Wavetek 143 Instruction Manual page 21

20 mhz function generator
Hide thumbs Also See for 143:
Table of Contents

Advertisement

Table 5-1. Calibration Chart (Continued)
Cal
Control
Desired
Step
Check
Tester
Points
Settings
Adjust
Results
Remarks
13
1000:1
Scope
FUNC-
R16
Symmetrical wave·
NOTE:Steps 13and
Sym-
TION
BOD
form
14 are interactive.
me
try
OUT
Sym
14
Main
FREQ VERNIER:
R35
Symmetrical wave-
Sym-
full cw
TOD
form
metry
Dial: 2.0
Sym
FREQ MULT: 1 K
15
Sine
Distortion
FUNCTION:
'\J
R120
Symmetrical
Connect FUNCTION
Distor-
Analyzer,
Triangle
residue
OUT to distortion
ti on
Scope
Balance
analyzer and distor·
tion analyzer output
to scope. Set scope
to .1V/div.Syncscope
to SYNC OUT BNC
loaded into
son.
-
16
R93,
Minimum sine
If either adjustment is
R107
distortion
going near a stop, re-
Triangle
center both pots and
Peaks
return to step 15.
17
Main
Frequency
FUNCTION:
'1.i
R4
2000 ±10 Hz
Remove SYNC OUT
Freq
Counter/
TOD
cable.
Timer
Freq
Adj
18
Cap
FREQMULT: 10
R48
20 ±0.1 Hz
Mult
Freq
19
X 10M
FREQMULT: 10M
C40
Best frequency
Freq
Dial: Vary
tracking over
X 10M range
20
X1M
FREQMULT: 1M
C34
Best frequency
This adjustment must
Freq
Dial: Vary
tracking over
be made each time
X 1M range
step 20 is done.
21
Trigger
Scope
FUNCTION:
"v
R162
Minimum shift of
Baseline
GENERATOR
baseline around
MODE: TRIG
0 Vdc
Dial: Vary
5-3