Page 2
Reproduction of this document in whole or in part is not permitted. Copyright Without the prior written consent of Witschi Electronic AG, no section may be copied, edited or electronically reproduced, transmitted or recorded on magnetic media or in any other way.
Working area of the operator........................... List of equipment..............................Initial operation ..............................3.8.1 Current..............................3.8.2 Compressed air ..........................3.8.3 Thermal printer........................... 3.8.4 Network ..............................Description ................................ProofMaster CP ..............................Included accessories ............................Optional accessories............................Procedure of a tightness test ..........................06.05.2022 3 / 88...
Page 4
Switch on the ProofMaster CP...................... 5.2.2 Enabling screen saver or standby ....................5.2.3 Switching off the device........................5.2.4 Determining test pressure values....................5.2.5 Selecting the program (ProofMaster CP) .................. 5.2.6 Starting the test ..........................5.2.7 Cancelling the test..........................5.2.8 Positioning the watch ........................5.2.9 Manually lowering/lifting the sensor ..................
Page 5
Table of contents Device..................................7.4.1 Opening speed of the bell cover....................7.4.2 Start by pressing the bell cover....................7.4.3 Sensor lighting............................ 7.4.4 Quiet pressure release........................7.4.5 Unit of leak rate..........................Import/export ................................7.5.1 Importing/exporting settings ......................7.5.2 Exporting measurements........................ 7.5.3 Resetting the device settings ......................
These instructions enable safe and efficient handling of the ProofMaster CP leak tightness testing device (referred to in the following as "device" or "ProofMaster CP"). These instructions are an integral part of the device and must be kept near the device so that it can be accessed by personnel at all times.
General information Technical support Rating plate The rating plate of the ProofMaster CP is located on the reverse of the device and contains the following information: • Manufacturer • Device name • Type • Serial number • Year of manufacture...
General information Guarantee Witschi Electronic AG grants two years warranty on the new ProofMaster CP. Further information can be found in the Warranty information document supplied with the ProofMaster CP. 8 / 88 06.05.2022...
Safety Safety This section provides an overview of all the important safety aspects that ensure personal protection and safe and trouble-free operation. The relevant sections contain additional warnings on the various tasks. Mandatory All safety and protection regulations described here must be read and observed in order to avoid damage to equipment, persons and the environment.
2.3.1 Labels on the device The stickers listed below are affixed to the ProofMaster CP. Calibration of the The sticker on the underside of the device shows when the device was last calibrated.
Page 12
Caution - material Material damage due to short circuit! Damage to the insulation on the power cord or the power supply unit can result in a short circuit and damage the ProofMaster CP. Prohibition Never immerse the device in water! It is not designed for immersion.
Page 13
Prohibition Never open the housing yourself! Opening the housing will void the warranty and Witschi Electronic AG will not be liable for any accident occurring after the housing has been opened by an unqualified person.
The watch can only be safely immersed in water for as long as there is positive pressure in the housing. Witschi Electronic AG accepts no liability for watches that are penetrated by water. If the display is red, remove the watch from the water! The positive pressure in the watch is too low and water can penetrate.
Safety Personnel qualifications CAUTION Risk of injury in the event of inadequate personnel qualifications! If unqualified personnel carry out work on the device or stay in the device's danger zone, there is a risk of injury and significant material damage. Mandatory Follow the obligations set out below: All activities may only be carried out by qualified personnel!
Info For a list of spare parts, see Accessories and spare parts [} Page 53]. Noise emissions The noise emissions from the ProofMaster CP are below 75 dB(A) and have been measured according to the guidelines EN ISO 3740:2001-03 and EN ISO 11200:2014-10. 16 / 88 06.05.2022...
Do not remove the ProofMaster CP from its packaging until shortly before commissioning! The packaging protects the ProofMaster CP from external damage. Always carry the ProofMaster CP by the carrying handle or with both hands! Falling over is prevented. Only technical personnel are authorised to install the ProofMaster CP! The technical personnel are trained for proper installation.
Make a note of the extent of the damage on the transport company's delivery note. c) Write a complaint. Mandatory Report damage or missing parts to Witschi Electronic AG immediately! Claims for damages can only be made within the applicable claim periods.
