Table of Contents Preface ......................Clear and Sanitize Procedures..................Memory Devices....................Data Export Devices................... Troubleshooting..................... How to Clear or Sanitize a Non-Functional Instrument ........... How to Recover from Clearing or Removing Memory from the Instrument ......WVR4000 and WVR5000 Declassification and Security...
Page 4
Table of Contents WVR4000 and WVR5000 Declassification and Security...
The instructions also describe how to declassify an instrument that is not functioning. These procedures can be useful for customers with data security concerns. Products The following Tektronix products are covered by this document: WVR4000 WVR5000 Related Documents...
Page 6
Preface WVR4000 and WVR5000 Declassification and Security...
20 seconds SDRAM 16 M X 16 CPU RAM CPU access Main board None Remove the power source from the instrument for at least 20 seconds During normal instrument operation. WVR4000 and WVR5000 Declassification and Security...
Page 8
Set backlight Serial I2C brightness brightness to 100% brightness to 100% EEPOT 256 X 1 Battery volts low Factory/Service Backplane board None None Serial I2C calibration Battery voltage procedure error During normal instrument operation. WVR4000 and WVR5000 Declassification and Security...
Page 9
Press the SEL button to remove all entries in the diagnostic log. c. Press the > button until the box by “Exit” is highlighted. d. Press the SEL button to exit the log display. 5. Press the CONFIG button to exit the configuration menu. WVR4000 and WVR5000 Declassification and Security...
Page 10
Enter “000.000.000.000” for the IP address. d. Repeat for the Subnet Mask, Gateway Address, and SNMP Trap Address 1 through SNMP Trap Address 4. 5. Press the CONFIG button to exit the configuration menu. WVR4000 and WVR5000 Declassification and Security...
6. Press the CONFIG button to exit the configuration menu. Enable Ethernet and SNMP To enable Ethernet and SNMP access, use the same procedures you would use to disable these devices, but select On to enable each device. Access Procedure WVR4000 and WVR5000 Declassification and Security...
Page 12
Clear and Sanitize Procedures WVR4000 and WVR5000 Declassification and Security...
How to Clear or Sanitize a Non-Functional Instrument To sanitize a non-functional instrument, remove the Main board and return the instrument to Tektronix for installation of a new Main board. How to Recover from Clearing or Removing Memory from the Instrument Reload the system software according to the loading instructions.