Keysight N1900 Series Installation Manual
Keysight N1900 Series Installation Manual

Keysight N1900 Series Installation Manual

Physical layer test systems
Table of Contents

Advertisement

Quick Links

Keysight N1900 Series
Physical Layer Test Systems
N1930B Physical Layer Test System Software
which supports VNA-Based and TDR-Based
Physical Layer Test System Hardware
Installation
Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the N1900 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Keysight N1900 Series

  • Page 1 Keysight N1900 Series Physical Layer Test Systems N1930B Physical Layer Test System Software which supports VNA-Based and TDR-Based Physical Layer Test System Hardware Installation Guide...
  • Page 4 FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S.
  • Page 5 Assistance Product maintenance agreements and other customer assistance agreements are available for Keysight Technologies products. For information about these agreements and for other assistance, contact Keysight. Refer to Contacting Keysight. Safety and Regulatory Information The safety and regulatory information pertaining to this product is located in...
  • Page 6 Click Help on the menu bar. This Installation Guide helps you to with PLTS software and hardware installation. PLTS Documentation Online All PLTS documentation, including the very latest online version of the PLTS Help file, can be found at: http://na.support.keysight.com/plts/help...
  • Page 7: Table Of Contents

    Contents Contents Installing the VNA-Based Physical Layer Test System Hardware ..................10 Step 1. Set Up the Personal Computer ................ 12 Step 2. Verify your System Shipment................13 Step 3. Set Up the Network Analyzer ................16 Step 4. Attach the Network Analyzer to the Test Set ........... 17 (N4420B or N4421A/B Test Set Only) ..................
  • Page 8 3.4 If you didn’t add the license before the first start of PLTS ..........70 Troubleshooting ................ 71 Electrostatic Discharge ....................72 PLTS Installation Troubleshooting ................73 Connection Manager working with .NET and the T&M Toolkit ..........74 PLTS does not recognize the Keysight USB to GPIB adapter..........74...
  • Page 9 Other Keysight USB to GPIB Adapter Problems ..............75 Problems with the VISA Layer ....................76 Problems with VISA COM (SOAP Error) .................. 76 Problems with using National Instruments I/O Devices ............80 I/O Troubleshooting Steps ......................81 Missing Microsoft Component ....................85 PLTS Hardware Troubleshooting ..................
  • Page 10: Installing The Vna-Based Physical Layer Test System Hardware

    Installing the VNA-Based Physical Layer Test System Hardware...
  • Page 11 Installing the VNA-Based Physical Layer Test System Hardware The Physical Layer Test System (PLTS) consists of the following items: • Personal computer (PC) • VNA-based system (Network analyzer and S-parameter test set) • PLTS software The installation procedure in this chapter will lead you through setting up the hardware (the PC and the VNA-based system).
  • Page 12: Step 1. Set Up The Personal Computer

    4 GB 6 GB+ 6 GB+ Virtual Memory GPIB Interface Keysight 82357A USB/GPIB Interface for No GPIB connection is required to utilize Windows or supported GPIB card (any PLTS in the off-line mode. Saved (stored) National Instruments or Keysight 82340/41...
  • Page 13: Step 2. Verify Your System Shipment

    20 lb. (9 kg). 2. Carefully inspect the system hardware to make sure that it was not damaged during shipment. NOTE If your test system was damaged during shipment, contact Keysight Technologies. Refer to Contacting Keysight.
  • Page 14 Installing the VNA-Based Physical Layer Test System Hardware Step 2. Verify your System Shipment 4. Check the accessories that were shipped with your system. Your network analyzer accessories will be checked during the network analyzer setup. Item Nr Part Number Part Description 8120-3445 GPIB Cable (3 feet)
  • Page 15 Installing the VNA-Based Physical Layer Test System Hardware Step 2. Verify your System Shipment 5. If you ordered any of the following options, check the parts. Option 1CP is shipped in a separate container. Option Number Item Part Description Number Rack mount flange kit (For use with handles) Handles (set of 2) Precision 50-ohm cables (4)
  • Page 16: Step 3. Set Up The Network Analyzer

