ADJUSTMENT
Adjust VR53 (""CH1G" - VERT PCB) for 5 divisions height.
After adjustment, check the channel A variable gain control.
With the control in minimum position, the waveform must
decrease to less than 2 divisions.
Switch back to position CAL.
REQUIREMENTS
Check all positions of the AMPL/DIV switch.
input voltage
AMPL/DIV setting
display
+3 %
20 mV
5 mV
4 div
50 mV
10 mV
5 div
0,1 V
20 mV
5 div
0,2 V
50 mV
4 div
0,5V
0,1 V
5 div
1V
0,2 V
5 div
2V
0,5V
4 div
5V
1V
5 div
10 V
2V
5 div
20 V
5V
4 div
SETTING
Vertical mode - B
Channel B coupling - DC
Channel B AMPL/DIV - 0,1 V
Variable control in CAL.
Apply a 1 kHz square-wave signal to channel B, peak to peak
amplitude 500 mV.
ADJUSTMENT
Adjust VR153 ("CH2G" - VERT PCB) for 5 divisions height.
After adjustment, check the channel B variable gain control.
With the control in minimum position, the waveform must
decrease to less than 2 divisions.
Switch back to position CAL.
REQUIREMENTS
Check all AMPL/DIV steps of channel B as specified for A.
5.3.11.
Vertical x5 MAG gain
TEST EQUIPMENT
Square-wave calibration generator
SETTING
Vertical mode - A
Channel A coupling - DC
Channel A AMPL/DIV - 0,1 V
Pull the channel A variable control into x5 position.
Apply a 1 kHz square-wave signal to channel A, peak to peak
amplitude 0,1 V.
ADJUSTMENT
Adjust VR30 ("x5G" - VERT PCB) for a waveform of 5 divisions.
SETTING
Repeat for channel B.
ADJUSTMENT
Repeat for channel B by means of VR130.
5.3.12
Channel B invert adjustment
TEST EQUIPMENT
NO
SETTING
Vertical mode - DUAL
Set for two straight lines A and B and place the two position knobs
in centre position (12 o'clock). Check whether the knobs are
correctly mounted at the axis (well centered).
ADJUSTMENT
Adjust VR148 ("POS CENT" - VERT PCB) in that way, that trace B
does not move when switching the position knob to INVERT.
After that, adjust VR332 for coincidence of trace B and the X-axis
the graticule. After that, adjust VR48for coincidence of trace A and
the X-axis of the graticule.