X. CLEANING AND DISINFECTION
The thermometer is a multi-patient reusable device, and the cleaning and disinfection
instructions were validated. Please follow the instructions to clean/disinfect your thermometer
prior to use.
Cleaning:
1) Use a cotton swab moistened with hot water or neutral detergent to throughout clean the
casing of thermometer and measuring probe to remove stain or oil.
2) Inspect the thermometer visually after the cleaning process, ensure there is no soil or stain
is seen. If the device is determined not to be visually clean at the end of the cleaning step,
you should either repeat the relevant previous cleaning steps or safely dispose of the device
so that a visibly soiled device is not used again.
Disinfection:
1) Disinfect the thermometer casing and the probe for 30 seconds with a cotton swab or
non-dust cloth dipped in 75% ethanol.
2) Repeat the above steps at least 3 times, then wipe surface and probe dry with a clean
cloth, and wait about 1 minute for drying after disinfection process.
Note:
Ensure that no liquid enters the interior of the thermometer.
Never use abrasive cleaning agents, thinners, or benzene for cleaning, and never
immerse the thermometer in water or other cleaning liquids.
Take care not to scratch the surface of the probe lens and the display.
If the visual inspection identifies unacceptable deterioration such as corrosion,
discoloration, pitting, cracked, the device should not be further used and should be
discarded.
If the thermometer has been used in patients, please visually inspect the thermometer to
determine if there is stain or oil on the casing or probe after each use, and follow the
instruction of cleaning and disinfection to cleaning your thermometer.
Care and Maintenance:
Please clean the thermometer regularly, usually once a month.
If the probe tip is damaged, please contact the maintenance center.
If necessary, you can clean it properly according to the actual dirt of the product; if you
do not use this product for a long time (more than 2 months), replace all the batteries
take out.
Store the thermometer in a dry location free from dust and contamination and away from
direct sunlight.
Do not use other batteries than mentioned batteries, do not use rechargeable batteries,
do not throw in fire.