Page 1
G5102-i Analyzer for Isotopic N User Manual Picarro Inc. 3105 Patrick Henry Drive Santa Clara, CA 95054, USA Phone: +1 408 962 3944 www.picarro.com Document Number 40-0007 Rev. C...
Page 2
NOTICES DISCLAIMER AND RESERVATION OF RIGHTS Picarro has prepared this manual solely for the information and use by its customers as a guide for the selection, installation, operation, and maintenance of the products described.
ANALYZER BASIC OPERATION ............... 30 Startup Procedure ................30 Desktop Icons and Folders ..............31 Shutdown Procedure ................. 34 Recovery from Electrical Power Outage ..........35 LIST OF GUI FUNCTIONS ................36 GUI Overview ..................36 P/N 40-0007 Rev. C G5102-i User Manual...
Page 5
15. CONTROLLING EXTERNAL VALVES ............76 15.1 Setup ....................76 15.2 Display the Show/Hide Valve Sequencer GUI ........79 15.3 The External Valve Sequencer Window ..........80 15.4 File Menu ................... 80 15.5 Action Menu ..................81 G5102-i User Manual P/N 40-0007 Rev. C...
Page 6
15.10 Load and Run a Saved Sequence ............. 84 16. LIMITED WARRANTY ................85 REMOTE DATA ACCESS ............87 A.1 Picarro Serial Communication ............... 87 A.2 Remote Data Access ................87 THE DATA FILE VIEWER ............90 B.1 Introduction ................... 90 B.2 The File Menu ..................
In the cavity, a series of mirrors reflects the infrared light through the sample, increasing the path length. For a Picarro cavity of only 25 cm in length, the effective path length of the cavity can be over 20 kilometers.
A cavity containing gas that absorbs light will have a shorter ring- down time than an empty cavity. As the light circulates in a cavity with a gas sample, the molecular absorption by the gas results in a decrease of the light intensity. P/N 40-0007 Rev. C G5102-i User Manual...
Picarro analyzers calculate the baseline underneath a poorly resolved peak by modeling the absorption peaks from other surrounding molecules and subtracting contributions from neighboring peaks to the absorption intensity.
Page 11
For precise pressure control, the cavity pressure is monitored using a high- linearity pressure transducer. The system computer uses this pressure data in a feedback loop to control proportional valves that adjust the inlet and outlet gas flow of the cavity. G5102-i User Manual P/N 40-0007 Rev. C...
CAUTION alerts you of a potential danger to equipment or to the user. CAUTION WARNING indicates an imminent danger to the user. WARNING REMINDER is a helpful hint to procedures listed in the text. REMINDER P/N 40-0007 Rev. C G5102-i User Manual...
Data Acquisition System (the Analyzer) Pump Power Unit Graphical User Interface centimeters millimeters ” (as in 1/4”) Inches Nitrous Oxide Water Hotbox Warm box Parts Per Million Parts Per Billion ‰ per mil °C degrees Celsius G5102-i User Manual P/N 40-0007 Rev. C...
SAFETY General Safety CDRH Certification This Picarro Analyzer complies with 21 CFR Chapter 1, sub-chapter J, and is classified as a Class 1 laser system when all panels and covers are on. CE Certification This Picarro Analyzer complies with the European standards and the instrument is affixed with a CE label.
WARNING laser safe service area using appropriate laser-safety glasses. The following Laser Safety Label is affixed to the inside of the analyzer: G5102-i User Manual P/N 40-0007 Rev. C...
Page 16
Figure 4: Laser Safety Label Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure. WARNING P/N 40-0007 Rev. C G5102-i User Manual...
UNPACKING THE ANALYZER Inspect the Shipping Boxes Picarro products are inspected and tested before leaving the factory. Their packing containers have been designed to keep the equipment safe from damage during transit. Inspect the condition of the boxes upon arrival. The system is delivered in...
• Keep the shipping crate and all packaging to reuse when transporting the analyzer. • Contact Picarro for options on transporting systems to remote labs. This analyzer and pump weighs 95 lbs. (43.1 kg). Use the technique outlined in the General Safety section on page 15 when lifting or moving the analyzer.
