Heal Force Prince-180D Instructions To User

Easy ecg monitor

Advertisement

Quick Links

User Manual for Prince-180D Easy ECG Monitor
Instructions to User
Dear Users,
Thank you very much for purchasing our product. Please read the manual very carefully before using this device.
These instructions describe the operating procedures to be followed strictly. Failure to follow these instructions may cause
measuring abnormality, device damage, or personal injury.
The Manual is published in English and Shenzhen Creative Industry Co., Ltd. (Creative) has the ultimate right to explain the
Manual. No part of this manual may be photocopied, reproduced or translated into another language without the prior written
consent. Creative reserves the right to improve and amend it at any time without prior notice.
Version of This Manual: Ver1.2
All rights reserved.
3502-3430014
I

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Prince-180D and is the answer not in the manual?

Questions and answers

Summary of Contents for Heal Force Prince-180D

  • Page 1 User Manual for Prince-180D Easy ECG Monitor Instructions to User Dear Users, Thank you very much for purchasing our product. Please read the manual very carefully before using this device. These instructions describe the operating procedures to be followed strictly. Failure to follow these instructions may cause measuring abnormality, device damage, or personal injury.
  • Page 2 User Manual for Prince-180D Easy ECG Monitor WARNING 1. Check the device to make sure that there is no visible damage that may affect user’s safety and measurement performance. Stop using the unit, when there is obvious damage. 2. Stop using the ECG cable if it is damaged, please change a new one.
  • Page 3: Table Of Contents

    User Manual for Prince-180D Easy ECG Monitor Table of contents   1 Overview ......................................1   1.1 Appearance ....................................1   1.2 Name and Model ..................................2   1.3 Structure ...................................... 2   1.4 Features ....................................... 2   1.5 Intended Use....................................3  ...
  • Page 4 User Manual for Prince-180D Easy ECG Monitor   6 Data Management ..................................16   6.1 Storage Status .................................... 16   6.2 Delete ID ....................................17   6.3 Delete All ....................................18   6.4 Upload Data ....................................18   7 Measurement Option Setting ................................ 18  ...
  • Page 5 User Manual for Prince-180D Easy ECG Monitor   10.7 Additional Declaration ................................24   11 Maintenance ....................................25   12. Troubleshooting ................................... 26   13 Key of Symbols ..................................... 27   Appendix 1 Additional Measuring Method ............................ 28   Appendix2 Measurement Result Description ..........................30  ...
  • Page 6: Overview

    User Manual for Prince-180D Easy ECG Monitor 1 Overview 1.1 Appearance Power: Long time press (about 2 seconds) makes the device power on or off; short time press turns on/off the backlight. Back: back to upper level of menu. 3. Metal Electrode I Measurement: shortcut button for built-in electrodes measuring, press this button to start measuring.
  • Page 7: Name And Model

    Name: Easy ECG Monitor Model: Prince-180D 1.3 Structure Prince-180D Easy ECG Monitor mainly consists of circuit boards, display screen and metal electrodes/lead wires. 1.4 Features 1. Small in size, light in weight, and easy to carry. 2. One-key measurement, easy to operate.
  • Page 8: Intended Use

    User Manual for Prince-180D Easy ECG Monitor 1.5 Intended Use This Easy ECG Monitor is intended for measuring and recording the ECG waveform and average heart rate of adult patient. It is applicable for use in clinics and homes, and convenient to operate by the patients themselves.
  • Page 9: Default Menu Screen

    User Manual for Prince-180D Easy ECG Monitor 3 Default Menu Screen 3.1 Menu Screen Long time press the“ ”Power key (lasting for 2 seconds) until a beep sound is heard to indicate the unit is turned on, then the booting screen will appear, as shown in the figure 3-1, after that the device enters into the default menu screen, as shown in figure 3-2.
  • Page 10: Ecg Measurements

    User Manual for Prince-180D Easy ECG Monitor 4 ECG Measurements There are 2 measuring modes for the ECG measurement, which includes quick measurement by built-in electrodes and long term measurement by external lead wires. Quick measurement: press “ ” measure key to enter into the quick measurement directly; it can display single channel ECG signal measured by built-in electrodes, the signal can be lead I, lead II or chest lead according to the electrode placement.
  • Page 11 User Manual for Prince-180D Easy ECG Monitor B. Wrist placement: Hold the device with right hand; make sure that the fingers touch the metal electrode II and III reliably. Press the electrode I below the left wrist about 1 inch, as shown in figure 4-2.
  • Page 12 User Manual for Prince-180D Easy ECG Monitor the electrode I on the left chest align with the midclavicular line (V4 position) or the anterior axillary line(V5 position) and on the level of the 5 intercostal space. Note: keep at ease and clean the electrode I by using the wet cloth before measuring. If the waveform baseline wanders or noisy waveform appears, hold your breath for a moment to get stable and clean signal.
  • Page 13: Measuring Procedure

