Download Print this page

Reader Mode - National Instruments MICROPROSS MP500 TCL3 Manual

Advertisement

Improvements

READER MODE

MPC_GetDataRate: it allows to measure the VICC & VCD data rate.
Data-rate measurement: Accuracy improved for Type A & Type B - high data rate.
Reject PICC inverted modulation: New parameter on API MPC_ChangeProtocolParameters to reject
negative load modulation. See parameter CPP_REJECT_INVERTED_MODULATION (52)
Select I or Q channel: New parameter on API MPC_ChangeProtocolParameters to disable I or Q
demodulator channel. See parameter CPP_DISABLE_I_OR_Q_DEMODULATION (51)
Vicinity high data rate x8 supported. New parameter HIGH_DATA_RATE_X8 (4) for API
MPC_SelectVCCommunication.
Improvement of the reception at 6.8 Mbits/s.
Add a new polling sequence with an accurate delay. New parameter TSCN_DO_EMV_POLLING (71) for
API MPC_AddToScenarioPicc
Add DISTURBANCE_SINE2 (4) generation to reach a max value of amplitude to 2,000 ‰, and a minimum
value of offset to 0 ‰
CARD EMULATION
Frequency measurement: Accuracy improved for automatic mode.
Negative PICC Modulation generation: New parameter on API MPC_AddToScenarioPcd to enable the
negative modulation generation using T
Programming the load modulation amplitude added in the sequencer - API MPC_AddToScenarioPcd .
See parameter TSCN_SET_LMA_CARD_EMULATION (72)
WIRELESS POWER CHARGING
Add TPR-THERMAL-15W antenna.
MISCELLANEOUS
Support of new antenna: 907-1425 - NFC-Qi 15W antenna
MPC_GetInitDataImpedance (GIGI) and MPC_SetInitDataImpedance (GISI) APIs added to manage
impedance compensation data
Corrections
READER MODE
Generation of DISTURBANCE_SINE (2) corrected
MPC_GetMeasureResonanceFrequency (REGM) and MPC_GetMeasureS11 (S1GM) may crash when
measurement was performed beforehand. Fixed.
WIRELESS POWER CHARGING
WPID packet CRC computation corrected.
11-21 rue Hubble
Parc de la Haute Borne
59650 Villeneuve d'Ascq - France
. See parameter TSCN_DO_NEGATIVE_MODULATION (71)
LOAD
Tel: + 33 320 746 630
www.micropross.com
Fax: +33 320 746 635
www.micropross.com
Tel : + 33 320 746 630

Advertisement

loading
Need help?

Need help?

Do you have a question about the MICROPROSS MP500 TCL3 and is the answer not in the manual?

Save PDF