Page 19
For External Solenoid Valves Cable (1) (Two electronic cables (one 25-pin, and one 9-pin) are used in units manufactured prior to 2018.) Nut (1) and For connecting input line to analyzer Ferrules (2) INPUT G5102-i User Manual P/N 40-0007 Rev. C...
Page 20
PPU. Leave enough free space for ease in making connections. Leave 6” free space at the rear of the analyzer and 6” free space at the front of the analyzer. CAUTION P/N 40-0007 Rev. C G5102-i User Manual...
ANALYZER OVERVIEW Intended Use The G5102-i analyzer measures concentrations of Isotopic N N) using Picarro’s patented Cavity Ring-Down Spectroscopy (CRDS). The analyzer can be deployed for air monitoring applications in a lab or in the field, allowing in-situ analysis of trace and ambient amounts of Isotopic N...
The coordinator gets data from the GUI and creates a single file, the data is reported as a function of sample. 5. Before shutdown, Picarro recommends flowing clean dry air (CDA) through the analyzer for several minutes. This prevents moist gas in the cavity from condensing on the optics as the analyzer cools.
INSTALLATION This section describes the setup and installation of the Picarro Analyzer. Please read and understand this section thoroughly before proceeding with the installation. Using this analyzer in a manner not specified by Picarro may result in damage to the analyzer and render it unsafe to operate.
This requires two persons for lifting the hardware. This analyzer and pump weighs 95 lbs. (43.1 kg). Use the technique outlined in the General Safety section on page 15 when lifting or moving the analyzer. WARNING P/N 40-0007 Rev. C G5102-i User Manual...
Placement 3. Unpack the analyzer connection accessories. The Certificate of Compliance and USB drive should be stored in a safe place and may be required if you contact Picarro with questions about the analyzer. Connections Figure 11 shows the analyzer rear panel with the connections labeled. The numbered steps in below correspond to the numbered list below.
Page 26
8. Leave the exhaust port open to the atmosphere. Use the AC power cables supplied with the analyzer or a similarly rated cable. Check with Picarro technical support if you have questions about power cable replacement. An inadequately rated CAUTION power cable can result in equipment damage.
¼" PFA inlet nut and ferrules. When using compressed gases, follow all appropriate safety conventions, including use of eye protection, physical restraint of cylinders, etc. WARNING Making a New Connection: When using new tubing, follow these steps. G5102-i User Manual P/N 40-0007 Rev. C...
3. Using a 5/8" wrench (not included), tighten the nut approximately one flat (60 degrees). Water Removal from the Gas Stream The G5102-i analyzer requires water removal from the gas stream prior to its admission to the analyzer. In order to achieve this purpose, Picarro has ®...
5. Select N2O from the Data Key dropdown menu in the CRDS Data Viewer screen. Figure 14. 6. Restart the analyzer: double-click the Picarro Mode Switcher icon on the desktop and select Isotopic N2O. If you get logged off of Windows at some point, the username you log into is Picarro and Password: Extreme_Science! P/N 40-0007 Rev.
3. Picarro Mode Switcher: When clicked, you will be led to a window that will allow you to switch between various measurement modes. Most analyzer models are configured for one mode and may not include the Mode Switcher.
Page 32
Figure 15: Picarro Mode Switcher 4. Picarro Controller: When clicked, you will be led to a useful diagnostic panel allowing the user to see the analyzer’s internal temperatures, pressure, and spectroscopy in real time. This program has user-accessible functions, but it cannot change anything related to analyzer functionality.
Page 33
(typically 12VDC <1A max). It is important to electrically isolate the solenoid valves from the CAUTION analyzer’s ground to avoid electrical interference which could compromise the analyzer’s accuracy. G5102-i User Manual P/N 40-0007 Rev. C...
FOR G2xxxx ANALYZERS: If you have trouble turning off the analyzer software, do not kill the process(es) in the task manager; rather, double-click on the "Stop Instrument" icon in the Diagnostics folder on the desktop. CAUTION P/N 40-0007 Rev. C G5102-i User Manual...
• Data, analyzer diagnostics and other parameters recorded up to the time of power outage are retained. If short power outages will be a routine operating environment, Picarro recommends using power conditioning or an uninterrupted power supply to help prevent damaging operating system and software corruption problems that can occur with repeated crashes.
