THERMOCOUPLE AND FINE WIRE WELDER
The TL-WELD welder is designed for sensor manufacturers
to produce commercial grade thermocouple junctions, and by
users of large numbers of exposed junction thermocouples
such as test and development laboratories where multipoint
temperature sensing of test pieces is required.
No special skills are required and most people will be
capable of producing quality work with minimal practice. A
satisfactory thermocouple junction is produced without using
argon, but where argon is available a momentary purge is
automatically triggered immediately prior to the weld to give
optimum weld integrity.
Safety Note
1. Always protect the eyes with a suitable filter during
welding – never view the weld discharge with the naked
eye.
2. Avoid touching the rear of the welder during operation,
the power switch heat sink may run hot. This is a normal
operating condition.
3. Do not allow the hand to directly contact the welding
electrode during operation.
MODEL TL-WELD