    Installing the VNA-Based Physical Layer Test System Hardware Step 3. Set Up the Network Analyzer Step 3. Set Up the Network Analyzer 1. Using the network analyzer’s Installation and Quick Start Guide, set up the network analyzer. 2. If you are installing your network analyzer in an equipment rack, be sure to leave at least 2 rack units of space below the analyzer to install the test set.
  • Page 17: Step 4. Attach The Network Analyzer To The Test Set

    Installing the VNA-Based Physical Layer Test System Hardware Step 4. Attach the Network Analyzer to the Test Set Step 4. Attach the Network Analyzer to the Test Set (N4420B or N4421A/B Test Set Only) If your test set is not an N4420B, an N4421A, or an N4421B, continue with Step 5.
  • Page 18: Preparing The Test Set

    Installing the VNA-Based Physical Layer Test System Hardware Step 4. Attach the Network Analyzer to the Test Set Preparing the Test Set 4. Remove the trim strip from the top of the front frame. 5. Install the four lock links to the top of the front frame using eight screws. Attaching the Network Analyzer to the Test Set 6.
  • Page 19 Installing the VNA-Based Physical Layer Test System Hardware Step 4. Attach the Network Analyzer to the Test Set 7. Secure the network analyzer’s lower locking feet to the test set’s upper locking feet by inserting the shorter two screws between the two pairs of locking feet, one on each side of the instrument as shown below.
  • Page 20: Step 5. Install The Test System On A Bench Top Or In An Equipment Rack

    Installing the VNA-Based Physical Layer Test System Hardware Step 5. Install the Test System on a Bench Top or in an Equipment Rack Step 5. Install the Test System on a Bench Top or in an Equipment Rack The test system can be installed on a bench top or in an equipment rack. In all installations, consider the following ventilation requirements when deciding where to set up your test system.
  • Page 21: To Install On A Bench Top

    Installing the VNA-Based Physical Layer Test System Hardware Step 5. Install the Test System on a Bench Top or in an Equipment Rack To Install on a Bench Top 1. Place the test set and the network analyzer on a bench top. In the example illustration shown below, the system is placed on a riser as an alternative to the bench top.
  • Page 22: To Install In An Equipment Rack

    • Installation Instructions 2 Flanges • Installation Instructions NOTE If any items are damaged or missing from a kit, contact us (refer to Contacting Keysight) to order a replacement kit. Items within these kits are not individually available. Chapter 1...
  • Page 23 Installing the VNA-Based Physical Layer Test System Hardware Step 5. Install the Test System on a Bench Top or in an Equipment Rack 2. Install the rails into the equipment rack using the instructions provided. Consider that the test set is two rack units high (3.5 inches). Mount the test set immediately below the network analyzer.
  • Page 24 Installing the VNA-Based Physical Layer Test System Hardware Step 5. Install the Test System on a Bench Top or in an Equipment Rack 4. Remove the feet before cabinet mounting the analyzer using the directions imprinted on the feet. 5. Ensure there is adequate clearance between the system cabinet and the sides and back of the test system for adequate ventilation.
  • Page 25 Installing the VNA-Based Physical Layer Test System Hardware Step 5. Install the Test System on a Bench Top or in an Equipment Rack 6. Lift the test set and slide it onto the rails that you installed earlier from the front of the equipment rack.
  • Page 26: Step 6. Make The Interconnections Between The S-Parameter Test Set And The Network Analyzer

    Installing the VNA-Based Physical Layer Test System Hardware Step 6. Make the Interconnections between the S-Parameter Test Set and the Network Analyzer Step 6. Make the Interconnections between the S-Parameter Test Set and the Network Analyzer 1. Locate your system or test set and network analyzer on the following pages for information describing the interconnections between the test set and the network analyzer.
  • Page 27 Installing the VNA-Based Physical Layer Test System Hardware Step 6. Make the Interconnections between the S-Parameter Test Set and the Network Analyzer Chapter 1...
  • Page 28 Installing the VNA-Based Physical Layer Test System Hardware Step 6. Make the Interconnections between the S-Parameter Test Set and the Network Analyzer CAUTION When connecting the interconnect cables described in the remaining pages of this section, be careful to install the interconnect cables correctly. The longer end of the interconnect cable connects to the network analyzer front panel connector.
  • Page 29: N1957B Test System Interconnections