149 mm 305 mm Requirements regarding the installation site The ProofMaster CP is designed for use on a worktable in as dry and dust-free an environment as possible. The ProofMaster CP is not suitable for outdoor use. Caution - material...
Initial operation Working area of the operator In order for the operator to work comfortably, we recommend leaving at least 50 cm space on each side of the ProofMaster CP in order to place the watch shop there. 50 cm...
1 ProofMaster CP Measuring the leak-tightness of watches. ProofMaster CP 2 Power supply unit Provides power to the ProofMaster CP. (12 VDC, 5 A) 3 Scratch Protect Lowerable watch underlay for efficient protection against unwanted scratches. 4 Watch support Stabilises watches <20 mm during the leak-...
Initial operation Initial operation Initial operation of the ProofMaster CP includes connecting the power supply and making all other connections for smooth operation (printer, network, etc.). 3.8.1 Current Caution - material Risk of material damage if an unsuitable power supply...
• Thermal printer (not included in the scope of delivery of ProofMaster • Standard USB cable type A–B or a Bluetooth set II (not included in the scope of delivery of the ProofMaster CP). See Accessories and spare parts [} Page 53]. a) Put the thermal printer into operation as specified in the Connection with USB cable manufacturer's instructions.
The following steps are necessary when using a PC within a network. • ProofMaster CP Required equipment • Ethernet cable (not included in the scope of delivery of ProofMaster CP) • PC incl. network environment (not included in the scope of delivery of ProofMaster CP) a) Ensure that the ProofMaster CP is switched off.
Description Description ProofMaster CP Front view Bell cover Deformation sensor Test chamber Compressed air inlet and outlet Flap over catch Touchscreen The bell cover covers the test chamber. During testing, the bell cover is closed and locked by the catch under flap 5.
Included accessories Power supply unit Power cable (country-specific) Transformer Power connection (on device) Item no.: JA01-PSD60E120K3 The ProofMaster CP is connected to the power supply using the power supply unit with the following key figures: Description Unit Value Input voltage...
Page 27
Description Description Unit Value Output voltage Output current Watch underlay Scratch Protect Watch support Item no.: 31.24.702 (watch support) and 31.24.71.40 (Scratch Protect) The standard support points are suitable for all watches > 20 mm. The Scratch Protect can also be used with these. This scratch protection only lowers the watch once the sensor head is moved to the support points.
Optional accessories Thermal printer The test results can be printed out on the thermal printer. The thermal printer can be connected directly to the ProofMaster CP via the USB cable type A–B or the USB Bluetooth adapter. See Thermal printer [} Page 23].
Test pressure Test pressure Automatic measuring time The ProofMaster CP has a new evaluation algorithm which detects elastic deformation in watches and automatically compensates it. In the standard setting, the ProofMaster CP automatically determines the required measuring time for each measurement. The software continues the measurement until a meaningful result is obtained.
Description Definitions Deformation measurement The measuring method of the ProofMaster CP is based on the principle of deformation. The watch is subjected to positive pressure to compress the watch housing, or a vacuum to expand it. The faster the deformation is compensated, the larger the leak in the watch case. The deformation of the case within a certain time serves as the basis for the measurement to determine whether the watch is leakproof or not.
• Elastic deformation of the test specimen; • Thermal influences. Over a longer measurement duration, the ProofMaster CP measurement algorithm compensates for these influences and displays a more accurate result. This makes it more reliable and faster than an algorithm used for conventional deformation measurement.
ð The home screen appears. ð The bell cover opens. Info When the ProofMaster CP is switched on for the first time, the display language can be selected on the displayed screen. The display language can be changed at any time, see Language selection [} Page 56].
Page 33
ProofMaster CP to standby. ð The bell cover closes and is locked. ð The ProofMaster CP is in standby or screen saver mode (standby screen is always black). ð The screen saver is displayed (date and time in this example).
Enabling screen saver or standby [} Page 32]. ð The bell cover closes and is locked. b) Switch off the ProofMaster CP by pressing the switch in the direction of the arrow. c) Cover the ProofMaster CP with the corresponding dust cover.