LIST OF GUI FUNCTIONS GUI Overview The features of the Picarro analyzer GUI are identified in Figure 18 and described in the following sections. Note that the types of data shown in the following illustrations may differ from this model. This section generally shows the available features common to most models.
Figure 21: Tools Menu 1. User Calibration: Opens the user calibration window (default password is “picarro”). The password can be reset in the QuickGui.ini file in the instrument directory: “C:\Picarro\G2000\AppConfig \Config\QuickGUI\” under the section: [Authorization] UserCalPassword = Picarro Show/Hide Valve 2.
Type the value you wish to set the alarm to and press the “OK” button, or press “Cancel” if you do not wish to change the alarm value. If you do nothing, the dialog box will disappear and the alarm value will remain unchanged. P/N 40-0007 Rev. C G5102-i User Manual...
The user can select which data stream is displayed using combinations from the Data Source and Data Key pull- down menus. The precision displayed can be adjusted using the “Precision” menu and Auto-scaling of the ‘Y’ axis is also available. G5102-i User Manual P/N 40-0007 Rev. C...
Page 40
4 digits of precision or “auto.” The currently selected precision is displayed during operation. This does not affect the precision of the saved data in the data log files or results files. Figure 28: Precision Pulldown P/N 40-0007 Rev. C G5102-i User Manual...
Page 41
To auto scale, the y-axis of either graph, use the auto-scale buttons below the graph. To lock or unlock the time axis of each graph during zooming, select that menu item in the ‘View’ menu. G5102-i User Manual P/N 40-0007 Rev. C...
Page 42
Press this button if you would like the instrument to start recording a separate data file. A dialog box will appear prompting you for a filename and location. Press this button again to stop recording the data file. P/N 40-0007 Rev. C G5102-i User Manual...
The valve control system begins to allow flow through the analyzer and stabilizes the pressure inside the cavity. 6. Measuring This is the normal mode of operation after startup has completed. Figure 32: Status Log Window G5102-i User Manual P/N 40-0007 Rev. C...
10. MAINTENANCE Picarro analyzers require minimal service or maintenance. With the exception of the particulate filter, and refilling the chiller, and replacing fuses, the analyzer is not user serviceable. Before performing any maintenance shut down the analyzer. 10.1 Cleaning Clean the outside of the analyzer with a clean dry cloth. Users should never access or clean the inside of a Picarro analyzer.
Gasses are filtered by two in-line, sub-micron particulate filters before the reaching the measurement cavity. The outer filter is user-replaceable. Replacement filters can be purchased from Picarro and installed by users. This section describes outer filter replacement. The inner filter is NOT user replaceable. Do NOT open the hot box.
1. Move the analyzer to a clean environment. 2. Using the 2mm hex driver, remove analyzer top lid by removing six (3 per side) M3 x 6mm socket flathead screws. 3. Lift the top cover off of the analyzer. P/N 40-0007 Rev. C G5102-i User Manual...
Page 47
Filter Foam Location 4. Remove the foam sheet from the top of the analzyer. Figure 37: Filter Exposed Filter Location 5. Loosen the filter output nut. (Fig 37) Figure 38: Filter Outlet Remove G5102-i User Manual P/N 40-0007 Rev. C...
• Picarro recommends inspecting the coolant level in the buffer tank and adding or refreshing in case of coolant loss or contamination.
Page 49
5. Disconnect the quick connectors from each other, as they are shipped plugged into each other. Then plug them into the two quick adaptors at the bottom of the buffer tank, as shown in Figure 42. G5102-i User Manual P/N 40-0007 Rev. C...
Page 50
3. Turn on analyzer to circulate coolant in tank for 5-10 minutes. 4. Check level and flow. Add more coolant if necessary. 5. If coolant is contaminated, turn off analyzer, drain buffer tank, and reconnect the tubes, before performing Steps 2 - 4. P/N 40-0007 Rev. C G5102-i User Manual...
