GSS7000 Signal Generator Hardware Manual
Covering Spirent's GSS7000 signal generator
C
C
OMMERCIAL IN
ONFIDENCE
Document number: DGP01430AAA
Document Issue: 2-05
Date of Issue: 21 February 2024
Document status: OFFICIAL
Published in England

Advertisement

loading
Need help?

Need help?

Do you have a question about the GSS7000 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Spirent GSS7000

  • Page 1 GSS7000 Signal Generator Hardware Manual Covering Spirent’s GSS7000 signal generator OMMERCIAL IN ONFIDENCE Document number: DGP01430AAA Document Issue: 2-05 Date of Issue: 21 February 2024 Document status: OFFICIAL Published in England...
  • Page 3: Copyright Notice

    In the folder of your installation, Spirent also have a dedicated support portal called the Customer Service Center (CSC). The support portal has a wealth of resources, including the latest software releases, online scenario generation tools, and our searchable Knowledge Base.
  • Page 4  Details of changes to the system, including any new software added and all upgrades (including Windows, new hardware, new drivers, Spirent software and so on) This will enable a swift response and improve the service to you. Document: DGP01430AAA...
  • Page 5: Table Of Contents

    GSS7000 Signal Generator Hardware Manual Contents Copyright notice............................i Contact us ..............................i 1. I ......................1-1 NTRODUCTION Document conventions........................1-1 Referenced documents ........................1-2 2. S ....................2-1 AFETY INFORMATION General safety information for your equipment..................2-1 Hazardous voltages ..........................2-2 Mains power inlet..........................2-2 Earth (ground) connection........................2-2...
  • Page 6 GSS7000 Signal Generator Hardware Manual Rear panel LED indicators ....................... 4-7 Model number and serial number ..................... 4-9 Configuring the dual-output GSS7000 signal generator..............4-9 5. U GSS7000 ....................5-1 SING THE Starting the signal generator ....................... 5-1 Controlling the signal generator......................5-1 Synchronising from external equipment .....................
  • Page 7: Introduction

    Thank you for purchasing a system from Spirent, world leaders in global navigation satellite system (GNSS) simulation. The combination of Spirent signal generator hardware and software provides a powerful and flexible tool for testing all forms of GNSS technology. With this power, there is a degree of complexity. This user manual lets you exploit the full capability of your system and provides all the information you need to use your software.
  • Page 8: Referenced Documents

    MS7000 GSS7000 Datasheet and product specification DCS0055D GSS7000 Simulator And Embedded Host Update Procedure DCS0059C GSS7000 Customer Calibration Procedure DCS0059E Replacing the Bios Battery in the embedded controller of the GSS7000 Gen 2 Simulator DCS0060B GSS7000 Channel Bank Field Upgrade Procedure Document: DGP01430AAA...
  • Page 9: Safety Information

    Do not use your equipment if it appears damaged. If the equipment does not work when you first switch on AC power, completely isolate it from the AC mains supply by removing the plug from the AC power outlet, then contact Spirent Global Services.
  • Page 10: Hazardous Voltages

    2.6 Fuse rating Spirent fits T xAH yVAC fuses to the AC power inlets. You must replace fuses with an approved fuse of the same type and rating. For users located outside the USA, you must fit a fuse approved to EN60127. For users located in the USA, you must fit a fuse approved to UL248.
  • Page 11: Ac Power Cable Assemblies

    -40 to 60°C (-40 to 140°F) 20 to 90% RH (non-condensing) Spirent recommend you always operate your equipment in a controlled environment that conforms to the parameters in the table above. Do not use your equipment outside the operating parameters in the table above (for example, in a tropical or cold environment).
  • Page 12: Ventilation

    Remove any accumulated dust using a vacuum cleaner. Spirent recommend you inspect the heat sink fitted to each signal generator card on a regular basis. Dust build-up on these heat sinks could adversely affect the operation of your signal generator, eventually leading to thermal shutdown.
  • Page 13: Dimensions