    Installing the VNA-Based Physical Layer Test System Hardware Step 6. Make the Interconnections between the S-Parameter Test Set and the Network Analyzer N1957B Test System Interconnections (N4421B Test Set with E8364B/C) CAUTION Damage to the interconnect cable can result from improper orientation of the cable. Refer to page for detailed information regarding the correct cable orientation.
  • Page 30 Installing the VNA-Based Physical Layer Test System Hardware Step 6. Make the Interconnections between the S-Parameter Test Set and the Network Analyzer Call Out Cable Part From Sequence Number Network Analyzer Test Set Z5623-20215 REF 1 SOURCE OUT REF 1 R1 OUT Z5623-20215 REF 1 RCVR R1 IN REF 1 RCVR R1 IN...
  • Page 31: N1955B Test System Interconnections

    Installing the VNA-Based Physical Layer Test System Hardware Step 6. Make the Interconnections between the S-Parameter Test Set and the Network Analyzer N1955B Test System Interconnections (or N4420B Test Set with E8363B/C Network Analyzer) CAUTION Damage to the interconnect cable can result from improper orientation of the cable. Refer to page for detailed information regarding the correct cable orientation.
  • Page 32 Installing the VNA-Based Physical Layer Test System Hardware Step 6. Make the Interconnections between the S-Parameter Test Set and the Network Analyzer Call Out Cable Part From Sequence Number Network Analyzer Test Set Z5623-20215 REF 1 SOURCE OUT REF 1 R1 OUT Z5623-20215 REF 1 RCVR R1 IN REF 1 RCVR R1 IN...
  • Page 33: N1935A And N1953B Test System Interconnections

    Installing the VNA-Based Physical Layer Test System Hardware Step 6. Make the Interconnections between the S-Parameter Test Set and the Network Analyzer N1935A and N1953B Test System Interconnections (or N4419B Test Set with E8362B/C or N5230A/C Option 225 Network Analyzer) CAUTION Damage to the interconnect cable can result from improper orientation of the cable.
  • Page 34 Installing the VNA-Based Physical Layer Test System Hardware Step 6. Make the Interconnections between the S-Parameter Test Set and the Network Analyzer Call Out Cable Part From Sequence Number Network Analyzer Test Set AD00756-1 REF 1 SOURCE OUT REF 1 SOURCE OUT AD00756-1 REF 1 RCVR R1 IN REF 1 RCVR R1 IN...
  • Page 35: Step 7. Set Up Pc To Plts Communication

    Installing the VNA-Based Physical Layer Test System Hardware Step 7. Set up PC to PLTS Communication Step 7. Set up PC to PLTS Communication Below are several different configurations depending on the following: • Whether the PNA has a 1.1 GHz CPU (2 GPIB ports) •...
  • Page 36: 1.1 Ghz Cpu Without Opt 550/551 - Legacy Mode

    Installing the VNA-Based Physical Layer Test System Hardware Step 7. Set up PC to PLTS Communication • ONLY 4 ports allowed. 1.1 GHz CPU WITHOUT Opt 550/551 - Legacy Mode 1. Using a GPIB cable, connect the PNA GPIB Talker/Listener port to the test set GPIB port. 2.
  • Page 37: 500 Mhz Cpu With Opt 550/551

    Installing the VNA-Based Physical Layer Test System Hardware Step 7. Set up PC to PLTS Communication 500 MHz CPU with Opt 550/551 1. Connect a 82357A/B to the PNA USB and to the test set GPIB port. 2. Using a GPIB cable, connect the PNA GPIB port to the PC for control of the system. Or a LAN cable (dashed line) can be used in this configuration to control the PLTS system in either SICL-LAN, COM, or DCOM.
  • Page 38 Installing the VNA-Based Physical Layer Test System Hardware Step 7. Set up PC to PLTS Communication 2. Using a second GPIB cable, connect the PNA Talker/Listener GPIB port to the PC for control of the system. Or a LAN cable (dashed line) can be used in this configuration to control the PLTS system in either SICL-LAN, COM, or DCOM.
  • Page 39: Step 8. Power Up The S-Parameter Test Set