This section lists problems that may occur during installation and operation of the analyzer, and step-by-step procedures that will provide resolution in most cases. If these instructions do not solve the problem, contact Picarro Technical Support. 11.1 Power LED on Analyzer Does Not Illuminate Pressing the front panel ON button does not start the analyzer.
Check the vacuum line between the analyzer and the power vacuum unit for leaks. b. Check the vacuum pump. Make sure it is functioning correctly. c. Check the gas pressure inlet for excessive pressure. P/N 40-0007 Rev. C G5102-i User Manual...
1. Hold down the power switch on the front panel for a few seconds until the computer and the analyzer turn off. 2. After another few seconds, restart the analyzer using the power button. G5102-i User Manual P/N 40-0007 Rev. C...
File deletion frequency and details can be modified in the file: C:\Picarro\SI2000\AppConfig\Config\Archiver\Archiver.ini. See File Archival and Automatic Deletion of Old Files. During data acquisition, the analyzer creates directories to store the data, filed by the date the data were acquired.
For more information about how to include various columns of data from the DataLog_Private in the C:\UserData DataLog files, contact Picarro. The currently active data file can be found in C:\Picarro\ G5000\Log\DataLogger\. Archive directory contain subdirectories arranged by file type and internally organized by \year\month\day\hour.
Page 56
In addition to the automatic file and directory management described above, the analyzer also automatically deletes various files specified in C:\Picarro\ SI2000\AppConfig\Config\FileEraser\FileEraser.ini. There are various settings, as described below: runtime_interval_hrs = 0.5 Specifies how often (in hours) to run the file eraser.
12.2 Setup Tool In the desktop folder called Picarro Utilities, the Setup Tool can be launched by double clicking on its icon. The tool allows the user to configure data file saving details, including which data elements are written to data files, digital data output (via serial port or TCP/IP), remote data delivery (via email), and general GUI properties.
Page 58
Make sure there are no COM port conflicts before clicking “Apply.” After making the appropriate edits, click “Apply” to put changes into effect and then “Exit” to close the window. Figure 45: Serial/Socket Port Setup Window P/N 40-0007 Rev. C G5102-i User Manual...
Page 59
• Data Directory: Location of the data you want email. After making the appropriate edits, click “Apply” to put changes into effect and then “Exit” to close the window. Figure 46: Data Delivery Setup Window G5102-i User Manual P/N 40-0007 Rev. C...
Page 60
5. Click “Apply” to put changes into effect and then “Exit” to close the window. You should now be able to access the “Show/Hide Valve Sequencer GUI” menu from the main GUI under “Tools” Figure 47: GUI Properties Window P/N 40-0007 Rev. C G5102-i User Manual...
Page 61
Datalog_User and DataLog_User_Sync. After making the appropriate edits, click “Apply” to put changes into effect and then “Exit” to close the window. Figure 48: Command Interface Window G5102-i User Manual P/N 40-0007 Rev. C...
Page 62
Data Streaming outputs data continuously, whereas the Command Interface needs commands to prompt data output. After making the appropriate edits, click “Apply” to put changes into effect and then “Exit” to close the window. Figure 49: Data Streaming Window P/N 40-0007 Rev. C G5102-i User Manual...
Customizing Analog Output Channels The Picarro analyzer may be optionally configured with an Electrical Interface Card (EIC) that provides up to 8 analog signals available to the user for monitoring various measurements results and analyzer parameters. Electrical Interface The Electrical Interface tab allows you to customize each analog output channel.
13.1 Slope and Offset Every Picarro analyzer ships with a default user calibration value of 1 for slope and 0 for offset. While new user calibrations will likely result in changes to both, we recommend that only offset changes are applied during more frequent calibrations.
2. Plot these values versus the certified values from the gas supplier, and determine the linear relationship between the known calibration values and the analyzer’s reported values. A linear best-fit equation can be calculated from the data. G5102-i User Manual P/N 40-0007 Rev. C...
Page 66
Data corrected = 0.9866•Dataraw+5.268 Figure 51: Linear Calibration Example Calibration values are input into the analyzer by selecting Tools > User Calibration, and then entering the slope and intercept for each species. P/N 40-0007 Rev. C G5102-i User Manual...
