Fluke 8050A Instruction Manual page 5

Hide thumbs Also See for 8050A:
Table of Contents

Advertisement

8050A
ERRATA
#3
On
pages
5-8
and
601-8,
Table
5-2
and
601-3, change
R3,
FROM:
R31
RES, MTL. FILM,
1000
+5%,
1/10W| 5142651
89536]
514265]
1
TO:
R3
]
RES, MTL. FILM,
1000 +.05%,
1/lOW] 514265]
89536] 5142651
1
On page
5-9,
Table
5-2,
add the static sensitive symbol
to U18, U19,
and
U20
.
On page 601-9, Table
601-3,
add the following:
U17^!
IC,
C-MOS,
SI,
N-CHANNEL,
40
PIN
DIP] 524900]
89536] 524900]
1|
1
Ul8*i
IC,
SELECTED
{
8
0
50A-450
6
)
I
5159991
89536]
5159991
1]
1
U19*i
IC,
SELECTED
(8050A-4506-Q1)
I
516005] 89536] 516005]
2|
1
U20*l
IC,
SELECTED
(8050A-4506-01)
j
5160051 89536] 516005] REF
ERRATA
#4
On page
4-7,
Table
4-5,
change the
DISPLAY
READING FOR steps
2
and
3,
FROM:
-37
.28
to -38.28
TO;
-37.63
to -37.93
CHANGE
#1
-
27416
On pages
5-3
and
601-3,
Tables
5-1
and
601-2,
CHANGE:
MP25ITEST
LEAD W/PROBE
(Y8132)
|
516666
I
89536
I
516666
]1
TO:
MP25ITEST
LEAD
(TL70)
i
642033
I
89536
!
642033
1
ERRATA
#5
On page
1-4,
Table
1-2,
for
both DC Current and
AC Current,
change
the
BURDEN VOLTAGE specifications,
FROM:
0.3V max
0
.
9V
max
TO;
0.3V typical
0.9V typical
ERRATA
#6
On page
4-2,
Table
4-1,
add the following
to the
bottom
of
the table:
AC
Calibrator
AC Voltage
100
mV
to
lOOV John Fluke Model 5200A
20
Hz
+/-.!%
On page
4-6,
add the following information:
4-35a.
Use the following procedure
to
verify proper operation
of
the
low frequency ac voltage
measurement functions:
7/85
2

Advertisement

Table of Contents
loading