    Installing the VNA-Based Physical Layer Test System Hardware Step 8. Power up the S-Parameter Test Set Step 8. Power up the S-Parameter Test Set 1. Ensure the available ac power supply meets the Power Source Requirements and the operating environment meets the Operating Environment Requirements listed below. Power Source Requirements Input Voltage Range 100 - 120 Vac...
  • Page 40 Installing the VNA-Based Physical Layer Test System Hardware Step 8. Power up the S-Parameter Test Set NOTE Perform Step 6 ONLY when both of the following are true. Otherwise, skip to Step * When using the Reference channel paths through the test set. See page 1-17 for more information.
  • Page 41: Installing The Tdr-Based Physical Layer Test System Hardware

    Installing the TDR-Based Physical Layer Test System Hardware...
  • Page 42 PLTS software • TDR-based PLTS hardware (one of the following systems) — Keysight 86100A/B/C Infiniium DCA Wide-Bandwidth Oscilloscope equipped with one or two 54754A Differential 18 GHz TDR/TDT Plug-in Modules using the following firmware revisions: Keysight 86100A/B: Firmware Revision 03.06 or greater Keysight 86100C: Firmware Revision 04.00 or greater...
  • Page 43 Installing the TDR-Based Physical Layer Test System Hardware NOTE If you have the VNA-based Physical Layer Test System, refer to 1 Installing the VNA-Based Physical Layer Test System Hardware for instructions on setting up that system. The following is a list of the installation steps to set up your TDR system hardware: •...
  • Page 44: Step 1. Set Up The Personal Computer

    4 GB 6 GB+ 6 GB+ Virtual Memory GPIB Interface Keysight 82357A USB/GPIB Interface for No GPIB connection is required to utilize Windows or supported GPIB card (any PLTS in the off-line mode. Saved (stored) National Instruments or Keysight measurement files can be recalled at any 82340/41 or 82350 GPIB card) time for analysis.
  • Page 45: Step 2. Set Up The Tdr System

    Step 1. Set Up the Personal Computer. CAUTION Both the Keysight and Tektronix TDR systems, although not required, may be connected to a computer network. Connecting to a computer network may present security risks to your TDR system. Chapter 2...
  • Page 46 Keysight 86100A/B/C Infiniium Digital Communications Analysis Wide-Bandwidth Oscilloscope with: — Firmware: 86100A/B revision 03.06 or later / 86100C: revision 04.00 or later — 1 or 2 Keysight 54754A 18 GHz Differential TDR/TDT Plug-In Modules installed • Tektronix CSA8000 Communications Signal Analyzer Oscilloscope with: —...
  • Page 47: Step 3. Set Up The General Purpose Interface Bus (Gpib)

    Each test system device must have a unique GPIB address. Check the GPIB address of your Keysight TDR system by selecting Remote Interface from the Utilities menu on the TDR display. To check the GPIB address of your Tektronix TDR system, refer to the programming manual for information.
  • Page 48: Step 4. Power Up The Tdr System

    Installing the TDR-Based Physical Layer Test System Hardware Step 4. Power up the TDR System Step 4. Power up the TDR System If you have not previously powered on your TDR system, start with step 1. If you have already powered on your TDR system, just review steps 1, 2, and 3 before continuing with step 4.
  • Page 49: Installing The Plts Software

    Installing the PLTS Software...
  • Page 50: Important Plts Installation Notes

    Modify, Repair, or Remove the application. Please select Repair. • Microsoft .NET Framework • Keysight IO Libraries & VisaCom Troubleshooting if you encounter problems with PLTS Installation. Chapter 3...
  • Page 51: Step 1. Install Or Upgrade Plts Software

    Step 1. Install or Upgrade PLTS Software Step 1. Install or Upgrade PLTS Software 1. You can download the PLTS software from the Keysight Software Management System (KSM) at http://www.keysight.com/find/softwaremanager. For instructions on downloading the software, refer to your entitlement certificate that you received via email.
  • Page 52 Installing the PLTS Software 5. At the Custom Setup dialog click Keysight IO Libraries Suite <current rev> and select: This feature will not be available 6. Click Next and continue with the installation process. 7. When the installation has completed, click Finish.
  • Page 53: Step 2. License The Physical Layer Test System Software

