Optex TC-Mi Instruction Manual page 5

Turbidity checker
Table of Contents

Advertisement

How to open / close the Cover
Open/close the cover by loosening/fastening the four cover fixing screws.
Screws for fixing the cover
Caution
Fasten the cover fastening screws surely. If these screws are loose, turbidity
might not be measured correctly due to the light entered in the measurement
bath or dust mixed in the measurement water.
How to open / close the Panel
Open/close the panel by loosening/fastening the four cover fixing screws.
Screws for fixing the cover
Caution
Screws must be tightened securely. Otherwise, sufficient protection cannot
be obtained.
6
How to connect / disconnect the tube
Caution
•Do not apply strong impact or excessive force to the joint. Otherwise, it may damage the
joint, as well as may cause crush, rupture, or disconnection of the tube.
•Water temperature must be between 0 and 40ºC.
•The unit must be isolated not to damage other machines or equipment in case of
water leakage.
•Use insert rings if necessary. To determine whether to insert rings or not, check the
specification of the tube. Selection of insert rings unmatched to the specification of
the tube may cause tube coming-off and water leakage.
•Follow the precautions below when attaching the tube.
- The cutting section must be right angle.
- Outer circumference must have no damage.
- The tube must not be shaped like an ellipse.
- The tube must be fully inserted to the end.
- Ensure that the tube is securely inserted and cannot come off after attached.
MO DE
Water inlet
Water outlet
Connecting
Insert the tube to the tube
fitting.
Tube fitting
Applicable tube diameter for water inlet:
Nylon tube: 8 +/- 0.1 mm dia.
Urethane tube: 8 +/- 0.15 mm dia.
Applicable tube diameter for water outlet:
Nylon tube: 10 +/- 0.1 mm dia.
Urethane tube: 10 +/- 0.15 mm dia.
Disconnecting
Pull out the tube while push the
lip of the tube fitting.
7