    User Manual for Prince-180D Easy ECG Monitor use (ie. by doctor). The location of the electrodes is shown in figure 4-6. R (red): place on interior of right arm (between the wrist and the elbow); L (yellow): place on interior of the left arm (between the wrist and the elbow);...
  • Page 14: Quick Measurement By Built-In Electrodes

    User Manual for Prince-180D Easy ECG Monitor prompt:“Check well-connection with lead wires.” 4.2.2 Quick Measurement by Built-in Electrodes Please refer to the section 4.1 “Hand measurement” to make the electrodes well-connection with skin. At default menu screen, press “ ” measure key can enter into quick measurement preparing screen directly, after 10 seconds, it will enter into the measuring screen, as shown in Figure 4-8.
  • Page 15 User Manual for Prince-180D Easy ECG Monitor measuring screen, as shown in Figure 4-10. On-screen display: (1 ) “00:00:01”: Measuring time (2) “X1”:“X1”: ECG waveform amplitude scale; “X2”: twice size of the nominal scale; “X4”: four times size of the nominal scale.
  • Page 16 User Manual for Prince-180D Easy ECG Monitor ② Press the “ ” ok key or “ ” return key to stop measuring and return to the default menu screen directly, the measuring data would be saved. Note: ① If the storage is full, the device will stop to measure and it prompts “Fail to save this record because of full data memory”, at this moment, you should delete some records to make sure enough storage space, then measuring again.
  • Page 17: Data Analysis And Measurement Result Description

    User Manual for Prince-180D Easy ECG Monitor 4.3 Data Analysis and Measurement Result Description When the measuring is finished, the device will analyze the measurement result, as shown in figure 4-11/4-12/4-13; the measurement result description refers to the index 2.
  • Page 18: Waveform Review

    User Manual for Prince-180D Easy ECG Monitor 5 Waveform Review 5.1 Record List Select “ ” review key on the default menu screen, then press “ ” ok key to enter into the user ID list screen, as shown in Figure 5-1.
  • Page 19: Waveform Review

    User Manual for Prince-180D Easy ECG Monitor 5.2 Waveform Review In the record list, selecting a record, then press “ ”ok key, the quick measurement record is shown in the figure 5-3; the external lead wire measuring records is shown in figure 5-4.
  • Page 20 User Manual for Prince-180D Easy ECG Monitor The chest lead review screen is shown in figure5-7; the limb lead review screen is shown in figure 5-8. In the figure 5-7/5-8/5-9 screen, press the “ ” ok key can review the result description for this ID record, as shown in figure 5-10/5-11.
  • Page 21: Data Management

    User Manual for Prince-180D Easy ECG Monitor Figure 5-10 No irregularity found (Ext.) Figure5-11 (irregularity found in Ext.) 6 Data Management Select “Memory” icon on the default menu, then press “ OK” button to enter into data management screen, as shown in Figure 6-1.
  • Page 22: Delete Id

    User Manual for Prince-180D Easy ECG Monitor 6.2 Delete ID 1. On data management screen, select “Delete” with Navigation Key, then press “ ” ok key to enter into the ID list screen, as shown in figure 6-3. 2. Press the Navigation Key “ , , ”...
  • Page 23: Delete All

    When measuring is finished, on data management screen, select “Upload data” to enter into upload data screen, as shown in figure 6-5. On this screen, you can transmit data to PC for data management, and interpretation through Prince-180D data cable. Please Figure 6-5 refer to the ECG Viewer Manager for detailed operation.
  • Page 24: Creating New Id

    User Manual for Prince-180D Easy ECG Monitor professional use (ie. by doctor). The “limb lead” or quick measurement by built-in electrodes is recommended for normal use. The default setting is “limb lead”. Chest lead: when the measuring mode is set to “chest lead”, V lead is defaulted setting. If you want to obtain specific Chest Lead (V1-V6), select the lead name (V1-V6) from the pull-down menu.
  • Page 25: System Setup

    User Manual for Prince-180D Easy ECG Monitor finished. 8 System Setup On the default menu screen, select “Setting” with Navigation Key, then press “ ”ok key button to enter into system setup screen, as shown in Figure 8-1. 1.Press the navigation key “ , , or ”:Move the cursor upwards/backwards.
  • Page 26: Time/Date Setting

    User Manual for Prince-180D Easy ECG Monitor 8.3 Time/Date Setting Date and time setting screen is shown as figure 8-4. 1. press navigation key“ , , or ”and “ ” ok key to select the date and time you need.
  • Page 27: Help

    User Manual for Prince-180D Easy ECG Monitor 9 Help On default menu screen, select “Help”, then press “ ” ok key to enter into help info screen. In the info screen,you can browse the “FAQ”, “How to”, “ECG info” and software version information.
  • Page 28: Technical Specifications