Graphical User Interface (GUI) The Picarro Data Recalibration (“Data Recal”) software can be found in the “Picarro Utilities” folder on the desktop. The Data Recal software utility can be launched by double-clicking on the Data Recal Icon in the folder. Shown below is a screenshot of the window that will open.
Page 68
Certified and CRDS reported values are entered in the white boxes. Standards that will be used for instrument recalibration are then selected by checking the corresponding box in the first column, labeled “Used for Recal.” P/N 40-0007 Rev. C G5102-i User Manual...
Page 69
Once these values are entered or selected, the “Compute” button will be active. Click “Compute” to calculate the new calibration parameters. These parameters will appear in the Calibration output section of the GUI, G5102-i User Manual P/N 40-0007 Rev. C...
Page 70
The top graph presents measurements for the recalibrated standards (calibration and QC), as well as a line fit through those points. The bottom graph shows the residuals, which shows how far each measurement is from the best-fit line. P/N 40-0007 Rev. C G5102-i User Manual...
Page 71
“Apply New Cal” button. Figure 58: Apply new calibration slope and intercept Then you will be prompted to enter a Calibration Password. The default Calibration Password is PICARRO. Figure 59: User Authorization Dialog box Next, a “Recalibration Confirmation” pop-up window will appear, displaying the new offset and slope (when applicable) values.
Page 72
Figure 61: Delta Recalibration Load File Next, select the file that you want to load. Figure 62: File Browser Delta Recal File Option The recalibration file for this example will contain the following information: P/N 40-0007 Rev. C G5102-i User Manual...
Page 73
Figure 64: Recalibration Exit Confirmation When you are finished with the calibration, you can exit the Delta Recal software utility by clicking on the “Exit” button at the bottom part of the Recal GUI. G5102-i User Manual P/N 40-0007 Rev. C...
Page 74
(3) the “offset + slope” NOTE calibration option is selected. This error occurs because the entered values will lead to an erroneous zero-slope value. Figure 66: Invalid New Slope Entry P/N 40-0007 Rev. C G5102-i User Manual...
Startup Pressure Stabilizing/Locked The valve control system begins to allow flow through the analyzer and stabilizes the pressure inside the cavity. Measuring This is the normal mode of operation after startup has completed. G5102-i User Manual P/N 40-0007 Rev. C...
This section explains how to control external valves using the built-in External Valve Sequencer GUI. If you are using the optional Picarro 16 Port Distribution Manifold with the analyzer, see the 16 Port Distribution Manual User Guide for instructions on using the External Valve Sequencer GUI.
The current step, elapsed duration, and valve states are shown in the top most row of the valve sequencer command window. The G5102-i User Manual P/N 40-0007 Rev. C...
Page 78
For each individual measurement the analyzer makes, the valve codes and rotary valve positions corresponding to the valve state(s) at that point in time are saved alongside the concentration data. P/N 40-0007 Rev. C G5102-i User Manual...
1. Start the analyzer by double-clicking the Start Analyzer shortcut on the desktop. 2. Wait for the analyzer to begin measuring. 3. On the status bar menu of the analyzer Data Viewer screen, select Show/Hide Valve Sequencer GUI dropdown (Figure 68). G5102-i User Manual P/N 40-0007 Rev. C...
File Menu Load a Valve Sequence File Click here to load saved files. Sequence Files are in: C:\Picarro\G2000\InstrConfig\ValveSequencer\Name of the Sequence File Save a Valve Sequence File Click Action/Save Valve Sequence File to save a valve sequence file. P/N 40-0007 Rev. C...
Click here to change current valve states to All off. Hide Sequencer Interface This toggle shows or hides the Sequencer Interface window. To reveal the window click on the Show Sequencer Interface. Figure 72: Valve Sequencer Action Menu G5102-i User Manual P/N 40-0007 Rev. C...
Enter how long you want the step to last. Rot. Valve Code Used only with the gas distribution manifold 15.8 Bottom Panel Use the bottom section of the window (Figure 69) to configure the sample. P/N 40-0007 Rev. C G5102-i User Manual...