    SUS subscription, refer to the PLTS User Manual. This can be viewed in the PLTS software on your PC (typically C:\Program Files\Keysight\PLTS_Revxx\Help) with or without an active product license. Also, it can be viewed online at http://na.support.keysight.com/plts/help.
  • Page 54: Setting Up The Fixed (Node-Locked) License

    The default directory that the software is loaded on is: C:\Program Files (x86)\Keysight. Therefore, enter the following command to change to the default directory: cd c:\Program Files (x86)\Keysight\PLTS_<software rev #>\License 3.
  • Page 55 The certificate lists the Order Number and the Certificate Number for your software. These two numbers will also be used to request a license for your software. 6. Go to the Keysight KSM (Keysight Software Management) Web site listed on the “Software License and Software Update Service Subscription Entitlement Certificate.” Follow the instructions at the Web site to receive your license file for the software.
  • Page 56 Installing the PLTS Software Step 2. License the Physical Layer Test System Software Finding the correct Host ID when more than one are listed Sometimes several Host IDs are returned from the lmhostid program, as in the following image. In the case of this laptop computer there are three IDs listed: an Ethernet connection, a Wireless connection, and Ethernet Adapter.
  • Page 57: Setting Up The Networkable (Floating) License

    Installing the PLTS Software Step 2. License the Physical Layer Test System Software Setting Up the Networkable (Floating) License The Networkable (Floating) license allows you to use PLTS on one of many PCs (or clients), utilizing a single license on a common server. The server issues the license to any of the clients that attempt to run PLTS.
  • Page 58 1. Copy the files listed below (in Preparing the License Server) to your server and restart the new version on the server. 2. If already using FlexLm 11.4, then replace the Keysight daemon file and use the new license file. Notes: •...
  • Page 59 The Host ID is the MAC (physical) address without the hyphens. When you request a Networkable license in the Keysight Software Management System (KSM), you are prompted to provide the server MAC (physical) address. For a Windows system, enter the 12 character host ID as the MAC address (do not insert hyphens).
  • Page 60 5. Locate the “Software License and Software Update Service Subscription Entitlement Certificate” that was mailed to you from Keysight. The certificate lists the Order Number and the Certificate Number for your software. These two numbers will also be needed to request a license for your software.
  • Page 61 LMTools dialog box. Select the Config Services tab. On this tab: a. In the Service Name list, select or enter KEYSIGHT License Manager. b. In the Path to the LMgrd.exe file text box, enter the path to the server’s PLTS license directory followed by the lmgrd.exe file name.
  • Page 62 Step 2. License the Physical Layer Test System Software 11. Select the Start/Stop/Reread tab. Select KEYSIGHT License Manager . Then click Start Server. This step must be repeated each time the server is powered up if you did not click the Start Server at Power Up check box earlier in this procedure.
  • Page 63: Setting Up The Transportable (Usb Key) License

    (dongle) here: 2. Locate the “Software License and Software Update Service Subscription Entitlement Certificate” that was mailed to you from Keysight. The certificate lists the Order Number and the Certificate Number for your software. These two numbers will also be used to request a license for your software.
  • Page 64 Installing the PLTS Software Step 2. License the Physical Layer Test System Software 3. Go to the Keysight KSM (Keysight Software Management) Web site listed on the “Software License and Software Update Service Subscription Entitlement Certificate.” Follow the instructions at the Web site to receive your license file for the software. While in the KSM system, you are automatically registered for a 1-year SUS (Software Update Service) subscription.
  • Page 65: Important Plts License Notes

    PLTS orders after July 1, 2020, PLTS uses the KAL license. For the existing PLTS users before July 1, 2020, you can continue to use the legacy Flexlm license or contact Keysight to upgrade to KAL license. The “PLTS License Management” and “PLTS License Reset” names have been changed to “PLTS Legacy License Management”...
  • Page 66: Step 3. Start The Physical Layer Test System Software

