Wide List Mode Example

Below is a method to load a list via the GUI using a basic text editor like notepad.
Example CSV list for GUI text window: "100, MHz, 10.00, dBm, 10, ms"
Example "Wide List Mode" CSV file with 5 points that step from 95 MHz to 115 MHz in 5
MHz steps at an amplitude of +5 dBm with 100 ms dwell between samples (see figure
below).
If dwell time is not specified on every line of your list, the dwell time is set in the main
menu.
Narrow band files are created and modified the same as the wide list except the
amplitude parameters (C & D) are not available.
NOTE 1: As with the Wide List mode command line format, setting any dwell time (E) inside the
loaded list mode file will cause the global dwell time setting (at list mode control panel) to be
ignored.
NOTE 2: The 5% frequency range limitation for Narrow List mode is defined as: the center
frequency (fc) ± 2.5%.
SELECT FILE: Once the Wide List or Narrow List radio button has been selected, an Open
File window is displayed for the user to select and load a preconfigured, comma-delimited list
file. Selecting a file or pressing the Cancel button will then initiate the list mode control panel as
shown above.