≥ V2.3
Firmware
A. Calibration Tips
B. Required Parts
C. Calibration Set Up
30 Washington Ave Ste D Haddonfield, NJ 08033
tel (800)707-5751 ● fax (800)721-2377 ● service@covita.net ● www.covita.net
+™
piCO
Smokerlyzer
© Copyright 2013 coVita LLC All Rights Reserved v020313
1. coVita calibration instructional videos can be found
under the support section of www.covita.net .
2. You must calibrate every 6 months to correct for
sensor drift using 20 ppm Carbon Monoxide (CO)
calibration gas. On occasion your piCO+ may prompt
you for an intermittent calibration.
3. Your piCO+ a has built in calibration
reminders to prompt you to calibrate
every 6 months.
4. Be sure to identify your flow meter-regulator for
correct instructions.
5. Remove flow meter-regulator from can for proper
storage and replace dust cap. (You may hear a pop
and this is normal.)
6.
Ensure that the caliadaptor tubing has no kinks while
the gas is flowing.
A. 20 ppm
Carbon Monoxide (CO)
B. Flow meter-regulator
C. Caliadaptor Tubing
D. D-Piece
™
E. piCO+
Smokerlyzer
1. Remove dust cap from can of gas
2. Ensure the find control valve on the flow meter-
regulator is in the off position.
3. Screw the regulator-flowmeter onto the gas can (some
slight pressure may be required). This can best done
by screwing the gas can into the valve.
4. Attach the small end of the tubing onto the regulator-
flowmeter.
5. Insert a D-piece into the piCO+
6. Insert the large end of the caliadaptor tubing onto the
D-piece.