Page 35
40.2 N 69.6 N 100.0 N 200.1 N 499.2 N 998.3 N Vacuum The ProofMaster CP can generate a maximum vacuum of 85% compared to the ambient pressure. The maximum vacuum that can be generated therefore varies depending on the height at which the ProofMaster CP is operated.
5.2.5 Selecting the program (ProofMaster CP) Preset programs The vacuum test can be activated or deactivated. a) Press ð The Select program screen appears. b) Determine whether the vacuum test is to be performed. ð It is possible to deactivate the display for this pop-up.
Due to the residual pressure inside the watch, the results would be distorted. Open the watch before the second test to allow the positive pressure to escape. g) Remove the watch from the ProofMaster CP. 5.2.7 Cancelling the test a) Press to stop the test.
Temperature fluctuations in the housing during testing may affect the test result. • Keep the watch next to the ProofMaster CP before testing. • If the watch has come directly from the user, wait until it has reached the ambient temperature.
5.2.9 Manually lowering/lifting the sensor Depending on the selected menu, it is possible to lower and lift the sensor manually. Lowering the sensor a) Press to lower the sensor head onto the watch 2. Lifting the sensor a) Press to lift the sensor head 1. 06.05.2022 39 / 88...
5.2.10 Identifying the tests All measurements are identified by the measured test specimen (model). It is possible to add further identification elements: • Prefix (for example, part number) • User • Comment (4 lines with maximum 32 characters) Enabling identification a) Press to open the identification window.
Measurement results 5.3.1 Displaying test results Info The only result of a leak test to be displayed is whether the test is good or bad. However, it is also possible to enable leak rate measurement. See Creating a program. Test area When the bell cover is opened after a leak test, the test area is illuminated green if the test result is good and red if the test result is bad.
Page 42
Measured deformations for each test Leak rate (only in Precise mode) End result of the test Total testing period The following examples show possible display phases. Test running Test OK Test rejected Test manually aborted Reduced display The reduced display provides the following information: Test specimen prefix Name of the operator (only with enabled identification, in the example above Opera)
In the examples below, the first test is running, the second test result is OK and the third test result is not OK. Test running Test OK Test rejected Test manually aborted 5.3.2 Selecting the results display The results can be displayed in numeric or reduced form. a) Press (at the bottom right of the screen) to show the results in reduced or a numerical display.
Programming 5.4.1 Menus and operating buttons The various function menus can be accessed by pressing the corresponding button 1: Programs menu WiCo menu (appears only when WiCoTRACE is enabled) Functions of the buttons Start a test sequence Stop a test sequence End the countdown Numerical display Reduced display...
3 ATM/30 m. Without vacuum test. Universal NoVac For all types of watch that are classified as water-resistant to 5 bar 5 ATM/50 m. Without vacuum test. Calibrated Leak For testing the function of the ProofMaster CP in combination with Witschi reference leaks. 06.05.2022 45 / 88...
Page 46
Parameters for the preset programs Abbreviation Program name Pressure values Limit values Watch size (bar) (µg/min) (mm) Universal 1 bar -0.2 / 1.0 200 / 50 20 - 40 Universal 2 bar -0.3 / 2.0 200 / 50 Universal 3 bar -0.4 / 3.0 200 / 50 Universal 5 bar -0.5 / 5.0...
Switching on the ChronoMaster Air (switch under the base). ð The ChronoMaster menu is displayed in the bottom left of the screen. ü The ProofMaster CP is switched on, see Switch on the With the USB cable ProofMaster CP [} Page 32].
a) Open the clamping jaws. b) Insert the calibre, ensuring that the winding stem is next to the microphone. c) Release the clamping jaws. d) Press to start a measurement. Display of the current measurement data Measuring position of the calibre Graphic display of the diagram 5.5.3 Configuring the measurement...
Page 49
a) Press Beat rate. Beat rate ð The selection of the test beat rate appears. Auto: Automatic determination of beat rate; Freq: Automatic determination of beat rate for rate deviation = 0 s/d; Manual: Manual determination of beat rate. Range from 3600 to 7200 A/h;...
Danger of material damage due to short circuit! Damage to the insulation on the power cord or the power supply unit can result in a short circuit and damage the ProofMaster CP. Prohibition Never immerse the ProofMaster CP in water! This will destroy the device.
