Cable connections
in the power cabinet
4-3
Fig. 4
Fig. 5
Shift the pallet with high-voltage
transformer
Hl in front of the cabinet.
insert the brass rod from the service accessory
pack through the two rings on
the transformer.
Insert the Hl in the installation
location in the power cabinet;
make sure that the transformer
is inserted correctly
in the holders (8iFig.5)
;
otherwise it will tilt when swivelling out.
Reattach safety bracket (6IFig.4) to the Ml6 frame.
Primary
leads on the high-voltage
transformer
(Fig.4)
l
Swivel out high-voltage
transformer
Hl;
for this, unlock the locking pin (1) on the left side of the Hl by withdrawing
it
and swivel the Hl to the front until the locking pin locks in.
l
Connect the ground wire cable (160) coming from the ground wire bus to
Hl. @ (7).
l
Fix the leads of the inverter modules (2) according
to designation:
Hl .Ul,
Hl .Vl, Hl .U2, Hl .V2
l
Connect the leads of the cable harness according
to designation:
a) Primary leads of the filament (3) of D822 are plugged into: Hl .l 1,12,21,22
b) Leads of the tube selection
(4) of D801 are plugged into:
H1.50,51,52
l
Check the oil level (approx. 1 cm) in the high-voltage
plug connectors
;
if necessary,
fill up with oil from the Hl
Siemens
AG
Medical
Engineering
Group
RLL5310.031.01.01.02