Sequencer window. This section explains how to set up a test sequence using the External Valve Sequencer window. To set up a sampling sequence: 1. On the Picarro analyzer GUI, go to: a. Tools b. Show/hide valve sequencer c. The External Valve Sequencer window displays. See Figure G5102-i User Manual P/N 40-0007 Rev.
Load saved sequence files by clicking File/Load Valve Sequence. All the Sequence Files are in C:\Picarro\G2000\InstrConfig\ValveSequencer\Name of the Sequence File. 1. Click Apply to run the sequence. 2. To skip the next step click Run Next Step (even while in middle of running a step): 3.
Product specifications for twelve (12) months from shipment. This warranty is the only warranty made by Picarro with respect to its Products and no person is authorized to bind Picarro for any obligations or liabilities beyond this warranty in connection with its Products.
Page 86
The following warranty conditions shall apply to all Picarro, Inc. products unless amended by a written instrument signed by a duly authorized officer of Picarro: ADJUSTMENT – No electrical, mechanical or optical adjustments to the product(s) are permitted. PARTS AND LABOR - New or factory-built replacements for defective parts will be supplied for twelve (12) months from date of shipment of the product.
Guide (included in pdf format on the installation CD). This command set may also be used across a TCP/IP interface through an Ethernet connection. Please contact Picarro for further details. A.2 Remote Data Access Using the RemoteAccess.ini file, the analyzer can be configured to automatically: •...
Page 88
RFC2554. If such authentication is not needed, the key UseAuthentication is set to false. If this key is set to true, two additional keys Username and Password must also be specified for the e-mail account. P/N 40-0007 Rev. C G5102-i User Manual...
Open H5 File > Open H5 opens a Picarro data file (HDF5 format) for data analysis and visualization. After opening the data file, you can create a new time series plot. Refer to New Time Series Plot for more information.
H5 file. The variables are then listed in the left panel, and users can use “>>” button to move variables to the right panel for concatenation. Figure 75: Select Variables Form G5102-i User Manual P/N 40-0007 Rev. C...
Page 92
Range Data files are saved in directory trees named by date and time option. Picarro software saves data in a directory tree that is named by the creation year, month, and day. (See example directory tree in the following image.) Select this option if the target folder has this file structure. This way, Data File Viewer will only search folders within the desired date range, which can substantially reduce processing time.
Page 93
Convert H5 to DAT Select File > Convert H5 to DAT to convert a file in a HDF5 format to DAT. These formats are described in Convert DAT to H5. G5102-i User Manual P/N 40-0007 Rev. C...
Page 94
Because the data interval is normally not a constant (unless interpolation is performed), fluctuations in the data interval will affect block averaging if the block size is comparable to the average data interval. P/N 40-0007 Rev. C G5102-i User Manual...
The Time Series Viewer form includes the following menus: Figure 79: Time Series Viewer Menus Time Series Viewer File Menu Use the File menu to save a configuration or take a screenshot. Figure 80: Time Series Viewer – File Menu G5102-i User Manual P/N 40-0007 Rev. C...
Viewer and save it as a .png to a specified file. Time Series Viewer Analysis Menu Use the Analysis menu to calculate statistics, generate a histogram, and to plot correlations and Allan Standard deviations. P/N 40-0007 Rev. C G5102-i User Manual...
Page 97
• Red Line: A Gaussian function fitted to the histogram. Fitting results of μ and σ are shown in the top-left corner of the plot. • Bin: Specifies the number of intervals that the range of values is divided into. G5102-i User Manual P/N 40-0007 Rev. C...
Page 98
Viewer – View Menu When switching from DateTime to Minute or Hour, the X-axis data is subtracted from the earliest point shown in the panel and then converted to the desired unit. NOTE P/N 40-0007 Rev. C G5102-i User Manual...
• Right-click: Opens an additional menu. Refer to the next section. Right-click Menu Right-clicking on the canvas provides opens a pop-up menu. Figure 86: Canvas Right- click Menu Export Image: Exports the current plot as a jpeg, png, or pdf file. G5102-i User Manual P/N 40-0007 Rev. C...
Page 100
• Label: Specify a label for the X-axis and the Y-axis. Figure 87: Image Editor Form • Statistics: Calculates the mean, standard deviation, and peak to peak for data in the current view. P/N 40-0007 Rev. C G5102-i User Manual...
