    CALIBRATION PROCEDURE NI 9231 8 AI, ±5 V, 24 Bit, 51.2 kS/s/ch Simultaneous, AC/DC Coupling, IEPE AC Coupling This document contains the verification and adjustment procedures for NI 9231. For more information about calibration solutions, visit ni.com/calibration. Contents Software............................ 1 Documentation..........................
  • Page 3: Test Equipment

    NI cDAQ-9174/9178 USB Chassis Quick Start NI-DAQmx installation and hardware setup NI 9231 Getting Started Guide NI 9231 connection information NI 9231 Datasheet NI 9231 circuitry and specifications NI-DAQmx Readme Operating system and application software support in NI-DAQmx LabVIEW Help LabVIEW programming concepts and reference information about NI-DAQmx VIs and functions NI-DAQmx C Reference Help...
  • Page 4: Test Conditions

    Connect the power source to the NI cDAQ-9178 chassis. Launch Measurement & Automation Explorer (MAX). Right-click the device name and select Self-Test to ensure that the module is working properly. NI 9231 Calibration Procedure | © National Instruments | 3...
  • Page 5: Verification

    Verification The following performance verification procedure describes the sequence of operation and provides test points required to verify the NI 9231. The verification procedure assumes that adequate traceable uncertainties are available for the calibration references. Accuracy Verification Complete the following procedure to determine the As-Found status of the NI 9231. Set the calibrator to Standby mode (STBY).
  • Page 6: Adjustment

    Set the calibrator to a reference value determined by the array of recommended calibration voltages. Set the calibrator to Operate mode (OPR). Call and configure the NI 9231 adjustment function according to NI 9231 Adjustment Configuration table. NI 9231 Calibration Procedure | © National Instruments | 5...
  • Page 7: Eeprom Update

    Table 5. NI 9231 Adjustment Configuration Physical Channel Reference Value cDAQ1Mod8/AI The actual value of the voltage output of the calibrator Set the calibrator to Standby mode (STBY). Repeat steps d through g for each point in the array. Close the calibration session. Disconnect the NI 9231 from the calibrator.
  • Page 8: Worldwide Support And Services

    1 866 ASK MYNI (275 6964). For support outside the United States, visit the Worldwide Offices section of ni.com/niglobal access the branch office websites, which provide up-to-date contact information. Range equals 5 V. NI 9231 Calibration Procedure | © National Instruments | 7...
  • Page 9 For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your patents.txt ni.com/patents software, the file on your media, or the National Instruments Patent Notice at . You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer ni.com/legal/export-compliance...