    Help file Home page, click Tools and Utilities. 3.2 Procedure Using Keysight Advanced Licensing (KAL) 1. The first time you start the software, you will need to open Keysight License Manager 6 in your Windows start menu to add your KAL licenses:...
  • Page 67 Installing the PLTS Software Step 3. Start the Physical Layer Test System Software 2. Make a selection based on license usage. 3. Select PLTS then click OK. Chapter 3...
  • Page 68: 3.3 Procedure Using Legacy Flexlm Licensing

    Installing the PLTS Software Step 3. Start the Physical Layer Test System Software 3.3 Procedure Using Legacy Flexlm Licensing 1. The first time you start the software, you will need to open PLTS legacy License Management in your Windows start menu to add your legacy Flexlm licenses: 2.
  • Page 69 Installing the PLTS Software Step 3. Start the Physical Layer Test System Software 3. Then, one of the following dialog boxes are shown depending on your type of license: • License If you own a fixed (node-locked) license: Click Browse and select the license file in the directory.
  • Page 70: 3.4 If You Didn't Add The License Before The First Start Of Plts

    Installing the PLTS Software Step 3. Start the Physical Layer Test System Software • If you own a networkable (floating) license: Enter the network computer server name that you recorded earlier. • When the license is found and identified, click Finish. 3.4 If you didn’t add the license before the first start of PLTS When PLTS starts and can't find qualified license, PLTS will display a dialog with instructions.
  • Page 71: Troubleshooting

    Troubleshooting...
  • Page 72: Electrostatic Discharge

    Figure 4-1 Static-Safe Workstation • static-control table mat and earth ground wire: part number 9300-0797 • wrist-strap cord: part number 9300-0980 • wrist-strap: part number 9300-1367 • heel-straps: part number 9300-1308 • floor mat: not available through Keysight Technologies Chapter 4...
  • Page 73: Plts Installation Troubleshooting

    This section provides some background and gives solutions for correcting PLTS installation problems. The following issues are covered in this section: • Connection Manager working with .NET and the T&M Toolkit • PLTS does not recognize the Keysight USB to GPIB adapter. • Problems with the VISA Layer •...
  • Page 74: Connection Manager Working With .Net And The T&M Toolkit

    Connection Manager working with .NET and the T&M Toolkit PLTS Connection Manager communicates with instruments and scans for instruments connected to the bus. The PLTS Connection Manager is built on Keysight's Test and Measurement Programmers Tool Kit (T&M Toolkit), which requires Microsoft's .NET package. Figure 4-2 shows PLTS and its dependencies on other software products and how it communicates with instruments.
  • Page 75: Other Keysight Usb To Gpib Adapter Problems

    To get the adapter to work for either case, make sure the PC is fully powered up and Windows is functional, then unplug the USB connection and reconnect the adapter. The Microsoft Operating System will find the new hardware and the details will be displayed in the Keysight Connection Expert dialog box similar to Figure 4-3.
  • Page 76: Problems With The Visa Layer

    When loading the Keysight I/O libraries second (as with PLTS), the Keysight VISA should be set as the primary VISA. Problems may occur leaving the Keysight VISA as secondary VISA. Refer to Figure 4-4 through Figure 4-8 for more details. To quickly check the...
  • Page 77 The VISA COM must be loaded for PLTS Connection Manager to work. If the IO libraries are already installed, uninstall then re-install the IO libraries. This can be done by re-installing PLTS or by down loading the IO libraries from the Keysight IO Libraries Suite web page: http://www.keysight.com/find/iolib Choose the Typical selection during the PLTS installation.
  • Page 78 PLTS Installation Troubleshooting Figure 4-6 User Analytics Options Click Install. The Typical selection automatically loads the Keysight VISA COM and sets it as the primary VISA. See Figure 4-7. Figure 4-7 Keysight IO Libraries Suite 2019 Start Copying Files Screen If you chose the Custom selection, you must now choose how to install the VISA.
  • Page 79 Troubleshooting PLTS Installation Troubleshooting Figure 4-8 Keysight IO Libraries Custom Installation Screen Chapter 4...
  • Page 80: Problems With Using National Instruments I/O Devices

