Philips PM3208 Service Manual page 14

20 mhz dual channel oscilloscope
Table of Contents

Advertisement

CHARACTERISTICS
SPECIFICATION
ADDITIONAL INFORMATION
2.4 EXTERNAL
INPUT
Input connector
BNC
Input impedance:
R parallel
1MQ+
2%
Measured at frequencies
< 1 MHz.
C parallel
30 pF
Input voltage limits
+100 V
Instrument should be properly
(d.c. + a.c. peak)
grounded through the
protective-ground conductor of
the power cord
2.5
HORIZONTAL
DEFLECTION
OR
X AXIS
Deflection sources
MTB
EXT. through Ch. A
See Section 2.7
Through EXTernal input
Selected by MTB trigger source
selector
Line
2.6
MAIN
TIME
BASE
Modes
repetitive
single
Deflection coefficient:
TB magnifier: OFF
steps
0,2 us/div...0,5 s/div
In a 1-2-5 sequence of 20 steps
error limit
3%
Measured over central 8 div
TB magnifier: x10
steps
20 ns/div...50 ms/div
In a 1-2-5 sequence of 20 steps
error limit
+5 %
Measured over central 8 div
VARiable gain range
2,5: 1 (minimum)
Uncalibrated, continuously
variable between steps
Hold-off:
variable between
1...2x minimum hold-off
2.4 EXTERNAL INPUT

Advertisement

Table of Contents
loading

Table of Contents