Electrolux RCB 24 P Maintenance And Repair Manual page 4

Table of Contents

Advertisement

@
Make sure that no objects or liquids enter the appliance through the vents.
©
Do not constrict or damage the connecting cable when installing or moving the appliance.
©
Have the appliance serviced once a year by a service engineer.
@
Remove the lid before disposing of the appliance, to prevent children from becoming trapped inside.
Dimensions :
TECHNICAL DATA
Length
:
Width
:
Height
:
Gross volume :
Storage and transport capacity :
43 litres
30 whole blood bags, each containing 500 ml
50 packs of red blood corpuscles, each containing 270 ml
Weight:
Energy supply of box :
Connected load :
Energy consumption over
24 hours:
Energy supply of
alarm system :
Safety category :
empty
:
loaded :
29 kg
44 kg
(with 30 whole blood bags, each containing 500 ml)
AC supply voltage :
frequency:
DC battery voltage :
AC supply voltage :
DC battery voltage :
90 V to
264 V
48 Hz to 62 Hz
11V
to
42V
130 W
120W
1.92 kW/h at an ambient temperature of 25°C (77°F)
(mains and battery operation)
4 nickel cadmium storage batteries
Type AA (Mignon 1,2V/R6) 500 to 750 mAh
operating time without mains connection
< 24 hours
charging time during operation: 48 hours
charging time in battery charger: 14 hours at 50 mA
splash-proof protection in accordance with IP 54
Interference suppression :
Low-voltage :
EMC
in accordance with MIL standard 461-462
704D and EC directive 89/336 EEC
Ambient conditions :
in accordance with EC directives 73/23/EEC and 93/68/EEC
temperature range: -32°C (-25°F) to +44°C (111°F)
protect box from heavy rain, sand and snow