SECTION
4
MAINTENANCE
INTRODUCTION
This
section
provides
instructions for
testing, calibra-
ting,
troubleshooting,
and
repairing
the
Model
130.
The
instructions are
concise
and
for
the experienced
electronics technician or
field
engineer.
Wavetek
main-
tains a factory-repair
department
for
those customers
not
possessing the necessary personnel or
test
equip-
ment
to
maintain the
instrument.
If
an instrument
is
returned
to
the factory
for calibration
or
repair, a
detailed description of
the
specific
problem
should be
attached
to
facilitate
the
turnaround
time. Test point
and adjustment
locations
are illustrated
in
the
Data Package.
RECOMMENDED
TEST
EQUIPMENT
Table
4-1
contains
a
list
of
recommended
test
equip-
ment.
Any
test
equipment
having equivalent accuracies
may
be
substituted
for
those
listed.
Table
4-1.
TEST
EQUIPMENT
Name
Required
Characteristics
Recommended
Manufacturer
Model
Oscilloscope
To
30
MHz
Tektronix
544
Plug-In
Dual
Channel
Tektronix
1A1
Plug-In
Peak
mV
measuring
capability
Tektronix
1A5
Distortion
Analyzer
To 600 kHz
Hewlett-Packard
334A
Spectrum
Analyzer
Display
To
50
MHz
Hewlett-Packard
141S
IF
Section
Hewlett-Packard
8552A
RF
Section
Hewlett-Packard
8553L
Voltmeter
M
illivolt
dc
measurement
.01
Accuracy
Counter
to
10
MHz
0.1%
of
reading
accuracy
CHECKOUT
AND
CALIBRATION
The
following paragraphs provide
complete
sequential
calibration
procedures
for
the
Model
130.
instrument
checkout procedures
are indicated
by
a
checkmark
(
v/
)
following the
procedure
title.
A
quick checkout
of the
instrument can be
performed by comparing
the
indicated
parameters with the
tolerances given
in
the
Specifications of Section
1.
NOTE
The
entire calibration
procedure
must
be
read
first
to
determine
initial
control
settings
and
test
equipment
connections
before attempting checkout.
Preliminary
Procedures
1.
Set
FREQ HZ
selector
to
the
XI
K
position.
2.
Allow
one-half
hour
for
warmup.
Removal
of
Dust Cover
To
gain access for calibration or
maintenance, proceed
as follows:
a.
Remove
power
cord.
b.
Loosen
the
two
knurled
captive
screws
on
the
rear
panel.
c.
Pull
off the
rear
panel.
d.
Remove
the
cover.
Power Supply
Regulation
1.
Connect
voltmeter
between TP1
(common)
and
TP2(-t)
on Main
Board. Adjust
R104
for
-H5
Vdc
±100 mV.
2.
Connect
voltmeter
between TP1
(common)
and
TP3(—
).
Since the
negative
supply
is
referenced to
the -H5-volt supply, the
voltmeter should
indicate
-15 Vdc
±100
mV.
Square
Wave
Amplitude
Symmetry
1.
Set function
selector to
Tj
.
2.
Connect
oscilloscope,
with
1A5
plug-in,
to coaxiai-
wire
lug
on
function
switch.
3.
Adjust
R121
until
square-wave
negative
peak
is
equal
to
amplitude
to positive
peak
±5
mV.
6
7/71