Ethernet Remote Communication Interface - LACO Technologies LHCT03D Operation And Maintenance Manual

Helium concentration tester
Table of Contents

Advertisement

Ethernet Remote Communication Interface

ID
0
1
2
3
4
5
6
7
8
LACO Technologies
The ethernet settings for the remote communication interface require the following
settings:
The IP address of the HCT for communication is 192.168.1.100 on port 100.
Set controlling hardware to IP address 192.168.1.3 and listen on port 100.
Note: if a different IP address is needed, contact LACO and we will set a software
modification for install (see section 10.3)
All send and reply commands end with a [CR] character. All read commands start with
a "?" and requests or set commands start with a "!" character. Refer to the table below
for all commands and syntax.
Command Description
Start cycle
Start calibration cycle
Abort cycle/ acknowledge
fault
System status
Test Data ready for upload ?DR
Read last test result array
Read current leak rate
Read current % helium
concentration
Read max % helium
concentration after test
*Test ID data will only be displayed if configured to be logged.
LHCT03D & LHCT03D-R Operations and Maintenance Manual
9 ETHERNET REMOTE COMMUNICATION INTERFACE
RS-232
Response Example
Command
!CS
CS
!CC
CC
!CA
CA
?SS
SS: Recovery or SS: Ready
or
SS: In Cycle or SS: Fault
DR: YES or DR: NO
?TD
TD: (Date)- (Time)-
(TestID#1)-(TestID#2)-
(TestID#3)-(TestID#4)-
(TestID#5)-(Max % Helium
Concentration)
?LR
LR: 1.23E-9
?HC
HC: 12.3
?MH
MH: 10.8
Validity
Ready only
Ready only
In-cycle only
All
All
Test Data Ready
= Yes.
All
In-cycle only
Test Data Ready
= Yes.
35