Cot does not lower and lock
into the transfer
Scenario
Trolley arms do not lower to lock the
cot into the transfer when loading.
1. Firmly press the manual release button on the
Power-LOAD control panel. The cot will then
lower and lock into the transfer.
Trolley assembly
Transfer assembly
Power-LOAD control panel:
Manual
release
button
Battery
Error
LED
LED