5186-01 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1...
  • Page 14 Product description Tornado 1 with membrane drying unit Electrical data 5182-02 Rated voltage 100 - 110 100 - 127 Mains frequency Nominal current at 8 bar (0.8 MPa) 8.6 - 9.0 9.1 - 8.0 Motor protection switch, recommended settings 8.6 - 10 9.1 - 9.1...
  • Page 15: Ing Unit

    Product description Tornado 1 with membrane drying unit Electrical data 5182-03 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1...
  • Page 16: Tornado 2

    Product description Tornado 2 Electrical data 5280-01 5285-01 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Max. permissible mains impedance in accordance with EN 61000‑3‑11 (0.276 + j0.172) Circuit breaker fuse characteristics B, C or D in acc.
  • Page 17: Tornado

    Product description Tornado 2 Electrical data 5280-03 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Max. permissible mains impedance in accordance with EN 61000‑3‑11 (0.276 + j0.172) Circuit breaker fuse characteristics B, C or D in acc.
  • Page 18: Tornado 2 With Membrane Dry

    Product description Tornado 2 with membrane drying unit Electrical data 5282-01 5286-01 5282100029 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Max. permissible mains impedance in accordance with EN 61000‑3‑11 (0.276 + j0.172) Circuit breaker fuse characteristics B, C or D in acc.
  • Page 19: Ing Unit

    Product description 4.10 Tornado 2 with membrane drying unit Electrical data 5282-03 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Max. permissible mains impedance in accordance with EN 61000‑3‑11 (0.276 + j0.172) Circuit breaker fuse characteristics B, C or D in acc.
  • Page 20: Filter

    Product description 4.11 Filter Filter mesh size Air intake filter µm Fine filter µm Virus bacteria filter µm 0.01 Coalescence filter µm 0.01 4.12 Ambient conditions Ambient conditions during storage and transport Temperature °C -10 - +55 Relative humidity < 95 Ambient conditions during operation Temperature °C...
  • Page 21: Distance Between Rubber Feet

    (cm) (cm) 20 l Type plate for the complete system With noise reducing hood Compressor unit Tornado 1 / 2 The type plate of the compressor unit is located on the crankcase below the cylinder. Compressor unit type plate (cm)
  • Page 22: Evaluation Of Conformity

    Product description Membrane drying unit The type plate of the membrane drying unit is located on the membrane drying unit. Membrane drying unit type plate 4.15 Evaluation of conformity This device has been subjected to conformity acceptance testing in accordance with the cur- rent relevant European Union guidelines.
  • Page 23: Operation

    Product description Operation Unit with membrane-drying unit Unit without membrane-dry- ing unit Compressor unit Air intake filter Compressor unit Carry handles Air intake filter On/off switch Carry handles Pressure switch On/off switch Pressure gauge/display Pressure switch Pressure vessel Pressure gauge/display Compressed air connection (quick release Pressure vessel coupling)
  • Page 24: Assembly

    Assembly Assembly ≥ 40 °C Requirements The unit must not be set up or operated within the vicinity of the patients (within a radius of 1.5 m). The unit can be installed either at the same level as the surgery room or on a floor below (e.g. cel- lar).
  • Page 25: Transport

    Assembly Transport Installation Remove the transport locks WARNING For safe transport, the unit is secured with foam Risk of explosion of the pressure tank padding blocks and a transport strap. and pressure hoses Cut and remove the transport strap. ❯ The pressure tank and the pressure ❯...
  • Page 26: Place A Collector Tray Underneath

    Assembly Electrical connections Safety when making electrical connections The unit has no main power switch. For this reason it is important that the unit is be set up in such a way that the plug can be easily accessed and unplugged if required.
  • Page 27: Commissioning

    Assembly Commissioning Secure the mains connector using the cable tie ❯ provided. Before taking the unit into operation, check for ❯ any damage. Damaged units must not be used. In many countries technical medical prod- ucts and electrical devices are subject to regular checks at set intervals.
  • Page 28: Checking The Switch-On/Cut-Off Pressure

    Assembly Measure the maximum current consumption At maximum tank pressure, slowly open the ❯ ❯ (this is the value just before the cut-off pressure condensate drain valve. is reached). Close the condensate drain valve as soon as ❯ If the reading deviates from the recommended all of the accumulated condensation water has setting, then the motor protection switch needs been blown out.
  • Page 29: 10 Adjustment Options

    Assembly 10 Adjustment options Adjust the start-up pressure via the pressure ❯ difference ∆p at the adjustment screw. The pressure difference increases in the "+" 10.1 Adjusting the pressure switch arrow direction and decreases in the "-" arrow direction. WARNING The maximum permissible pressure difference Risk of explosion of the pressure ves- must not be set to more than 3 bar.
  • Page 30: Adjusting The Motor Protection Switch

