Guidance and manufacture's declaration – electromagnetic immunity
The ri-former
Diagnostic Station is intended for use in the electro-
®
magnetic environment specified below. The customer or the user of the
Diagnostic station ri-former should assure that it is used in such an
environment.
IEC 60601
Immunity test
test level
Conducted RF
IEC 61000-4-6
Pass
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2,7 GHz
Proximity fields
Pass
from RF wireless
communi-cations
equipment
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a.)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cord-
less) telephones and landmobile radios, amateur radio, AM and FM radio broadcast and
TV broadcast cannot be predicted theoreticallywith accuracy. To assess the electroma-
gnetic environment due to fixed RF transmitters, an electromagnetic sitesurvey should
be considered. If the measured field strength in the location in which the Diagnostic
station ri-former is used exceeds the applicable RF compliance level above, the Dia-
gnostic station ri-former should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting
or relocating the ri-former
Diagnostic Station.
®
b.)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Compliance
Electromagnetic
level
environment - guidance
Portable and mobile RF
communications equipment
Pass
should not be used closer to
any part of the ri-former
Diagnostic Station, including
the cables, than the recom-
mended distance, which is
calculated using the equation
applicable to the transmitter
frequency.
Recommended separation
distance
3 V/m
d= 1.2√P 150 KHz to 80 MHz
d= 1.2√P 80 MHz to 800 MHz
d= 2.3√P 800 MHz to 2,7 GHz
Where P is the maximum
Pass
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in metres (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
survey,a should be less than
the compliance level in each
frequency range.b
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Where P is the maximum out-
put power of the transmitter
in watts (W) according to the
transmitter manufacturer and
the recommended distance is
given in meters (m).
Field strengths from fixed RF
transmitters determined by
an electromagnetic site
survey a, should be less than
the compliance level in each
frequency range b.
Interference may occur in the
vicinity of devices marked with
the following symbol:
Recommended separation distances between
portable and mobile RF communications equipment and the
Diagnostic Station.
ri-former
®
The
Diagnostic Station is intended for use in an electroma-
ri-former
®
gnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the
prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (trans-
mitters) and the
ri-former
according to the maximum output power of the communications equip-
ment.
Rated maximum
®
output power of
transmitter
(W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d inmetres (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is themaximum output power rating of the trans-
mitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption andreflection from structures, objects and people.

11. Accessories

Art. No.: 3652-600
Art. No.: 3652-500
Art. No.: 3652-400
Art. No.: 10384
Art. No.: 3655-103
Art. No.: 3655-106
Art. No.: 3655-109
Art. No.: 3655-123
Art. No.: 3655-130
Art. No.: 3656
Art. No.: 3656-301
Art. No.: 3654
Art. No.: 14065-531
Art. No.: 14065-534
Art. No.: 10563
Art. No.: 10563-301
Art. No.: 10580
Art. No.: 10580-301
Art. No.: 10565
Art. No.: 10565-301
Diagnostic Station can help
ri-former
®
Diagnostic Station as recommended below,
®
Separation distance according to frequency of
transmitter
(m)
150 kHz to 80
80 MHz to 800
MHz
MHz
d =12√P
d =1,2√P
0,12
0,12
0,38
0,38
1,2
1,2
3,8
3,8
12
12
with foot and big ben® 3.5 V / 100-240 V
ri-former
®
with foot 3,5 V / 100-240 V
ri-former
®
anesthesia model without universal
ri-former
®
clamp 3.5 V / 100-240 V
Universal clamp
Extension module, big-ben
big ben® Wickelman. adult
Hakenman. adult
Klettenman. adult
Klettenman. strong arms
Klettenman. children
Extension module, infrared thermometer
ri-thermo®N
Extension module, infrared thermometer ri-ther-
mo®N without anti-theft device
Extension module, infrared thermometer ri-ther-
mo®N with anti-theft device
ri-spec Ear specula Dispenser Packed in a polybag
25 pcs.
1 box 40 poly bags 1000 pcs. 2.5 mm hopper Packed in
a polybag 25 pcs.
1 box 40 poly bags 1000 pcs. 4 mm funnels Packed in
a polybag 25 pcs.
Instrument heads
ri-former
®
without anti-theft device with anti-theft device
ri-scope® L F.O. Otoscope
L1 3,5 V XL
L1 3,5 V XL with anti-theft device
L2 3,5 V XL
L2 3,5 V XL with anti-theft device
L2 3,5 V LED
L2 3,5 V LED with anti-theft device
800 MHz to
2,7 GHz
d =2,3√P
0,23
0,73
2,3
7,3
23
10