8 years, at the message "Inspection required. Call Witschi service" appears when latest starting the ProofMaster CP. To do so, contact the point of sale. See Technical support [} Page 7]. Comment The number of tests since initial commissioning can be found in the menu Settings >...
Maintenance Cleaning the seal (O ring) ü The ProofMaster CP is switched off and the bell cover is open. a) Clean the seal with a microfibre cloth. Replacing the seal (O ring) ü The ProofMaster CP is switched off and the bell cover is open.
Maintenance Accessories and spare parts Ordering spare parts To order accessories or spare parts, please contact Technical support [} Page 7]. Your nearest point of sale can be found on our website www.witschi.com. Printing Accessories Item no. Thermal printer JB01-SLK-TE25-S Bluetooth Set II to Printer 95.1511...
Settings Settings Quick settings menu ü The ProofMaster CP is switched on, see Switch on the Opening the quick settings ProofMaster CP [} Page 32]. a) Swipe the touchscreen with your finger in the direction of the arrow. ð The quick settings appear.
Settings Settings The settings of the ProofMaster CP can be changed in the settings. Proceed as follows to access the settings: The ProofMaster CP offers quick access – see the Quick settings menu [} Page 54] – and conventional access to the configuration menu settings.
Settings System 7.3.1 Language selection a) Select System from the list. ð System is highlighted in blue and the right window displays the system options. b) Tap the display language to open the selection window. c) Select the desired display language from the list. 56 / 88 06.05.2022...
Settings 7.3.2 Screen saver If the automatic screen saver function of the ProofMaster CP is activated, the bell cover closes after the "Enable after" time set on the ProofMaster CP and the screen saver is displayed. a) Select System from the list.
Page 58
Settings Screen Supported Optimum Description saver formats resolution Images PNG, BMP, 800x480 Images with other resolutions must be resized. JPG, JPEG Videos 800x480 Videos in another aspect ratio are displayed with a black bar. It may not be possible to play high-resolution videos. Deleting a screen saver Info Only imported screen savers can be deleted.
Settings 7.3.3 Display and audio settings a) Select System from the list. ð System is highlighted in blue and the right window displays the system options. b) Configure the display and audio settings as needed. ð See the details below for the settings of each option. a) Move the cursor left to decrease the brightness and right to Brightness increase the brightness.
Mandatory The ProofMaster CP must have an Internet connection! a) Enable or disable the automatic setting of the date and time. ð The button is blue when the function is enabled and grey when it is not.
Select the desired time zone. 7.3.5 Software update With the further development of the ProofMaster CP, the software will also be improved. It is possible to configure an automatic update of the software. The ProofMaster CP must have an Internet connection in order to be able to communicate with the Witschi Electronic AG update server.
Page 62
When the update is accepted, the device will automatically install the new software. The ProofMaster CP must not be switched off during the update. a) Insert the USB stick with the update files for the ProofMaster CP. With the USB stick b) Press Update.
Settings Device 7.4.1 Opening speed of the bell cover a) Select Device from the list. ð Device is highlighted in blue and the right window displays the device options. b) Open the bell cover speed selection by tapping the current selection.
Settings ð Device is highlighted in blue and the right window displays the device options. b) Activate or deactivate the function Start by pressing the bell cover. ð The button is blue when the function is enabled and grey when it is not. 7.4.3 Sensor lighting a) Select Device from the list.
Settings 7.4.5 Unit of leak rate a) Select Device from the list. ð Device is highlighted in blue and the right window displays the device options. b) Open the unit selection for the leak rate by tapping the current selection. c) Select the desired leak rate unit.
Press Export settings. ð A window appears confirming the successful completion of the settings export. ð The settings of the ProofMaster CP were exported to the USB stick in the config.json file. d) Remove the USB stick. 66 / 88...
Page 67
Settings Importing a) Insert a USB stick with the settings into the ProofMaster CP. Mandatory The settings to be imported must be combined in a file with the file name config.json. The file is created by carrying out an export.
This function is intended for analysis by Witschi Electronic AG Customer Service. The ProofMaster CP allows you to export measurement data using a USB stick. This enables the measurements that have been performed to be analysed by Witschi Electronic AG.