    Assembly 10.2 Adjusting the motor protec- tion switch NOTICE Danger of overheating if the motor protection switch is set too high If the motor protection switch is set too high, motor damage can occur as a result of overheating. Set the motor protection switch prop- ❯...
  • Page 31: 11 Circuit Diagrams

    Assembly 11 Circuit diagrams 11.1 230 V units Units without a membrane-drying unit P> Condenser Compressor motor Fan motor, noise reduction hood (if required) Pressure switch Mains connection 1/N/PE AC 230 V Compressor motor plug connection Plug connection for fan motor, noise reduction hood (if required) 9000-610-60/02 2305V010...
  • Page 32 Assembly Unit with membrane-drying unit P> Condenser Compressor motor Cooling fan motor, membrane drying unit Fan motor, noise reduction hood (if required) Pressure switch Mains connection 1/N/PE AC 230 V Plug connection for fan motor, cooling, membrane drying unit Plug connection for compressor motor and cooling, membrane drying unit Plug connection for fan motor, noise reduction hood (if required) 9000-610-60/02 2305V010...
  • Page 33: Devices With 110 - 127 V

    Assembly 11.2 Devices with 110 - 127 V Units without a membrane-drying unit P> I > I > I > Condenser Compressor motor Fan motor, noise reduction hood (if required) Pressure switch Mains connection 1/N/PE AC 110 - 127 V / 230 V Compressor motor plug connection Plug connection for fan motor, noise reduction hood (if required) 9000-610-60/02...
  • Page 34 Assembly Unit with membrane-drying unit P> I > I > I > Condenser Compressor motor Cooling fan motor, membrane drying unit Fan motor, noise reduction hood (if required) Pressure switch Mains connection 1/N/PE AC 110 - 127 V / 230 V Plug connection for fan motor, cooling, membrane drying unit Plug connection for compressor motor and cooling, membrane drying unit Plug connection for fan motor, noise reduction hood (if required)
  • Page 35: 12 Operation

    Usage Usage 12 Operation Prior to working on the unit or in case of danger, disconnect it from the mains. 12.1 Switching the unit on/off Switch on the unit at the pressure switch by ❯ turning it to the position "I AUTO". The compressor unit will run automatically and the pressure tank will be filled.
  • Page 36: 13 Maintenance

    Usage 13 Maintenance Prior to working on the unit or in case of danger, disconnect it from the mains. WARNING Risk of infection due to burst filters Particles enter the compressed air network and can therefore enter the mouth of the patient. Replace filters in accordance with the maintenance schedule.
  • Page 37: Wear Parts And Replacement Parts

    Usage Unit with membrane-drying unit Maintenance Maintenance work interval At regular intervals Empty the collector tray under the membrane drying unit (the interval may vary ❯ depending on the ambient conditions and method of working; empty it daily if the humidity is high). Annually Replace the air intake filter in the compressor unit –...
  • Page 38: Replacing The Air Intake Filter

    Usage Release the filter cover by rotating it anti- ❯ 13.3 Replacing the air intake filter clockwise and then take it off. Units without a noise reduction hood Switch off the compressor at the pressure ❯ switch. Disconnect all power from the device. ❯...
  • Page 39: Replacing The Filter Of The Membrane Drying Unit

    Usage 13.4 Replacing the filter of the 13.5 Checking the safety valve membrane drying unit The functioning of the safety valve must be checked at regular intervals in accordance with Fine or virus bacteria filter national regulations. Unscrew and remove the filter cover. ❯...
  • Page 40: 14 Taking Out Of Use

    Usage 14 Taking out of use 14.1 Taking the unit out of use If the unit is not to be used for a prolonged period of time, we recommend that it is properly shut down and taken out of use. To do this, the accumulated condensation water from the unit must be drained.
  • Page 41: Troubleshooting

    Troubleshooting Troubleshooting 15 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. Prior to working on the unit or in case of danger, disconnect it from the mains. Error Possible cause Remedy...
  • Page 42 Troubleshooting Error Possible cause Remedy Reduced delivery. Compres- Air intake filter dirty Replace the air intake filter at ❯ least 1x per year. The air sor needs longer to charge the pressure tank, see charg- intake filter must never be ing times in "4 Technical data"...
  • Page 43: Appendix

    Appendix Appendix 16 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)
  • Page 44 Appendix 9000-610-60/02 2305V010...
  • Page 45: 17 Country Representatives

    Appendix 17 Country representatives Country Address UK Responsible Person: Duerr Dental (Products) UK Ltd. 14 Linnell Way Telford Way Industrial Estate Kettering, Northants NN 16 8PS Уповноважений представник в Україні: Приватне підприємство “Галіт” вул. 15 квітня, 6Є, с. Байківці, Тернопільський р-н, 47711, Україна...
  • Page 48 Hersteller / Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...