2. Product Specification Dimensions:17.5cmX13.5cmX14cm Weight:1100g I/P: 100-120Vac, 50/60 Hz Sine Wave, 180W or 200-240Vac, 50/60 Hz Sine Wave, 180W (Please refer to the mark on device.) Device disconnection:pull out the power cord of this device from socket. Max. operating pressure:20kPa Warm-up time:less than 30min.
Page 5
Symbols Glossary Symbol Title and Meaning CE Mark Manufacturer 2460 Date of manufacture Power On/Off (Standby) Protective earth Consult instructions for use Medical device Caution Fragile, handle with care Temperature limit Storage Humidity limitation Keep dry Batch code Use no hooks Catalogue number Handle with care Unique device identifier...
Page 6
(Power Level : 9, room temperature 23℃± 2℃, 120cm-silicone circuit) Performance : Temperature difference between heater plate and patient’s side (room temperature 23℃± 2℃, silicone circuit length 22mmID / 120cm, VADI chamber G-314004): Temperature at Heater plate Flow rate range Power Level patient Y(℃)
3. Troubleshooting Alarm Priority: MID ALARM CONDITIONS REQUIRED ACTION The Low Temperature alarm Check the temperature of activates when the humidifier aluminum tray is normally. detects a low temperature in the aluminum tray. The High Temperature alarm Check the temperature of activates when the humidifier aluminum tray is normally.
2. For patients with low temperature (ex. infants) respiratory humidifier with servo-control is recommended for usage. 3. The VH-1500 can be used in all age which is upper airway malfunction, include in neonatal, child, and adult. 4. The VH-1500 can be applied on any ventilator which include in servo control, invasive ventilation.
5. Controls and Indicators Device control panel has 3 main buttons, power level corresponding with temperature scheme. +Heating Power on Power up (silence button) PWR Lv. PWR Lv. ℃ ℃ -Heating 68 71 68 71 60 64 60 64 Power down Scheme of Power 52 56 52 56...
6. Installation 6.1 Slide the device on the bracket . (Fig.1) (Fig.1) 6.2 Slide chamber onto the heater base. (Fig.2) (Fig.2) Note -Choose the appropriate chamber. 6.3 Add distilled water into chamber.(Fig.3) or stab distilled water with the needle of IV-SET and connect the tube to humidification chamber.(Fig.4) (Fig.3)
Page 11
6.4 Connect the tube from gas supply to inlet port of chamber and the inspiratory part of circuit to the outlet port. Notes 1. Water level should not exceed the highest level and should not be less than the lowest level. 2.
Page 12
6.5 To change power level, press heating-power up/down buttons accordingly. Notes 1. Water vapor in humidification chamber is observed during normal operation. 2. If there is no flow, turn off the humidifier. 3. Water traps are recommended for use to collect water condensation due to differences in temperature.
7. Disposition Please dispose humidifier according to the local rules of the disposition of electronic devices. 8. Clean The heated plate may be wiped with a damp cloth with using clean water or 75% ethyl alcohol. 9. Maintenance 1. Check the power cord for damage, replace if necessary. 2.
Do not use the receptacle that is difficult to disconnect the humidifier. 6. Warning against servicing and maintenance while the ME EQUIPMENT is in use. 7. Serious incident in relation to this product must be reported to VADI, distributor and applicable competent authorities. -12-...