– NOTICE These installation and operating instructions Risk of extensive material/property damage apply to: Tornado 1 Other symbols REF: 5180-01; 5180-02; 5180-03; 5185-01; These symbols are used in the document and on 5182-01; 5182-02; 5182-03; 5186-01...
Important information Safety Conformity mark for the United Kingdom of Great Britain and Northern Ireland, with The unit has been developed and designed in the number of the designated authority such a way that dangers are effectively ruled out if used in accordance with the Intended Use. Authorised representative for Switzerland Despite this, the following residual risks can remain:...
Important information Electrical safety WARNING Observe and comply with all the relevant elec- ❯ Risk of explosion due to ignition of trical safety regulations when working on the combustible materials unit. Do not operate the unit in any rooms in ❯...
Important information 2.10 Disposal The unit must be disposed of properly. Within the European Union, the unit must be disposed of in accordance with EU Directive 2012/19/EU (WEEE). If you have any questions about the correct ❯ disposal of parts, please contact your dental trade supplier.
Product description Technical data Tornado 1 Electrical data 5180-01 5185-01 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1...
Page 11
Product description Tornado 1 Electrical data 5180-02 Rated voltage 100 - 110 100 - 127 Mains frequency Nominal current at 8 bar (0.8 MPa) 8.5 - 8.9 9.0 - 7.9 Motor protection switch, recommended settings 8.5 - 10 9.0 - 9.0...
Page 12
Product description Tornado 1 Electrical data 5180-03 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1...
Page 13
Product description Tornado 1 with membrane drying unit Electrical data 5182-01 5186-01 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1...
Page 14
Product description Tornado 1 with membrane drying unit Electrical data 5182-02 Rated voltage 100 - 110 100 - 127 Mains frequency Nominal current at 8 bar (0.8 MPa) 8.6 - 9.0 9.1 - 8.0 Motor protection switch, recommended settings 8.6 - 10 9.1 - 9.1...
Product description Tornado 1 with membrane drying unit Electrical data 5182-03 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Circuit breaker fuse characteristics B, C or D in acc. with EN 60898-1...
Product description Tornado 2 Electrical data 5280-01 5285-01 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Max. permissible mains impedance in accordance with EN 61000‑3‑11 (0.276 + j0.172) Circuit breaker fuse characteristics B, C or D in acc.
Product description Tornado 2 Electrical data 5280-03 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Max. permissible mains impedance in accordance with EN 61000‑3‑11 (0.276 + j0.172) Circuit breaker fuse characteristics B, C or D in acc.
Product description Tornado 2 with membrane drying unit Electrical data 5282-01 5286-01 5282100029 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Max. permissible mains impedance in accordance with EN 61000‑3‑11 (0.276 + j0.172) Circuit breaker fuse characteristics B, C or D in acc.
Product description 4.10 Tornado 2 with membrane drying unit Electrical data 5282-03 Rated voltage Mains frequency Nominal current at 8 bar (0.8 MPa) Motor protection Motor winding overheat protector Type of protection IP 24 Mains fuses * Max. permissible mains impedance in accordance with EN 61000‑3‑11 (0.276 + j0.172) Circuit breaker fuse characteristics B, C or D in acc.
(cm) (cm) 20 l Type plate for the complete system With noise reducing hood Compressor unit Tornado 1 / 2 The type plate of the compressor unit is located on the crankcase below the cylinder. Compressor unit type plate (cm)
Product description Membrane drying unit The type plate of the membrane drying unit is located on the membrane drying unit. Membrane drying unit type plate 4.15 Evaluation of conformity This device has been subjected to conformity acceptance testing in accordance with the cur- rent relevant European Union guidelines.
Product description Operation Unit with membrane-drying unit Unit without membrane-dry- ing unit Compressor unit Air intake filter Compressor unit Carry handles Air intake filter On/off switch Carry handles Pressure switch On/off switch Pressure gauge/display Pressure switch Pressure vessel Pressure gauge/display Compressed air connection (quick release Pressure vessel coupling)
Assembly Assembly ≥ 40 °C Requirements The unit must not be set up or operated within the vicinity of the patients (within a radius of 1.5 m). The unit can be installed either at the same level as the surgery room or on a floor below (e.g. cel- lar).
Assembly Transport Installation Remove the transport locks WARNING For safe transport, the unit is secured with foam Risk of explosion of the pressure tank padding blocks and a transport strap. and pressure hoses Cut and remove the transport strap. ❯ The pressure tank and the pressure ❯...
Assembly Electrical connections Safety when making electrical connections The unit has no main power switch. For this reason it is important that the unit is be set up in such a way that the plug can be easily accessed and unplugged if required.
Assembly Commissioning Secure the mains connector using the cable tie ❯ provided. Before taking the unit into operation, check for ❯ any damage. Damaged units must not be used. In many countries technical medical prod- ucts and electrical devices are subject to regular checks at set intervals.
Assembly Measure the maximum current consumption At maximum tank pressure, slowly open the ❯ ❯ (this is the value just before the cut-off pressure condensate drain valve. is reached). Close the condensate drain valve as soon as ❯ If the reading deviates from the recommended all of the accumulated condensation water has setting, then the motor protection switch needs been blown out.