    Including feet: 198 (7.8) 2.13 Embedded Host BIOS battery The Embedded Host uses a battery to store BIOS settings. Should this battery fail, your GSS7000 signal generator will continue to function as normal. Spirent recommend replacing the BIOS battery every three years. The installed PosApp software contains a warning feature that alerts the requirement to change the battery three years after date of build, calibration or BIOS battery replacement.
  • Page 14: Disposal

    Spirent simulator products meet the requirements of the European Union (EU) Waste Electrical and Electronic Equipment (WEEE) Directive. The WEEE Directive requires Spirent to label products placed on the EU market after 13th August 2005 with this ‘wheeled bin’ symbol: This symbol indicates the product contains material that presents an environmental concern.
  • Page 15: Gss7000 Introduction

    The GSS7000 signal generator supports the following channels on an individually licensed basis:  BeiDou B1I, BeiDou B2I, BeiDou B3I, BeiDou B1C and BeiDou B2a  Galileo E1, Galileo E5 (currently, the GSS7000 signal generator is limited to 16 channels for Galileo E5)  GLONASS L1, GLONASS L2 ...
  • Page 16: Embedded Host

    USB2 and USB3 ports  Two Ethernet ports, one for general network connectivity, one for use with Spirent products PosApp running on the Embedded Host requires that the engine runs on a real-time Linux operating system, while the GUI runs on the Windows operating system. Separate Embedded Host processor cores (one each for engine and GUI)
  • Page 17: Channel Bank

    85°C. In the unlikely event of a thermal shutdown, Spirent recommends you do not restart the signal generator for at least one hour. If the problem persists, contact Spirent Global Services.
  • Page 18: Rf Combiner

    3.4 DC power rail indicators DC power rail indicators are only visible with the GSS7000 side covers removed. You must look through the slot in the side casing shown in the figure below (front panel is on the left-hand side) to see the DC power rail indicators.
  • Page 19 GSS7000 Signal Generator Hardware Manual GSS7000 Introduction Figure 3-3: DC power rail indicators DC power rail indicator Colour Description reference PECL_VCC Green - power rail on RF power rail Off - power rail off Green - power rail on RF power rail...
  • Page 20: Timer Indicators

    GSS7000 Signal Generator Hardware Manual 3.5 Timer indicators Timer indicators are only visible with the GSS7000 side covers removed. Look through the slot in the side casing shown in the figure below (front panel is on the left-hand side) to see the Timer indicators.
  • Page 21: Power Supply Unit

    GSS7000 Signal Generator Hardware Manual GSS7000 Introduction Indicator on Reference Colour Description Oscillator cover Green With no connection to the rear panel EXT REF IN port:  On - Internal reference OK  Flashing - Internal reference fail  Off - Internal reference state unknown 3.7 Power supply unit...
  • Page 22 GSS7000 Introduction GSS7000 Signal Generator Hardware Manual Document: DGP01430AAA Issue number: 2-05 Page 3-8 Spirent Communications plc...
  • Page 23: Gss7000 Overview And Installation

    4.1 Front panel GSS7000 signal generators have single, or (optionally) dual, RF Out ports. Each RF Out port is a "composite" output port, with all satellite channels available on this port, together with any signal you apply to the rear panel EXT RF IN port. Each RF out port is the normal connection point for your device under test.
  • Page 24: Front Panel Indicators

    SAFETY WARNING You must ensure any connection to a rear panel port is within the SELV limits of IEC62368. The following figure shows the rear panel of a GSS7000 signal generator. Spirent might not fit all connectors for certain configurations.
  • Page 25: Rear Panel Ports

    The following tables describe all rear panel connectors. It groups the connectors into nominal groupings as viewed when looking at the rear panel. Where a port is identified as having a terminator during testing, Spirent recommends that you keep the port terminated when not in use to maintain the system’s performance.
  • Page 26 In addition, you can set any of the above to run as:  Ungated: continuously running  Gated: only present while a scenario is running The default timer settings for a stand-alone GSS7000 signal generator are:  Timer 1: 1PPS UNGATED  Timer 2: 1PPS UNGATED ...
  • Page 27: Usb Port Assignment