    User Manual for Prince-180D Easy ECG Monitor 10 Technical Specifications 10.1 ECG Measurement 1. Amplification channels: 3. 2. Signal input: 3 embedded metal electrodes or external lead wire cable (with 3/4 lead wires). 3. Lead selections and measuring methods: Quick measurement by embedded electrodes; measurement by external electrodes with 3 lead wires (single channel for Lead I, II, or III) or 4 lead wires (3 simultaneous channels for Lead I, II and III) 4.
  • Page 29: Classification

    User Manual for Prince-180D Easy ECG Monitor 10.3 Classification 1. The type of protection against electric shock: Internally powered equipment 2. The degree of protection against electric shock: Type BF applied part 3. The degree of protection against harmful ingress of liquid: Ordinary equipment without protection against ingress of liquid.
  • Page 30: Maintenance

    User Manual for Prince-180D Easy ECG Monitor batteries (Capability at least 2200mAH) can be used for more than 8 hours. Maximal input DC offset voltage: ±300mV. 11 Maintenance 1. Please change the batteries, when appears low-voltage icon. 2. It is recommended to wipe the metal electrode with medical alcohol, when the measuring signal is too weak.
  • Page 31: Troubleshooting

    User Manual for Prince-180D Easy ECG Monitor 12. Troubleshooting Trouble The device can not turn on. Possible Reason Solution 1. The batteries are drained or almost drained. 1. Change batteries. 2. The batteries are not inserted 2. Reinstall batteries. properly.
  • Page 32: Key Of Symbols

    User Manual for Prince-180D Easy ECG Monitor 13 Key of Symbols Symbol Desc cription Symbol Description Heart Rate (Unit: beats per Batt tery indicator minutes) wer button shortcut utton measuring nfirmation Back button ▇ butto , , , vigation key...
  • Page 33: Appendix 1 Additional Measuring Method

    User Manual for Prince-180D Easy ECG Monitor Appendix 1 Additional Measuring Method Direction of using limb clip type electrodes: clean the electrodes by using wet cloth; plug the lead wires into socket; or affix the lead wires to clip, and then keep the clip in measuring position.
  • Page 34 User Manual for Prince-180D Easy ECG Monitor 4. F(green)electrode location of Chest lead measurement The location of ECG electrodes: electrode location for V1-V6 on the 4 intercostal space at right border of sternum on the 4 intercostal space at left border of sternum...
  • Page 35: Appendix2 Measurement Result Description

    User Manual for Prince-180D Easy ECG Monitor Appendix2 Measurement Result Description No. ECG Measurement Result Notes: (bpm:time/minute) Descriptions No irregularity found. 60 bpm ≤ HR ≤ 100 bpm, No irregular HR found. Suspected a little fast beat 100 bpm < HR <= 110 bpm Suspected fast beat HR>...
  • Page 36 User Manual for Prince-180D Easy ECG Monitor Suspected a little fast and interval A little fast beat and suspected Premature beat. with occasional short beat interval Suspected a little slow and interval A little slow beat and suspected Premature beat.
  • Page 37 User Manual for Prince-180D Easy ECG Monitor Small waveform or baseline wander is larger, Poor Signal, measure again average HR can not be measured.
  • Page 38: Appendix3 General Knowledge Of Ecg

    User Manual for Prince-180D Easy ECG Monitor Appendix3 General Knowledge of ECG ⅠNormal sinus rhythm: In sinus conditions, SA node paces the heart with the regularly rate and the normal rhythm. The heart rate is in the range of 60 to 100 times per minute and the rhythm is regular. P wave is normal and each one is followed by a QRS wave.
  • Page 39 User Manual for Prince-180D Easy ECG Monitor Symptom: heart rate<60bpm Indication: It occurs when healthy people fall asleep, and it can be found in athlete (or those who love doing sports frequently), old people, or vagus excitement person. Pathology: Sick sinus syndrome, Ischemic heart disease, Cardiomyopathy, intracranial hypertension, increased hypokalemia, Low temperature, period of convalescence of acute infectious disease or after use some medicines such as digitalis.
  • Page 40 User Manual for Prince-180D Easy ECG Monitor Testee’s Symptom: Normal heart beat is followed by a premature beat. 4) Bigeminy: It is a type of PVC in which a normal beat is coupled with a premature beat. Because of its trigger cardiac abnormal pacemaker in different position, it divided into arial, ventricular and node.
  • Page 41 User Manual for Prince-180D Easy ECG Monitor Indication:Most commonly found in healthy people, it causes by deep respiration, tachypnea, positional changes, swallow, rage etc. It also appears in functional cardiac disease,such as Wolff - Parkinson-White Syndrome, rheumatic heart disease, coronary heart disease, Cardiomyopathy, Congenital heart disease, medicinal reaction(digitalis toxicosis)etc.
  • Page 42 User Manual for Prince-180D Easy ECG Monitor...