Bedfont Micro+ Smokerlyzer Manual page 2

One co monitor for all your smoking cessation needs
Hide thumbs Also See for Micro+ Smokerlyzer:

Advertisement

Contents
Important Information............................................................................. Page 2
How to perform a bump test with a non-smoker...................................... Page 3
How to perform a bump test with verified gas......................................... Page 4
How to perform a calibration.................................................................... Page 5
How to perform a breath test................................................................... Page 6
Important Information:
NOTE: The Micro⁺™ should be wiped down before and after each test. For a full list of recommended
cleaning wipes and solutions, please visit: https://www.bedfont.com/cleaning-bedfont-monitors
How to perform a bump test with a non-smoker:
Prior to undertaking your first CO test, please check the Micro⁺™ Smokerlyzer® is reading correctly by performing a Bump Test.
This can be done by performing a breath test yourself or with a known non-smoker, or by using verified gas
1
Press the breath test icon to begin
2
Follow the instructions
The practitioner will be
on the Micro⁺™ to complete
prompted to inhale and
a breath test
hold their breath
3
If the device is reading correctly, a non-smoker reading will be shown
onscreen; this result should be green and under 6ppm.
If the reading comes back a little high, it could be due to environmental
CO; please verify the Smokerlyzer® with a second non-smoker.
If the reading is too high, you might need to calibrate your machine
There will be a 15 second
The practitioner will be prompted to
exhale slowly into the device, and
countdown
should aim to completely empty their
lungs