  • Page 1 Power-LOAD ® Cot Fastener System Troubleshooting guide 2018/12 6390-069-002 B.0 Refer to the Power-LOAD Cot Fastening System Operations Manual ®...
  • Page 2 Introduction disclaimer These instructions are only intended to serve as short-term solutions. Please contact your local Stryker Service Technician or Account Manager to schedule a maintenance inspection following any incident where troubleshooting is required. Please see your manual for complete troubleshooting information.
  • Page 3 Introduction diagram Power-PRO foot end Head end LED indicators lift controls Lifting arms Head end Transfer lock Triggers Power-LOAD control panel Red release lever Trolley Safety assembly hook Transfer assembly Transfer trolley Foot end lock Floor plate...
  • Page 4 Cot gets stuck when unloading Scenario 1. Pull the patient left transfer lock 2. Hold down the red release lever at the foot trigger and push the cot back into the end of the anchor with your thumb and use Cot gets stuck halfway out when you vehicle until it locks into place.
  • Page 5: When Loading

    Cot gets stuck when loading Scenario 1. Check for debris in the 2. Remove the debris so 3. Pull the cot and trolley all 4. If there is no debris in the transfer trolley lock. the roller will spring the way out of the vehicle transfer trolley lock, pull Trolley and cot get stuck up.
  • Page 6 LOAD control panel to release cot or press down (↓) on the trolley in the loading position pressure in the hydraulics. Power-LOAD control panel. and the error LED Note - You may need to apply pressure is illuminated. to the lifting arms while you press the manual release button to lower the lifting arms completely.
  • Page 7 Scenario 1. Firmly press the manual release button on the Power-LOAD control panel. The cot will then Trolley arms do not lower to lock the lower and lock into the transfer. cot into the transfer when loading.
  • Page 8 2. At the foot end of the cot, press the Scenario 1. Press up (↑) on the Power-LOAD retract (-) button to raise the cot control panel to raise the cot Cot begins to collapse and the legs.
  • Page 9 Power-LOAD lifting arms stop when loading Scenario 1. Realign the cot to center the 2. Press up (↑) on the Power- lifting arms between the cot LOAD control panel or the Lifting arms are not centered legs and the head end LED...