Assembly 10 Adjustment options Adjust the start-up pressure via the pressure ❯ difference ∆p at the adjustment screw. The pressure difference increases in the "+" 10.1 Adjusting the pressure switch arrow direction and decreases in the "-" arrow direction. WARNING The maximum permissible pressure difference Risk of explosion of the pressure ves- must not be set to more than 3 bar.
Assembly 10.2 Adjusting the motor protec- tion switch NOTICE Danger of overheating if the motor protection switch is set too high If the motor protection switch is set too high, motor damage can occur as a result of overheating. Set the motor protection switch prop- ❯...
Assembly 11 Circuit diagrams 11.1 230 V units Units without a membrane-drying unit P> Condenser Compressor motor Fan motor, noise reduction hood (if required) Pressure switch Mains connection 1/N/PE AC 230 V Compressor motor plug connection Plug connection for fan motor, noise reduction hood (if required) 9000-610-60/02 2305V010...
Page 32
Assembly Unit with membrane-drying unit P> Condenser Compressor motor Cooling fan motor, membrane drying unit Fan motor, noise reduction hood (if required) Pressure switch Mains connection 1/N/PE AC 230 V Plug connection for fan motor, cooling, membrane drying unit Plug connection for compressor motor and cooling, membrane drying unit Plug connection for fan motor, noise reduction hood (if required) 9000-610-60/02 2305V010...
Assembly 11.2 Devices with 110 - 127 V Units without a membrane-drying unit P> I > I > I > Condenser Compressor motor Fan motor, noise reduction hood (if required) Pressure switch Mains connection 1/N/PE AC 110 - 127 V / 230 V Compressor motor plug connection Plug connection for fan motor, noise reduction hood (if required) 9000-610-60/02...
Page 34
Assembly Unit with membrane-drying unit P> I > I > I > Condenser Compressor motor Cooling fan motor, membrane drying unit Fan motor, noise reduction hood (if required) Pressure switch Mains connection 1/N/PE AC 110 - 127 V / 230 V Plug connection for fan motor, cooling, membrane drying unit Plug connection for compressor motor and cooling, membrane drying unit Plug connection for fan motor, noise reduction hood (if required)
Usage Usage 12 Operation Prior to working on the unit or in case of danger, disconnect it from the mains. 12.1 Switching the unit on/off Switch on the unit at the pressure switch by ❯ turning it to the position "I AUTO". The compressor unit will run automatically and the pressure tank will be filled.
Usage 13 Maintenance Prior to working on the unit or in case of danger, disconnect it from the mains. WARNING Risk of infection due to burst filters Particles enter the compressed air network and can therefore enter the mouth of the patient. Replace filters in accordance with the maintenance schedule.
Usage Unit with membrane-drying unit Maintenance Maintenance work interval At regular intervals Empty the collector tray under the membrane drying unit (the interval may vary ❯ depending on the ambient conditions and method of working; empty it daily if the humidity is high). Annually Replace the air intake filter in the compressor unit –...
Usage Release the filter cover by rotating it anti- ❯ 13.3 Replacing the air intake filter clockwise and then take it off. Units without a noise reduction hood Switch off the compressor at the pressure ❯ switch. Disconnect all power from the device. ❯...
Usage 13.4 Replacing the filter of the 13.5 Checking the safety valve membrane drying unit The functioning of the safety valve must be checked at regular intervals in accordance with Fine or virus bacteria filter national regulations. Unscrew and remove the filter cover. ❯...
Usage 14 Taking out of use 14.1 Taking the unit out of use If the unit is not to be used for a prolonged period of time, we recommend that it is properly shut down and taken out of use. To do this, the accumulated condensation water from the unit must be drained.
Troubleshooting Troubleshooting 15 Tips for operators and service technicians Any repairs exceeding routine maintenance may only be carried out by qualified personnel or our service. Prior to working on the unit or in case of danger, disconnect it from the mains. Error Possible cause Remedy...
Page 42
Troubleshooting Error Possible cause Remedy Reduced delivery. Compres- Air intake filter dirty Replace the air intake filter at ❯ least 1x per year. The air sor needs longer to charge the pressure tank, see charg- intake filter must never be ing times in "4 Technical data"...
Appendix Appendix 16 Handover record This document confirms that a qualified handover of the medical device has taken place and that appropriate instructions have been provided for it. This must be carried out by a qualified adviser for the medical device, who will instruct you in the proper handling and operation of the medical device. Product name Order number (REF) Serial number (SN)
Appendix 17 Country representatives Country Address UK Responsible Person: Duerr Dental (Products) UK Ltd. 14 Linnell Way Telford Way Industrial Estate Kettering, Northants NN 16 8PS Уповноважений представник в Україні: Приватне підприємство “Галіт” вул. 15 квітня, 6Є, с. Байківці, Тернопільський р-н, 47711, Україна...
Page 48
Hersteller / Manufacturer: DÜRR DENTAL SE Höpfigheimer Str. 17 74321 Bietigheim-Bissingen Germany Fon: +49 7142 705-0 www.duerrdental.com info@duerrdental.com...