Philips PM3055 Operation Manual page 41

60 mhz dual time base oscilloscope
Hide thumbs Also See for PM3055:
Table of Contents

Advertisement

5-6E
5.4
TRIGGERING
CHARACERISTICS
SPECIFICATION
ADDITIONAL INFORMATION
5.4.1
MTB triggering
@
Trig- mode
AUTO
(auto free run)
Bright line in absent
of trigger signal
Triggered
Single
@ Trigger source
A, B, Composite (AB), EXT,
Line
@ Trigger coupling
Peak to peak (p/p), DC, TVL,
TVF
@
Level range
Peak to peak
Related to peak to
peak value.
DC
INTERNAL
> +/-8 div.
EXTERNAL
> +/— 800 mV
TVL/TVF
Fixed level
@ Trigger slope
ae
2
@ Trigger sensivity
INTERNAL
0 — 10 MHz
<0.5 div.
@
50 MHz
< 1,0 div.
@
100 MHz
< 2,0 div.
EXTERNAL
0 — 10 MHz
<
50mV
@
50 MHz
< 150 mV
@
100 MHz
< 500 mV
TVL/TVF
INTERNAL
<0,7 div.
EXTERNAL
< 70 mV
Auto free run start 100 ms (typ) after no
trig-pulse.
Switch automaticly to auto free run if one
of the display channels is grounded.
In multi-channe! mode (alternate) each channel
is armed after reset; if sweep is already started,
sweep is not finished.
Line trigger source always triggers on main
frequency.
Line trigger amplitude is depended on line
input voltage.
Approx. 6 div. @ 220 V a.c, input voltage.
p/p coupling is DC rejected.
Slope sign in LCD
+ for POS. and — for NEG. VIDEO
Trig. coupling DC.
Trig. coupling DC.
Trig. coupling DC.
Trig. coupling DC,
Trig. coupling DC.
Trig. coupling DC.
Sync. pulse.
Sync. pulse.
sree,
poco
poe
pee
pte
f
Satin
ry
Rees
ernencesat
peat
poner
((Sitee
A os ath
. a
wd
wore