Guidance and MANUFACTURER'S declaration – electromagnetic IMMUNITY
Guidance and manufacturer's declaration – electromagnetic immunity
The G10 is intended for use in the electromagnetic environment specified below. The customer or
the user of the G10 should assure that it is used in such an environment.
Immunity test
EC 60601 test level
Conducted RF
3 Vrms
150 kHz to 80 MHz
IEC 61000-4-6
6 Vrms 150 kHz to
80 MHz outside ISM
Radiated RF
bandsa
IEC 61000-4-3
10 V/m
80 MHz to 2.7 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
Compliance level
3 Vrms
150 kHz to 80 MHz
6 Vrms 150 kHz to 80
MHz outside ISM band-
sa
10 V/m
23
Electromagnetic environment
- guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the G10 ,
including cables, than the
recommended separation
distance calculated from the
equation applicable to the
frequency of the transmitter.
Recommended separation
distance
80MHz to 800MHz
800MHz to 2.7GHz
where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres(m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey, a should be less than
the compliance level in each
frequency range b
Interference may occur in the
vicinity of equipment marked
with the following symbol: