VADI VH-1500 Manual

Respiratory humidifier
Hide thumbs Also See for VH-1500:

Advertisement

Quick Links

VH-1500
RESPIRATORY HUMIDIFIER
Manual
Read instruction carefully before use
Made In Taiwan
VA-1-800-VH1500-110.EN_01

Advertisement

loading

Summary of Contents for VADI VH-1500

  • Page 1 VH-1500 RESPIRATORY HUMIDIFIER Manual Read instruction carefully before use Made In Taiwan VA-1-800-VH1500-110.EN_01...
  • Page 2 Table of contents Table of contents Table of contents Table of contents 1. 1. 1. 1. Intended Use Intended Use Intended Use ........ Intended Use ..............................1 1 1 1 - - - - 2 2 2 2 2.
  • Page 3 ● ● ● ● The VH-1500 can be used in all age which is upper airway malfunction, include in neonatal, child, and adult. ● ● ● ● The VH-1500 can be applied on any ventilator which include in servo control, invasive ventilation.
  • Page 4 2. The filter has to be sterilized before use and reuse. VADI only recommends steam autoclaving with wrapping the filter in sterilization pouch at temperature at 132℃ for 4 minutes (require at least 20 minutes drying cycle after sterilization), complying with the steam autoclave manufacturer’s...
  • Page 5 Type:VH-1500 Dimensions:145mm(H)*170mm(D)*135mm(W) Weight:1.1kg Electrical Ratings Supply voltage:100-120Vac ~1.6A or 200-240Vac ~0.8A (Please refer to the mark on device.) Supply Frequency:50/60Hz Sine Wave Max. power supply: 180W Device disconnection:pull out the power cord of this device from socket. Max. operating pressure:20kpa Warm-up time:less than 30min.
  • Page 6 (Power Level : 9, room temperature 23℃± 2℃, 120cm-silicone circuit) Performance : Temperature difference between heater plate and patient’s side Performance : Performance : Performance : (room temperature 23℃± 2℃, silicone circuit length 22mmID / 120cm, VADI chamber G-314004): Heater plate Flow rate range Temperature at Power Level Temperature (℃)
  • Page 7 3. Controls and Indicators Device control panel has 3 main buttons, power level corresponding with temperature scheme. + + + + Heating Power on Power up (silence button) - - - - Heating Scheme of Power Power down level corresponding (silence button) to temperature Power On: Press the “Power On”...
  • Page 8 4.1 Slide the device on the bracket . ( ( ( ( Pic.1 Pic.1) ) ) ) Pic.1 Pic.1 (Pic.1) 4.2 Slide chamber onto the heater base. ( ( ( ( Pic.2 Pic.2 Pic.2 Pic.2) ) ) ) (Pic.2) ▲Caution: -Choose the appropriate chamber.
  • Page 9 ▲Attention: - Water level should not exceed the highest level and should not be less than the lowest level. - Device cannot be lean over 10°. - Condensed water in circuit is not acceptable. - Connect water trap is recommended. ●Warning: -...
  • Page 10 4.6 To change power level, press heating-power up/down buttons accordingly. Attention: ● Water vapor in humidification chamber is observed during normal operation. ● If there is no flow, turn off the humidifier. ● Water traps are recommended for use to collect water condensation due to differences in temperature.
  • Page 11 5. Performance Data Follow the next steps to conduct product test: 5.1 Put temperature tester on heater plate. 5.2 Set temperature level as 3,check if the temperature is correspondent to the indicator on device panel. 5.3 Set temperature level as 6,check if the temperature is correspondent to the indicator on device panel.
  • Page 12 Absorbent paper for infant chamber Absorbent paper for infant chamber Absorbent paper for infant chamber 10. Manufacturer Information VADI Medical Technology Co., Ltd. Yangmei VADI Medical Technology Co., Ltd. Yangmei VADI Medical Technology Co., Ltd. Yangmei VADI Medical Technology Co., Ltd. Yangmei No.