Erform Ance Est - HP 435B Operating And Service Manual

Power meter
Table of Contents

Advertisement

P
erforma n ce
T
ests
Mo del 435 Β
4-7 .
I
N
ST
RUMEN
TATIO
N
ACC
UR
ACY T
E
ST W IT
H
CA
L
I
BR
ATO
R
(Cont'd)
R
A
N
G
E
Switc
h
P
ositio
n
fully ccw
1 step cw
2 steps εω
3 steps cw
4 step s cw
6.
Set th e
P
ower Meter RAN GE
s
witch to each p ossibl e
p
osition in
t
urn. Set the
11683Α RANGE
s
wi tch to th e same
p
osi tion
a
nd ve rify
t
hat t he DVM rea din g,
whic h includes
n
oise and drift,
i
s within
t
he limi ts shown in
t
he ta ble
b
e low.
Min .
mVdc
+975
+978
+981
+984
+987
R
es
u
lts
Act
u
al
mVdc
PERF
O
RM
A
N
C
E
T
E
STS
Μαχ.
mVdc
+1025
+1022
+1019
+1016
+1013
R
A
N
G
E
Switc
h
P
ositio
n
5 steps εω
6 steps εω
7 step s cw
8 steps εω
fully εω
Min .
mVdc
+998
+990
+990
+990
+990
R
es
u
lts
Act
u
al
mV dc
Μαχ.
m
V
dc
+1002
+1010
+1010
+1015
+1015

Advertisement