To Save Or Recall Front-Panel Setups - HP 54600A User's And Service Manual

Table of Contents

Advertisement

Operating Your Oscilloscope
To save or recall fro1t-pa1el setups
XY display •ode operatlag
IIIII
When you select the
'l:f
display mode, the time base is turned off. Channel1 is the
X-axis input, channel2 is theY-axis input, and channel4 (external trigger in the
HP 54600A) is the Z-axis input.
H
you only want to see portions
of
the V versus X
display, use the Z-axis input. Z-axis turns on and off the trace (analog oscilloscopes
called this Z -blanking because it turned the beam on and off). When Z is low
(<1.3 V), YversusX is displayed; when Z is high(> 1.3 V), the trace is turned off.
To save or recall front-panel setups
There are 16 memories for storing front-panel setups. Saving
front-panel setups can save you time in situations where several setups
are repeated many times. For example, production test lines.
1
Press
I
setup
I .
2 To change the selected memory location, press either the
left-most softkey or turn the knob closest to the Cursors key.
3 Press the
save
softkey to save a front-panel setup, then press
the
Recall
softkey to recall a front-panel setup.
60
'"-"
'--''
',_;
'._/
,_;
~
"-"
'-"
"-"
\..../
.._,
\..../
,_;
'._/
~I
,_;
'._/
..._,
.._,
_,
'"-"
'._/
\..../
_,
'"-..J
'-"
'"-"
'._/
_,