6 OPERATING PROCEDURES
28

Parameters Screens

Test Timer (sec) – Entered desired gas sampling time. Minimum time is 10
seconds and maximum is 300 seconds.
Test % Helium – This is the default or typical helium concentration % that is
measured.
% Helium Tolerance Range – This range helps the customer determine if their
gas sampling is in or out of their desired range. For example, if a customer is
measuring 50% helium concentration with a tolerance of 2% the system can
warn if the measured gas mixture is not between 48-52%.
Warning Out of Tolerance – When selected this option allows a screen
warning text to appear on the test screen during testing if the measured
mixture is out of tolerance.
Perform internal calibrations on all non-Alcatel leak detectors at least monthly
and preferably weekly.
Ensure the leak rate on the test screen matches that of the leak detector
output.
Ensure leak detector is on for at least 15 minutes before testing.
Change backing pump oil regularly.
Ensure leak detector is serviced regularly, whether by LACO or in-house.
Ensure leak detector is not in zero mode when in test.
If the leak detector is a TITANTEST there are three modes in the TITANTEST:
Vacuum, Auto Test, and Sniff. Ensure the leak detector is always set to
Vacuum test mode.
LHCT03D & LHCT03D-R Operations and Maintenance Manual