1. PERFORMANCE:
Measuring Range
and Sampling time
Flow Rate(L/min.)
Sampling amount
2nd Pressure
Operating Temperature
Colour Change
1. THE DETECTOR TUBE CONTAINS CHEMICAL REAGENTS.
2. DO NOT TOUCH THESE REAGENTS DIRECTLY ONCE TUBES WERE BROKEN.
3. KEEP THE TUBES OUT OF THE REACH OF CHILDREN.
NOTICE
1. USE ONLY WITH THE COMPRESSED BREATHING AIR SAMPLING SYSTEM (P-41R).
OTHERWISE, CONSIDERABLE ERROR IN INDICATION MAY OCCUR.
2. DO NOT USE THIS TUBE OUTSIDE THE STATED OPERATING TEMPERATURE RANGE.
3. STORE TUBES IN A COOL AND DARK PLACE (0-25 ℃/32-77°F), AND USE BEFORE
EXPIRATION DATE PRINTED ON THE TOP OF THE BOX.
4. PRIOR TO USE, READ ITEM 7. USER RESPONSIBILITY CAREFULLY.
5. READ THE CONCENTRATION IMMEDIATELY AFTER DRAWING THE SAMPLE.
2. SAMPLING AND MEASUREMENT:
1) Without connecting detector tube, make the position of Open/Shutoff valve ⑥ to (A)and open the Needle
valve ⑤ fully. Adjust the 2nd pressure to a preset condition while looking at the 2nd pressure gauge ②
through the pressure regulator ③.
INSTRUCTION MANUAL
COMPRESSED BREATHING AIR TEST TUBE
: 2 - 24 %
approx. 1 minute
2
50 mL
2
0.6 Kgf/cm
(0.059 MPa)
: 0 - 40 ℃ (32 - 104°F) (Temperature correction is necessary.)
: White → Brown
① Part to connect gas cylinder
② 2nd pressure gauge
③ Handle of pressure regulator
④ Flow meter
⑤ Needle valve
⑥ Open/Shutoff valve
⑦ Inlet to gas detector tube
⑧ Gas detector tube
Fig.1
OXYGEN
CAUTION
Fig.2
No.604SP

Advertisement

loading
Need help?

Need help?

Do you have a question about the OXYGEN and is the answer not in the manual?

Questions and answers

Summary of Contents for GL Sciences OXYGEN

  • Page 1 INSTRUCTION MANUAL COMPRESSED BREATHING AIR TEST TUBE OXYGEN No.604SP ★ READ THIS INSTRUCTION MANUAL AND THE INSTRUCTIONS OF THE COMPRESSED BREATHING AIR SAMPLING SYSTEM (P-41R) PRIOR TO USING THIS PRODUCT. ★ DO NOT DISCARD CAREFULLY THIS INSTRUCTION MANUAL UNTIL ALL THE TUBES IN THIS BOX ARE USED UP.
  • Page 2 WARNING THE TEMPERATURE INDICATOR IN THE OXYGEN DETECTOR TUBE BECOMES HOT AFTER HEATING. TO AVOID POSSIBLE INJURY, DO NOT TOUCH THE PART NEAR THE TEMPERATURE INDICATOR OR WEAR SAFETY GLOVES WHEN HANDLING THE OXYGEN DETECTOR TUBE.