Summary of Contents for Instrutech The Stinger CVM211GAA
Page 1
InstruTech CVM211GAA Stinger Convection Vacuum Gauge Limited Availability Used and in Excellent Condition Buy Today! https://www.artisantg.com/59806-1 A l l t r a d e m a r k s , b r a n d n a m e s , a n d b r a n d s a p p e a r i n g h e r e i n a r e t h e p r o p e r t y o f t h e i r r e s p e c t i v e o w n e r s .
Page 2
InstruTech® CVM211 Convection Vacuum Gauge Module The Stinger™ User Manual (Unit of measure in Torr / mTorr) InstruTech 1475 S. Fordham St. Longmont, CO 80503 Phone: +1-303-651-0551 Fax: +1-303-678-1754 E-mail info@instrutechinc.com www.instrutechinc.com p/n 000219-110...
1 Torr, convection-enhanced Pirani gauges increase the measuring range to just above atmosphere. The InstruTech® CVM211 Stinger™ module provides the basic signal conditioning required to turn a convection vacuum gauge into a complete measuring instrument. There are two different models of Stinger. One model provides a non-linear analog output, and one setpoint relay.
Input: 100 - 240 Vac Output: 24 Vdc @ 750 mA (18 W) with Universal European AC Plug Various AC plugs, 6 ft. cable length PS401-EU with UK AC Plug PS401-UK with China AC Plug PS401-C with Australian AC Plug PS401-SP InstruTech Page 4...
Page 9
The conventional IEC60320 AC power entry receptacle allows use with any user supplied AC mains power cord set available worldwide. Input: 100 - 240 Vac Output: 24 Vdc @ 2.5 A (60 W) Cable Length: 6 ft. InstruTech Page 5...
CVM211 Stinger Important Safety Information InstruTech has designed and tested this product to provide safe and reliable service, provided it is installed and operated within the strict safety guidelines provided in this manual. Please read and follow all warnings and instructions.
Do not use if the equipment grounding or electrical insulation has been compromised. Do not use if the unit has been dropped or the enclosure has been damaged. Contact InstruTech for return authorization and instructions for returning the product to InstruTech for evaluation.
WARNING! Do not attempt to use with gases other than nitrogen (N ) or air without referring to correction factor data tables. InstruTech gauges and modules are calibrated for direct readout of nitrogen or air. Do not attempt to use with other gases such as argon (Ar) or carbon dioxide (CO ) unless accurate conversion data for to other gas is properly used.
A good recommended practice is to remove power from any cable prior to connecting or disconnecting it. The InstruTech CVM211 will directly replace Granville-Phillips® Mini-Convectron® modules that have a 9-pin D-sub connector (DE-9P), and you can use your existing cables and electronics.
<select> key. Pressing the switch upward is referred to as pressing the <up> key. Pressing the switch downward is referred to as pressing the <down> key. During setup, and operation, be sure to consider the Torr/mTorr LEDs. InstruTech Page 11...
It is good practice to perform the sequence of checking and adjusting span (ATM) then zero (VAC) and then, finally re-checking the span setting to ensure that the circuitry is properly balanced for use in measuring pressure throughout the intended measurement range. InstruTech Page 12...
5 seconds has passed, the display will return to normal pressure display without resetting to factory defaults. If you reset all values to original factory default settings, you would need to repeat the initial setup procedure as described in section 4.1 and reprogram other parameters as required. InstruTech Page 13...
To return all settings to original factory defaults, press <up> key and hold for 5 seconds. Display will read "dEF". After 5 seconds, factory default settings will replace all user-settings, and readout will return to normal pressure display. InstruTech Page 14...
Under normal conditions the voltages and currents in InstruTech convection gauges are too low to cause ignition of flammable gases. However, under certain failure conditions, sufficient energy could be generated to cause flammable vapors or gases to ignite or explode.
Page 20
The X-axis is the pressure reading indicated by the convection gauge under test. This chart shows readings for an InstruTech convection gauge (CVG) and Granville-Phillips® Convectron® gauge to illustrate that the difference in the response for both of these types of gauges is virtually indistinguishable.
2) Gas used is argon (Ar). Display shows pressure measurement of 600 mTorr. True pressure of argon is 1 Torr. 3) Gas used is oxygen (O ). Display shows pressure measurement of 486 mTorr. True pressure of oxygen is 500 mTorr. InstruTech Page 17...
Log-Linear Output Output - Volts Output - Volts Many InstruTech modules and controllers also provide a log-linear output signal, as an alternative to the non- linear signal described above. This output, shown at right, is a 1 Volt per decade signal that may be easier to use for data logging or control.
= Pressure = 1.0E-03 Torr The equations listed above are used to calculate the non-linear voltage outputs for N /air shown in the table below. Non- linear voltage outputs for various other gases are also shown in the same table. InstruTech Page 19...
Page 24
An analog output of less than .01 volts to near 0 volt indicates a damaged or faulty sensor. Refer to the next page if you have a CVM211 Stinger with a log-linear analog output. InstruTech Page 20...
The chart on the following page shows the graphical results of the table and formulas given above for nitrogen. True pressure (N ) is plotted on the X-axis with a log scale. The output signal is plotted on the Y-axis on a linear scale. InstruTech Page 21...
Page 26
Instruction Manual CVM211 Stinger Log-Linear Analog Output Voltage vs. Pressure 0.0001 0.001 0.01 1000 Pressure (Torr) Chart of the calculated pressures using the formulas and data for the log-linear output signal for nitrogen from the previous page. InstruTech Page 22...
Care should be exercised when using gases other than nitrogen (N ) / air. Maintenance In general, maintenance is not required for your InstruTech module. Periodic performance checks may be done by comparing the gauge to a known reference standard. Troubleshooting...
Thermal vacuum gauges may be available with different sensor materials that are not as reactive with your particular process gases. The standard gold plated tungsten sensor used in the InstruTech convection gauge is offered for use with air and inert gases such as N , argon, etc.
The CVM211 module is factory calibrated for the specific sensor (gauge tube) installed in it. If the device fails for any reason, return the CVM211 to the factory to determine if either the sensor or the electronics could be replaced or if the entire module should be replaced. InstruTech Page 25...
(examination of which shall disclose to SELLER's satisfaction that it is defective) returned, to SELLER's plant, properly identified within twenty four (24) months (unless otherwise noted) after the date of shipment from InstruTech Plant. BUYER must obtain the approval of SELLER and a return authorization number prior to shipment.
Page 31
InstruTech® 1475 S. Fordham St. Longmont, CO 80503 Phone +1-303-651-0551 Fax +1-303-678-1754 E-mail info@instrutechinc.com www.instrutechinc.com p/n 000219-110...