Page 101
Analysis. For details about the File menu, see Save Configuration on page 96. The canvas in this plot is interactive. For details about the plot canvas, see The Time Series Viewer Canvas on Page 99. REMINDER G5102-i User Manual P/N 40-0007 Rev. C...
Polynomial fit: Specifies to fit polynomial function of degree n: y = Σc Curve fit: Specifies to use non-linear least squares to fit an arbitrary function to data. Figure 89: Analysis Menu P/N 40-0007 Rev. C G5102-i User Manual...
Page 103
After applying any of the above Analysis options, the results, statistics, or fitting function with coefficients are displayed in the lower portion of the Correlation Plot window. Figure 90: Results of Quadratic Fitting G5102-i User Manual P/N 40-0007 Rev. C...
LaserBasePlateTec chiller TEC is used to remove heat from chiller liquid Temperature reading of Laser Baseplate. The laser is LaserBasePlateTemp Celsius mounted on the baseplate. The baseplate temperature is set at 30 Celsius P/N 40-0007 Rev. C G5102-i User Manual...
Page 105
The amount of Wavelength Monitor (WLM) offset change adjust1 for virtual laser 1 after each spectral scan The amount of Wavelength Monitor (WLM) offset change adjust4 for virtual laser 2 after each spectral scan G5102-i User Manual P/N 40-0007 Rev. C...
Page 106
Peak corresponding to Alpha isotopic N2O derived from peak4 ppb/cm fitting Peak corresponding to Alpha isotopic N2O with N2O and peak4_spec ppb/cm CO2 corrections Peak corresponding to Beta isotopic N2O derived from peak5 ppb/cm fitting P/N 40-0007 Rev. C G5102-i User Manual...
Page 107
WLM offset for virtual laser 2 wlm3_offset WLM offset for virtual laser 3 Wlm4_offset WLM offset for virtual laser 4 wlm5_offset WLM offset for virtual laser 1 G5102-i User Manual P/N 40-0007 Rev. C...
G5000 instruments, a continuous correction cannot be made. Instead Picarro uses an automated program (called Maintainer) to run maintenance on the G5102-i at discrete times after a prolonged period of not being run. Maintenance is performed on a monthly basis. When the instrument software is started and it’s time for maintenance, the panel below will be shown.
Page 109
Figure 92: Type description here G5102-i User Manual P/N 40-0007 Rev. C...
PACKING AND SHIPPING The following describes how to prepare the G5102-i for shipment. Shown here is a step by step procedure which, if followed correctly, will help ensure the safe delivery of the system. Please follow the instructions carefully. 1. Ensure the analyzer has been fully powered down following the previous instructions.
Page 111
7. Secure the pump by rotating the black cover plate so that the screw holes in the black plate line up with the through holes in the bottle of the PPU cover and tightening the screws. The proper orientation is shown below. G5102-i User Manual P/N 40-0007 Rev. C...
Page 112
Figure 98: Bagging the 9. Place the analyzer in its plastic bag with two desiccant pouches as shown below. Again, ensure that the bag remains closed by sealing with 2” wide packing tape. P/N 40-0007 Rev. C G5102-i User Manual...
Page 113
Make sure analyzer is placed in this area such that it fits completely inside the packing foam (see figure below). Figure 100: Pump Analyzer Unit Packing Location Location Locations G5102-i User Manual P/N 40-0007 Rev. C...
Page 114
13. Place the plastic bag containing the two power lines, two electronic cables and one vacuum line on top of the PPU as shown in the figure below. Figure 102: Add Cable Bag and Foam Top Piece P/N 40-0007 Rev. C G5102-i User Manual...
Page 115
16. Place the lid on the wooden crate as shown in the figure below. The metal clamps shown in the figure are used to secure the lid to the wooden crate. Figure 104: Ready For Clamps G5102-i User Manual P/N 40-0007 Rev. C...
Page 116
19. When completely assembled, the packing crate should look like the system shown in the figure below. Figure 106: Ready to Ship 20. To finish, remove any old shipping labels from the crate and attach new labels as appropriate. P/N 40-0007 Rev. C G5102-i User Manual...