COM Object Reference
SCPI.CALCulate(Ch).SELected.TRANsform.DISTance.STATe
SCPI.CALCulate(Ch).SELected.TRANsform.DISTance.STA
Te
Object type
Property
Syntax
SCPI.CALCulate(Ch).SELected.TRANsform.DISTance.STATe = Status
Status = SCPI.CALCulate(Ch).SELected.TRANsform.DISTance.STATe
Description
For the active trace of channels 1 to 4 (Ch), turns ON/OFF the transformation function of
the fault location function.
You can enable the transformation function only when the sweep type is the linear sweep
and the number of points is 3 or more. If you execute this object to try to enable the
transformation function when the sweep type is other than the linear sweep or the number
of points is less than 3, an error occurs and the object is ignored.
When the sweep type is the power sweep, you cannot turn on the transformation function.
If you execute this object trying to turn on the transformation function during the power
sweep, an error occurs and the object is ignored.
Variable
For information on the variable (Ch), see Table C-2, "Variable (Ch)," on page 159.
Examples
Dim Trans As Boolean
SCPI.CALCulate(1).PARameter(1).SELect
SCPI.CALCulate(1).SELected.TRANsform.DISTance.STATe = True
Trans = SCPI.CALCulate(1).SELected.TRANsform.DISTance.STATe
Related objects
Equivalent key
[Analysis]
168
Status
Description
ON/OFF of the gating function
Data type
Boolean type (Boolean)
Range
Select from the following.
•True or -1
•False or 0
Preset value
False or 0
-
Fault Location
Turns ON the transformation function.
Turns OFF the transformation function.
-
Fault Location
Appendix C