Cot fastener system
Hide thumbs Also See for Power-LOAD:

Advertisement

Cot will not unlock
from the trolley
Scenario
Cot will not release from the
trolley in the loading position
and the error LED
is illuminated.
1. Firmly press the manual
release button on the Power-
LOAD control panel to release
pressure in the hydraulics.
Trolley assembly
Transfer assembly
2. To lower the lifting arms, press
the extend (+) button on the
cot or press down (↓) on the
Power-LOAD control panel.
Note - You may need to apply pressure
to the lifting arms while you press the
manual release button to lower the
lifting arms completely.
Power-LOAD control panel:
Battery
Error
LED
LED
3. Remove the cot from
the trolley.
Manual release
button