1.6 INITIAL CHECKOUT

1.6.1 Connect the power cord.
1.6. 2 Press the ON-OFF switch to the "ON" position and push
the desired unit of measurement switch.
1.6. 3 When the display appears, adjust the Zero Control until it
reads "0.00".
1.6.4 Apply a known pressure to the HI port and confirm that
a pressure reading appears. An alternative method is to
press a finger tightly over the HI port. The air in the port
will be compressed sufficiently to produce a reading.
1.6. 5 Plug the continuity test cable into the three test jacks.
Connect each color-coded wire to the matching, labeled
jack.
1.6. 6 Touch the Normally Open (NO) wire to the Common (C)
wire, and confirm that the NO LED (red) lights up.
1.6.7 Touch the Normally Closed (NC) wire to the Common (C)
wire, and confirm that the NC LED (green) lights up.
Instruction Manual DM-6650.02