Display Messages - Datron 1061A User Handbook Manual

Digital voltmeters
Hide thumbs Also See for 1061A:
Table of Contents

Advertisement

Display
messages
Alpha
numeric messages
will
appear on
the
display
when
an
input,
computational
or
memory
error
occurs.
A
message
will
also
appear
when
a
limit
is
transgressed
and
during
the
self test
routine.
Message
Cause
Correction
Error
0
L
Overload
199999(9)
Reduce
input
signal
or
up
range
Error
1
Arithmetic
Overflow
Adjust
C
store,
rear input or
ratio
input
signal
Error
2
Invalid
Data
Entry/ Recall
Adjust
Data
value/ Recall
Error
3
Uncertainty
>
100%
or unspecified
Error 4
Input
Zero
or
calibration
input
error
Adjust
applied input
and
repress key.
Error
5
DC
Self
Test
failure
1
Error
6
Error
7
kf2
Self
Test
failure
AC
Self
Test
failure
j
Refer to Calibration
&
Servicing
^
Handbook
Error
8
I
Self
Test
failure
i
Error 9
Arithmetic
Underflow
Adjust
C
store,
rear input or
ratio
input signal
IPO
Input
Zero
memory
error
Repeat
input zero procedure
FAIL
Calibration
memory
error
Refer to Calibration
&
Servicing
Handbook
Hi
Lt
Hi Limit
transgressed
Lo
Lt
Lo
Limit
transgressed
Hi
Lo
Lt
Hi
and
Lo
Limit
transgressed
Amend
limit
values
PASS
&
Legend*
Self
Test pass message
2-10