    4.2.2 USB port assignment You must assign each available GSS7000 signal generator USB port to either Linux or Windows whenever you insert a USB device. The GSS7000 Embedded Host automatically recognises connection of a USB keyboard and mouse; you do not need to assign these USB devices.
  • Page 28 USB port to Linux, such as when using a USB to RS232 adapter for GBAS messages when using a GSS4150 GBAS Landing System VDB signal generator with the GSS7000 signal generator. Following the steps in this section will reset the current USB port assignments for all USB ports on the GSS7000 Embedded Host.
  • Page 29: Rear Panel Led Indicators

    In normal operation, indicators 1 to 10 for each page will be on. An indicator that is off shows there is an error. Note: In the event of a fault, Spirent recommends you take a photograph of each page. Spirent might request copies of these photographs to assist in fault-finding.
  • Page 30 GSS7000 overview and installation GSS7000 Signal Generator Hardware Manual Indicator Description Off - 3V3 DC supply not present On - 3V3 DC supply present Off - 2V5 DC supply not present On - 2V5 DC supply present Off - 1V DC supply not present...
  • Page 31: Model Number And Serial Number

    Serial number ‘7000’ followed by four digits 70009876 Always quote the model number and serial number of your GSS7000 signal generator when discussing your signal generator with Spirent. 4.3 Configuring the dual-output GSS7000 signal generator You can use the dual-output system in the following ways, depending on the system you have: ...
  • Page 32 GSS7000 overview and installation GSS7000 Signal Generator Hardware Manual To remove the loop-back cable from your dual-output signal generator, reverse the procedure. Document: DGP01430AAA Issue number: 2-05 Page 4-10 Spirent Communications plc...
  • Page 33: Using The Gss7000

    5.3 Synchronising from external equipment GSS7000 signal generators incorporate input and output signal ports that you can use in various ways to synchronise time between your signal generator and the rest of your system. This section describes how to use the 1PPS IN and TRIGGER IN rear panel ports to achieve synchronisation.
  • Page 34: 1Pps In

    Using the GSS7000 GSS7000 Signal Generator Hardware Manual You can start a scenario either by appropriate timing of the Run command (Trigger Mode: Disabled) or by selecting Delayed Trigger Mode and applying a rising edge to the TRIGGER IN rear panel port. Both cause the scenario to start on the next one-second rollover of the timer.
  • Page 35: Trig In - Immediate Mode

    GSS7000 Signal Generator Hardware Manual Using the GSS7000 Figure 5-2: Time synchronisation of RF output using 1PPS IN (fixed wrt EXT REF IN) 5.3.2 TRIG IN - Immediate mode Immediate When using trigger mode , the timing requirements for the rising edge applied to the TRIGGER IN rear panel...
  • Page 36: Using An Ip Power Switch

    The Host Controller utility provided with the PosApp software enables you to reboot your system remotely by using an IP power switch. Your GSS7000 must be running back plane software v2.06 or later to use an IP power switch. For this functionality to work, you must do the following: ...
  • Page 37: Configuring Remote Power On The Embedded Host

    5.4.1 Configuring remote power on the Embedded Host You use the String Send utility to configure the Embedded Host of your GSS7000 Signal Generator. To configure the Embedded Host for remote power control: Ensure that the Embedded Host is running.
  • Page 38: Installing The Ip Power Switch

    Using the GSS7000 GSS7000 Signal Generator Hardware Manual If the version is not v2.06 or later, you cannot use the IP power switch; contact Spirent Global Services to discuss your options. Enter the following command in the text box below the message window, and then click W...
  • Page 39: Entering The Ip Addresses Into The Host Controller Utility

    GSS7000 Signal Generator Hardware Manual Using the GSS7000 Figure 5-7: IP power switch connections after installation 5.4.3 Entering the IP addresses into the Host Controller utility You must add the IP addresses of the IP power switch and remote signal generator to the Host Controller utility before you can control the power on and off remotely.
  • Page 40 Using the GSS7000 GSS7000 Signal Generator Hardware Manual Click and then close the Host Controller utility. On the system from which you want to control the remote units, start SimGEN. Tools > 7xxx/9xxx Utilities > Host Controller Select Click In the S panel, enter the IP address from step 5.