    5. Uninstall Keysight VISA Com using Start, Settings, Control Panel (Apps for Windows 10), and Add/Remove Programs (Uninstall for Windows 10). 6. Uninstall the Keysight T&M Toolkit using Start, Settings, Control Panel (Apps for Windows 10), and Add/Remove Programs (Uninstall for Windows 10).
  • Page 81: I/O Troubleshooting Steps

    8. Install the new NI driver. Reboot your computer if prompted. 9. Install NI VISA. Reboot your computer if prompted. 10. Install PLTS. Ensure that the Keysight VISA Layer is the primary VISA. See Figure 4-17. 11. For NI Labview code to run properly, you will need to enable TULIP Passport in NI Labview.
  • Page 82 Troubleshooting PLTS Installation Troubleshooting Use the Self-Test function in the Measurement & Automation Explorer to help troubleshoot any problems you encounter while using your GPIB hardware and the NI-488.2 software. Self- Test verifies that your GPIB hardware and the NI-488.2 software are installed correctly and able to perform basic I/O functions.
  • Page 83 PLTS Installation Troubleshooting Troubleshooting the Keysight I/O Libraries Troubleshoot the Keysight I/O libraries by doing the following: 1. When the Keysight IO Control is running, an IO icon is displayed in the task bar as shown in Figure 4-12. Figure 4-12 Keysight IO Control Icon If the Keysight IO Control is not already running, click Start, Keysight Connection Expert.
  • Page 84 Troubleshooting PLTS Installation Troubleshooting 3. To configure an interface, right-click the name of the interface and choose Edit. Figure 4-14 Edit settings for an 82357 USB/GPIB interface Dialog Box 4. For all GPIB interfaces used with PLTS, the VISA Interface Name and SICL Interface Name MUST BE THE SAME (including the case).
  • Page 85: Missing Microsoft Component

    Troubleshooting PLTS Installation Troubleshooting Figure 4-15 Interactive IO Missing Microsoft Component When installing PLTS, if the following error is displayed, comctl32.ocx has to be installed and registered. 'Component 'comctl32' or one of its dependencies not correctly registered: a file is missing or invalid' Chapter 4...
  • Page 86: Plts Hardware Troubleshooting

    Line fuse not installed, or Install the line fuse. the first time you use it. incorrect line fuse installed. Excessive ripple in data. Load termination damaged by Contact Keysight Technologies. excessive RF power. Contacting Keysight for more information. Loose connections between VNA Check and torque the connectors.
  • Page 87: Troubleshooting Plts Using A Tektronix Csa8000 Or Tds8000

    Troubleshooting PLTS Hardware Troubleshooting Troubleshooting PLTS Using a Tektronix CSA8000 or TDS8000 PLTS is designed to use the following Tektronix equipment. • Tektronix CSA8000 Communications Signal Analyzer equipped with one or two 80E04 Dual Channel, 20 GHz TDR Sampling Modules using Firmware Revision 1.3.3 or greater •...
  • Page 88 Troubleshooting PLTS Hardware Troubleshooting • Set the Units to volts (V) c. In the Preset area: • Click the C1 button 4. On the TDR sampling module, verify that the red TDR LED and the yellow SELECT ON/OFF LED for channel 1 (CH 1) is lit. 5.
  • Page 89 Make sure that GPIB Talk/Listen is selected in the GPIB Mode area. 2. Click the Start button in the lower left corner of the PC. Then select Programs, Keysight IO Libraries, then VISA Assistant to open the Visa Assistant dialog box.
  • Page 90 Troubleshooting PLTS Hardware Troubleshooting Figure 4-17 Tektronix CSA8000 Response to *IDN? Button b. Select IEEE 488.2 in the Inst. Lang. area. c. Click the *IDN? button. The PC queries the instrument at the address requesting instrument information. The Tektronix instrument should reply with a response that lists the manufacturer, the model number, and some additional information that includes the firmware version (FV).
  • Page 91 If it is found, continue with Verify the Tektronix Instrument Makes a PLTS Measurement. If it is not found, contact Keysight for assistance. Verify the Tektronix Instrument Makes a PLTS Measurement After PLTS has found and identified the Tektronix instrument, make a basic measurement using the following instructions.
  • Page 92 4-19. Each measurement plot has been autoscaled. Figure 4-19 PLTS-Tektronix Measurement Example Plots 5. If your measurement results are similar to the measurement results shown above, your Tektronix instrument operates with the PLTS software. If it is not found, contact Keysight for assistance. Chapter 4...
  • Page 93: Contacting Keysight

