Philips PM3055 Operation Manual page 39

60 mhz dual time base oscilloscope
Hide thumbs Also See for PM3055:
Table of Contents

Advertisement

5-4E
5.3
5.3.1
CHARACTERISTICS
SPECIFICATION
ADDITIONAL INFORMATION
Lower — 3 dB point
AC coupling EXT. input
<10 Hz
@ 25°C input sinewave 6 div.
10 Hz
Line jump trigger source
<2sd
OFFSET trig.point from screen
< 1,5 sd
centre
Delay EXT. triggerview and
< 10 ns
channel A or B
Dynamic range EXT. input
@ 1 MHz
> +/— 12 div.
@ 50 MHz
>
6 div.
HORIZONTAL DEFLECTION X AXIS
Main Time Base
Time coeff.
0,5 sec. ... 50 ns
Error limit
<3%
Horizontal position range
Start of sweep and
10th div. must be
shifted over screen
centre
Variable control ratio
11> 25
Time Base Magnifier
Expansion *10
Error limit
<4%
Horizontal magnifier balance
<2,5 sd
"10——>"1
Hold off
Minimum to maximum hold
1:>
10
off time ratio
Only when trigger mode is DC.
Jump between trigger source A, B, composite
and EXT.
1,2, 5 sequence (magnifier off).
Measured at —4 ... +4 div. from screen screen
centre.
;
Not valid in X-deflection.
Measured at +4 ... —4 div. from screen centre.
Shift start of sweep in *10 in mid-screen
position then switch to *1.
Minimum hold off time is related to time
base setting.