Settings 7.5.3 Resetting the device settings This process restores the factory settings of the ProofMaster CP. The programs and test results are retained. a) Select Import/export from the list. ð Import/export is highlighted in blue and the right window displays the import, export and recovery options.
Settings 7.5.4 Restoring factory settings This process restores the factory settings of the ProofMaster CP and deletes all program data, measurement results, etc. on the device. a) Select Import/export from the list. ð Import/export is highlighted in blue and the right window displays the import, export and recovery options.
Printer 7.6.1 Configuring the printer The configuration of the thermal printer is carried out on the ProofMaster CP. a) Select Printer from the list. ð Printer is highlighted in blue and the right window displays the device options. b) Enable the printer.
Settings 7.6.2 Configuring the label Configurable areas Logo Header Results Footer The figure above shows the different configurable areas of the label. ü Printer configured. a) Select Printer from the list. ð Printer is highlighted in blue and the right window displays the printer options.
Page 73
ð The Witschi logo is printed on the label. a) Open the logo selection by tapping the name of the displayed logo. Importing the logo b) Insert a USB stick with the desired logo into the ProofMaster CP (BPM or GIF, optimal width: 368 p). c) Press Load new logo.
Page 74
Settings c) Enter the header using the alphanumeric keypad. ð 4 lines with maximum 30 characters. ð The entered header is printed on the label. Selecting the results The measurement result can be displayed in numeric or reduced form. display See examples below. Numeric format Reduced format Not OK...
Settings Network The connection option allows automatic configuration of the ProofMaster CP with the help of a server. a) Select Network from the list. ð Network is highlighted in blue and the right window displays the network options. b) Enable or disable the network.
Read off and note down the IP address of the server. ProofMaster CP Note: The subnet mask must enable communication between the IP address of the PC or the server IP address of the ProofMaster CP. a) The network settings must be configured in the ProofMaster CP, see Network [} Page 75].
Page 77
You must also ensure that the communications port "1111" is not blocked by a firewall. Where necessary, this port number must be adapted in the ProofMaster CP and in the SMX service of the WicoTRACE software. a) Press to perform new measurements with the new programs.
ProofMaster CP 1. ð The Bluetooth dongle is automatically recognised by the ProofMaster CP. ð Switch on the ProofMaster CP and the ChronoMaster Air. b) Select ChronoMaster Air from the list. ð ChronoMaster Air is highlighted in blue and the right window displays the device options.
Settings a) Deactivate the ChronoMaster Air. ChronoMaster Air not ð The button is grey when the function is not enabled and blue coupled when it is. 7.10 System information This menu displays the following information: • Serial numbers • IP address •...
Troubleshooting Troubleshooting Error messages Info Error messages appear when there is a malfunction on the device. The following table contains information on the possible causes of an error message and information about troubleshooting. The most important malfunctions are described below. Contact customer service if the error continues after taking all troubleshooting measures.
Troubleshooting Error message Cause Remedy WiCoTRACE (SMX-Service) error! Ethernet cable faulty or not Check Ethernet cable. connected correctly. No reply received from the SMX- Service. WiCoTRACE configuration not set Check whether the SMX-Service of correctly. WiCoTRACE was started correctly. Check WiCoTRACE configuration. Vibration detected A vibration was detected during the Place the test device in a suitable,...
Switch off the compressor as specified in the manufacturer's documentation. b) Release the air from the compressed air hose of the compressor. c) Switch off the ProofMaster CP. ð The remaining pressure in the test chamber escapes. ð See Switching off the device [} Page 34].
Never force the device cover to open when the device is under pressure! Otherwise parts could be ejected. ü The ProofMaster CP is switched off. a) Switch off the other used accessories properly. b) Pull out all cables (USB/Ethernet). c) Disconnect the power supply unit connection from the local mains network.
Data sheet Data sheet Dimensions and weight Term Value Unit Weight Width Height with closed bell cover Height with open bell cover Depth Technical data Term Value Unit Vacuum range -0.1 to -0.5 * Pressure range 0.1 to 5 Accuracy of the test pressure ±...
Page 85
Data sheet Emissions Term Value Unit Noise emissions < 75 dB(A) Noise emissions for "Quiet pressure release" < 68 dB(A) See Quiet pressure release [} Page 64]. 06.05.2022 85 / 88...