Philips PM3208 Service Manual page 15

20 mhz dual channel oscilloscope
Table of Contents

Advertisement

2 CHARACTERISTICS
CHARACTERISTICS
SPECIFICATION
ADDITIONAL INFORMATION
2.7
EXTERNAL
HORIZONTAL
DEFLECTION
Deflection sources
Channel deflection:
deflection coefficient
error limit
Frequency response:
lower transition point of BW
upper transition point of BW.
Deflection via EXTernal
input:
deflection coefficient
error limit
Frequency response:
lower transition point of BW
upper transition point of BW
Max. phase diff. between
HOR. and VERT.
2.8
TRIGGER
MTB
Trigger sources
Trigger modes:
triggered
auto
single sweep
Coupling of channel
Channel A
EXTernal input
see channels A and B
+10%
see appropriate channel
> 1 MHz (—3 dB)
0,1 V/div
*+15%
d.c.
10 Hz (-—3 dB)
1A
= 1 MHz (—3 dB)
15 kHz
Channel A
Channel B
External
Line
TRIG
AUTO
SINGLE
DC
AC-LF
TV
2.7 EXTERNAL HORIZONTAL DEFLECTION
Selected by MTB trigger source
selector
Except error limit
Measured over central 8 div
Input coupling of channel and
coupling of MTB trigger are
cascaded
Coupling switch in DC position
Coupling switch in AC or AC-LF
position
Coupling switch in DC position
Coupling switch in AC-LF position
Up to 50 kHz
Line trigger derived from mains
Time-base starts when trigger
occurs
Same as TRIG, but bright baseline
is displayed if during 0,1 s no
trigger has occurred
Same as TRIG, but only for first
time after SINGLE has been
activated and after RESET
Approx. 10 Hz (— 3 dB)
Bandpass: approx.
10 Hz...15 kHz (— 3 dB)
TVF: 0,5 s/div...0,1 ms/div
TVL: 50 ws/div...0,2 s/div

Advertisement

Table of Contents
loading

Table of Contents