    Assistance with test and measurements needs and information on finding a local Keysight office are available on the Web at: http://www.keysight.com/find/assist If you do not have access to the Internet, please contact your Keysight field engineer. Make sure have the following information readily available: •...
  • Page 94: Safety And Regulatory Information

    Safety and Regulatory Information...
  • Page 95: Safety Information

    Safety and Regulatory Information Safety Information Safety Information Review to the safety information in this section before operating your physical layer test system. Safety Symbols The following safety symbols are used throughout this manual. Familiarize yourself with each of the symbols and its meaning before operating the physical layer test system. CAUTION Caution denotes a hazard.
  • Page 96: Instrument Markings

    Safety and Regulatory Information Safety Information Instrument Markings The instruction documentation symbol. The product is marked with this symbol when it is necessary for the user to refer to the instructions in the documentation. The AC symbol indicates the required nature of the line module input power. This symbol indicates separate collection for electrical and electronic equipment, mandated under EU law as of August 13, 2005.
  • Page 97 Safety and Regulatory Information Safety Information This symbol on all primary or secondary packaging indicates compliance to China standard GB 18455-2001. South Korean Certification (KC) mark; includes the marking's identifier code which follows the format: MSIP-REM-YYY-ZZZZZZZZZZZZZZ or KCC-REM-YYY-ZZZZZZZZZZZZ. Chapter 5...
  • Page 98: Safety Considerations

    Safety and Regulatory Information Safety Information Safety Considerations Familiarize yourself with each of the safety considerations before operating the physical layer test system. NOTE Positioning the Test System for Use When setting up the physical layer test system for use, position the equipment so that the front panel power switch is easy to reach.
  • Page 99 Safety and Regulatory Information Safety Information Before Applying Power CAUTION Install the instrument so that the ON/OFF switch is readily identifiable and is easily reached by the operator. The ON/OFF switch or the detachable power cord is the instrument disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the instrument.
  • Page 100 Safety Information General WARNING To prevent electrical shock, disconnect the Keysight Technologies (N4415A, N4416A, N4417A, N4418A, N4419A/B, N4420B, and N4421A/B) S-parameter test set from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally.
  • Page 101: Regulatory Information

    Operator position am Arbeitsplatz Normal position normaler Betrieb per ISO 7779 nach DIN 45635 t. 19 Declaration of Conformity A Declaration of Conformity is on file for the N1900 Series models, and a copy is available upon request. Chapter 5...
  • Page 102 Index compliance with Symbols .NET Framework Canadian EMC requirements German noise requirements Numerics computer requirements conformity, declaration of 50 ohm cable option cord, ac power CSA symbol C-Tick mark ac power cord frequency range damage from shipment AC symbol declaration of conformity accessories device under test acoustic noise emission...
  • Page 103 Index ground wire model number network analyzer test set handle kit handles N1953A interconnections hardware Error! Bookmark not defined. N1953B interconnections heel straps, ESD N1955B interconnections humidity N1957B interconnections network analyzer options ICES-001 setup icon, software networkable (floating) license identifying Error! Bookmark not defined.
  • Page 104 Index equipment installation VNA-based PLTS interconnections voltage requirements safety considerations before applying power general safety earth ground warnings before applying power, 99 servicing definition safety earth ground safety considerations general safety symbols sample device under test safety earth ground sample pc board servicing scanning Error! Bookmark not...
  • Page 105 This information is subject to change without notice. © Keysight Technologies 2009 - 2020 June 2020 *N1930-90020* N1930-90020 www.keysight.com...

This manual is also suitable for:

N1930b

Table of Contents