SECTION 3 Start-up and operating procedures (continued)
- calibration is then carried out and ends when playback is at ZERO CONTINUOUSLY.

Illustration 9 | Calibration completed to zero

3.4.3

Calibration In-Line Brix Meter failure

If calibration fails, the display shows "Failure to zero".
Illustration 10 | Failure to zero
In this case, repeat the procedure for calibrating the In-Line Brix Meter, refer to Section 3.4.2: Calibration
procedure for In-Line Brix Meter.
If the problem persists, contact your LAPIERRE distributor.

3.5 READ THE BRIX DEGREE OF CONCENTRATE

When the concentrator is switched on, the In-Line Brix Meter displays a Brix degree reading by default.
Illustration 11 | Reading the Brix level of concentrate
3.6 APPLY A CORRECTION VALUE TO THE ACTUAL VALUE READ BY
THE IN-LINE BRIX METER
This function is useful, during operation, for correcting the actual value read by the In-Line Brix Meter.
The range of possible correction values is from -1.0 to +1.0 Brix degrees, in 0.1 degree increments.
How to determine the value | The correction value can be determined using a refractometer (not supplied).
The difference between the refractometer value and the In-Line Brix Meter value becomes the correction value
to be applied to the meter.
Why apply a correction value | This correction can compensate for a distorted reading during operation.